Adam S Arthur1,2, Andy Molyneux3, Alexander L Coon4, Isil Saatci5, Istvan Szikora6, Feyyaz Baltacioglu7, Ali Sultan8, Daniel Hoit1,2, Josser E Delgado Almandoz9, Lucas Elijovich1,10, Saru Cekirge11,12, James V Byrne13, David Fiorella14.
Abstract
INTRODUCTION: The Woven EndoBridge Intrasaccular Therapy (WEB-IT) Study is a pivotal, prospective, single-arm, investigational device exemption study designed to evaluate the safety and effectiveness of the WEB device for the treatment of wide-neck bifurcation aneurysms.Entities:
Keywords: aneurysm; blood flow; coil; device; magnetic resonance angiography
Mesh:
Year: 2019 PMID: 30992395 PMCID: PMC6824604 DOI: 10.1136/neurintsurg-2019-014815
Source DB: PubMed Journal: J Neurointerv Surg ISSN: 1759-8478 Impact factor: 5.836
Figure 1Illustration of the WEB single-layer sphere (A) and the WEB single layer (B) shapes. Sizes range from 4×3 mm to 11×9 mm. The devices are braided composites made from nitinol and platinum that are highly conformable and visible on fluoroscopy.
Antiplatelet status of patients in the US WEB-IT Trial
| Visit | Antiplatelet therapy, % (n/N) | ||
| Dual | Single | None | |
| Baseline | 12.7 (19/150) | 14.7 (22/150) | 72.7 (109/150) |
| Procedure | 69.3 (104/150) | 27.3 (41/150) | 3.3 (5/150) |
| 30 Days | 31.3 (47/150) | 56.0 (84/150) | 12.7 (19/150) |
| 6 Months | 11.3 (16/142) | 62.7 (89/142) | 26.1 (37/142) |
| 12 Months | 7.0 (10/143) | 41.3 (59/143) | 51.7 (74/143) |
The WEB-IT angiographic scale
| WEB-IT scale | Definition | 1-Year angiographic outcome descriptions | 1-Year angiographic schematics | 1-Year angiographic WEB-IT examples |
| Complete occlusion | No contrast in contact with intracranial aneurysm neck or with wall of intracranial aneurysm sac | WEB-IT effectiveness endpoint |
|
|
| Residual neck | Some contrast in contact with intracranial aneurysm neck but no contrast in contact with wall of intracranial aneurysm sac or inside the WEB device | Adequately occluded patients when combined with completely occluded patients |
|
|
| Residual aneurysm | Apparent contrast in contact with intracranial aneurysm sac or inside the WEB device | Patients who may require re-treatment |
|
|
| The WEB-IT angiographic scale is described with definitions. This scale was used by the core laboratory for evaluation of angiographic follow-up at 1 year. Illustrations depict three separate basilar apex wide-neck bifurcation aneurysms and 1-year follow-up imaging from the trial. | ||||