Aortic prosthetic size predictor in aortic valve replacement.

Anh Tuan Vo1, Tomomi Nakajima2, Trang Thi Thu Nguyen1, Nguyen Thoi Hai Nguyen3, Nga Bich Le3, Tri Huu Cao4, Dinh Hoang Nguyen5,6.   

Abstract

BACKGROUND: Patient-prosthesis mismatch (PPM) is a major concern in aortic valve replacement (AVR) and leads to perioperative morbidity and rehospitalization. Predicting aortic annulus diameter pre-procedurally is crucial to managing patients with high-risk of PPM.
OBJECTIVES: To compare preoperative measurements of aortic annulus from echocardiography and CT scan with surgical sizing and develop an imaging-based algorithm to predict PPM.
METHODS: From January 2017 to December 2020, patients underwent AVR at a teaching hospital were examined. The relationship between imaging measurements with operative values was assesed using scatter plots and Pearson's correlation coefficient. Univariable linear regression was then used to build the predictive model.
RESULTS: A total of 144 patients underwent AVR during the study period. Suture types and surgical approaches were not significantly associated with prosthesis size. CT scan-based measurements showed strong correlation with prosthesis size: mean diameter (R = 0.79), perimeter-derived diameter (R = 0.76), and area-derived diameter (R = 0.75). Mechanical valve and tissue valve shared similar correlation coefficients. Prosthesis size predictive models based on CT scan were 12.89 + 0.335 × d for mean diameter, 13.275 + 0.315 × d for perimeter-derived diameter and 13.626 + 0.309 × d for area-derived diameter.
CONCLUSIONS: Preoperative CT scan measurements are a reliable predictor of aortic prosthesis size. Transthoracic echocardiography is a possible alternative, though it is highly performer-dependent and unable to represent the aortic annulus fully. Together, these two imaging modalities can be used to quantitatively anticipate PPM preoperatively.
© 2021. The Author(s).

Entities:  

Keywords:  Computed tomography; Patient-prothesis mismatch; Prosthetic size predictor; Transthoracic echocardiography

Year:  2021        PMID: 34348752      PMCID: PMC8336273          DOI: 10.1186/s13019-021-01601-z

Source DB:  PubMed          Journal:  J Cardiothorac Surg        ISSN: 1749-8090            Impact factor:   1.637


Introduction

Aortic valve replacement (AVR) remains the gold standard for patients with valvular lesions like aortic stenosis. A major complication following AVR is patient-prosthesis mismatch (PPM), a nonstructural dysfunction. First described in 1978 [1], PPM occurs when excessive pressure gradient is generated across a normally functioning prosthetic valve. Its severity is determined by the indexed effective orifice area (EOAi) as follows: not clinically significant (none or mild) when > 0.85 cm2/m2, moderate when between 0.65 and 0.85 cm2/m2, and severe when ≤ 0.65 cm2/m2. While the negative impact of PPM in the early recovery period is controversial, it generally increases perioperative morbidity and rehospitalization due to heart failure and lack of left ventricular mass regression, and eventually long-term mortality [2, 3]. Several factors were found to increase the likelihood of PPM, including female sex, younger age, high body surface area (BSA), left ventricular end systolic diameter, aortic root dimension [4], hypertension, diabetes, renal failure, and utilization of bioprothesis [5]. Among these, calculating BSA from a patient's height and weight has been suggested as the first in a simple three-step algorithm to determine the type and size of prosthesis according to EOAi's reference values [6]. The aforementioned risk factors can easily be recognized clinically and preoperatively; however, their qualitative nature prevents surgeons from knowing whether further operative considerations, such as aortic root enlargement (ARE), are warranted. With the above limitations in mind, multimodality imaging has been proposed as a powerful and comprehensive approach to identify and quantitate PPM [7]. 2D/3D transthoracic (TTE) and multidetector computed tomography (CT) remain the most widely used tools to measure the aortic annulus dimensions pre-procedurally [3]. Previous studies have documented the superiority of CT measurements as compared to echocardiographic values and recommended the former to be routinely included in prosthesis sizing [8]. In practice, however, some inconsistencies still exist between calculated values and actual manufacturers' prothesis size. Hence, a more direct and robust imaging-based algorithm would significantly improve the ability to predict PPM before implantation. Herein, we examined AVR cases from our institution and assessed the correlation between the aortic valve diameter as measured by echocardiography and CT scan preoperatively versus the true size of implanted valve. Our aim is to develop an imaging-based algorithm to predict the prosthesis valve size prior to AVR.

Methods

Study design

This retrospective cohort study included patients who underwent AVR at the Department of Cardiovascular Surgery, University Medical Center at Ho Chi Minh City from January 2017 to December 2020. All patients had 128-slice CT scan and TTE and both were used to calculate the aortic annulus. The aortic annulus is a crown-shaped structure that serves as the insertion point for the aortic cusps. Its highest and lowest points are located at each of the three commissures and between any two of them, respectively. The annulus, which lacks a planar structure, is compressed to the round-shaped prosthesis after conventional AVR. We therefore assumed that prothesis size is correlated with the plane passing all three nadirs of the aortic valve. Using TTE, the diameters of aortic annulus and left ventricular outflow tract (LVOT), as well as the sinus of Valsalva, sinotubular junction (STJ) and ascending aorta, were measured on the parasternal long axis view (PLAX) (Fig. 1). While reproducible, the results were greatly dependent on echocardiographers.
Fig. 1

Measurement of aortic annulus diameters on TTE

Measurement of aortic annulus diameters on TTE On CT scan, we employed multiplanar reconstruction (OsiriX™ software, Bernex, Switzerland) to map out the plane that passes through three nadirs of the aortic valve. The largest, smallest, average, perimeter-derived and area-derived diameters were then calculated (Fig. 2).
Fig. 2

Measurement of aortic annulus diameters on CT scan

Measurement of aortic annulus diameters on CT scan All diameters were measured at end-systole.

Surgical techniques

We performed AVR via three approaches: conventional full sternotomy, upper ministernotomy or second intercostal minithoracotomy. For the minimally invasive procedures, peripheral cardiopulmonary bypass (CPB) was established with femoral vessels. Custodiol® HTK Solution was delivered antegradely into the aortic root or the coronary ostia in patients with severe aortic regurgitation and was repeated every 120 min if necessary. We used single annular sutures for intra-annular valve replacement and ventricular-side pledgeted sutures for supra-annular valve replacement. Pledgeted sutures (supra-annular fashion) were utilized when severe calcifications were found on the valve leaflets and annulus. Finally, transesophageal echocardiography (TEE) was used to assess the surgery results. For mechanical valve, we used SJM™ Masters Series Mechanical Heart Valve (Abbott Laboratories, Chicago, Illinois, USA). For tissue valve, we used Carpentier-Edwards PERIMOUNT Magna Ease (Edwards Lifesciences, Irvine, California, USA).

Statistical analysis

Data analysis was performed using MedCalc version 19 (Medcalc Software Ltd, Ostend, Belgium). The associations between imaging measurements and valve size were evaluated using scatter plots and Pearson’s correlation coefficients. Linear regression model was used to predict the surgical sizing from imaging-based values. As strong correlations between parameters were found, we performed univariable linear regression and selected the highest correlated measurements to build the prediction model (outcome = a + b × measurement; a: intercept; b: slope). The 95% confidence interval (CI) of the slope and intercept are reported. We analyzed the whole dataset and then performed subgroup analyses for tissue valve and mechanical valve. Statistically significant p value was set to equal or less then 0.05.

Results

A total of 144 patients were included in our study. Baseline and operative characteristics are presented in Tables 1 and 2, respectively. Overall, we found no significant differences in the durations of cardiopulmonary bypass, cross-clamp, mechanical ventilation and ICU stay among three surgical techniques. Interestingly, patients undergoing second intercostal minithoracotomy had longer CPB and cross-clamp time but shorter mechanical ventilation length and ICU stay. Regarding adverse outcome, one death occured at postoperative day five due to cerebral hemorrhage. No aortic dissection was recorded. Nine patients required reoperation due to bleeding, including five cases of full sternotomy, three upper ministernotomies and one second intercostal minithoracotomy. Operative approach was changed to full sternotomy in two patients due to uncontrolled bleeding (upper ministernotomy, one case) and poor exposure (second intercostal minithoracotomy, one case).
Table 1

Baseline characteristics

VariableCohort (n = 144)
Demographic variables
Age (years)56.2 ± 14.6
Sex (female)60 (41.6)
Comorbidities, n (%)
 Hypertension65 (45.1)
 Type II diabetes mellitus31 (21.5)
 Preoperative atrial fibrillation49 (34.0)
AVR variables
Etiology of aortic valve lesions, n (%)
 Post rheumatic31 (21.5)
 Degenerative51 (35.4)
 Congenital62 (43.1)
Approach
 Full sternotomy68 (47.2)
 Upper ministernotomy47 (32.6)
 Second intercostal minithoracotomy29 (20.1)
Prosthesis type
 Mechanical valve62 (43.1)
 Tissue valve82 (56.9)
Valve size (mm), mean ± SD21.36 ± 1.8
 Mechanical valve21.2 ± 1.77
 Tissue valve21.5 ± 1.74

AVR aortic valve replacement

Table 2

Intraoperative details

VariableP value
Cardiopulmonary bypass time (minutes)
Total116.5 ± 35.1
 Full sternotomy104.5 ± 36.00.281
 Upper ministernotomy108.5 ± 21.7
 Second intercostal minithoracotomy126.6 ± 25.7
Cross-clamp time (minutes)
Total82.7 ± 21.6
 Full sternotomy69.6 ± 20.20.090
 Upper ministernotomy66.5 ± 16.1
 Second intercostal minithoracotomy92.3 ± 17.5
Mechanical ventilation time (hours)
 Total36.2 ± 28.6
 Full sternotomy43.9 ± 30.70.112
 Upper ministernotomy31.7 ± 30
 Second intercostal minithoracotomy23.5 ± 8.9
Length of ICU stay (days)
Total3.3 ± 1.4
 Full sternotomy3.7 ± 1.30.49
 Upper ministernotomy2.9 ± 1.7
 Second intercostal minithoracotomy2.9 ± 1.3
Postoperative complications, n (%)
Mortality1 (0.69)
Permanent stroke2 (1.4)
Aortic dissection0 (0)
Prolonged mechanical ventilation*21 (14.6)
Renal failure requiring dialysis6 (4.2)
Reoperation for bleeding9 (6.3)
 Full sternotomy5
 Upper ministernotomy3
 Second intercostal minithoracotomy1
Conversion to full sternotomy2
 Upper ministernotomy (n = 47)1 (2.2)
 Second intercostal minithoracotomy (n = 29)1 (6.9)

ICU intensive care unit

*P < 0.05 is considered significantly different

Baseline characteristics AVR aortic valve replacement Intraoperative details ICU intensive care unit *P < 0.05 is considered significantly different Table 3 illustrates prosthesis size as according to two suture techniques, namely intra-annular single suture and supra-annular pledgeted suture, and three surgical approaches. Neither parameters significantly affect the prosthesis size.
Table 3

Prosthesis size according to different suture techniques and surgical approaches

VariablesProsthesis sizeP-value
Suture technique
Intra-annular single suture (n = 73)21.1 ± 1.70.124 (t-test)
Supra-annular pledgeted suture (n = 41)20.8 ± 1.8
Surgical approach
Full sternotomy21.4 ± 1.670.349 (One-way ANOVA)
Upper ministernotomy21.4 ± 1.96
Second intercostal minithoracotomy20.7 ± 1.89
Prosthesis size according to different suture techniques and surgical approaches The relationship between diameters calculated by imaging modalities and the implanted prothesis size was computed using Pearson correlation coefficients (Table 4) and graphed as scatterplots (Fig. 3). TTE-measured diameters showed the weakest correlation, whereas those calculated from CT scan were strongly correlated with surgical prosthesis sizing (coefficients were greater than 0.70 for all diameters and for both mechanical and tissue valve). Hence, CT-based predictive model of aortic prothesis size were developed using univariable linear regression (Table 5).
Table 4

Pearson correlation coefficients between imaging-measured diameters and intraoperative prothesis size

Imaging modalityCalculated aortic annulus diameters
All patientsMechanical valveTissue valve
TTE diameter (PLAX diameter)0.590.650.49
CT scan
 Mean diameter0.790.790.79
 Perimeter-derived diameter0.760.770.76
 Area-derived diameter0.750.730.76

CT computed tomography, PLAX parasternal long axis view, TTE transthoracic echocardiography

Fig. 3

Scatterplots for the correlation of each calculated diameters with prosthesis size

Table 5

CT scan-based prediction model of aortic prosthesis size

Calculated diameterAll patientsMechanical valve groupTissue valve group
Mean diameter

12.89 + 0.335 × d

(R2 = 0.6241)

11.104 + 0.378 × d

(R2 = 0.6241)

13.068 + 0.341 × d

(R2 = 0.6241)

Perimeter-derived diameter

13.275 + 0.315 × d

(R2 = 0.5776)

11.213 + 0.371 × d

(R2 = 0.5929)

13.724 + 0.308 × d

(R2 = 0.5776)

Area-derived diameter

13.626 + 0.309 × d

(R2 = 0.5625)

11.593 + 0.363 × d

(R2 = 0.5329)

13.910 + 0.310 × d

(R2 = 0.5776)

CT computed tomography, d corresponding calculated diameter, R2 R-squared, coefficient of determination of the model

Pearson correlation coefficients between imaging-measured diameters and intraoperative prothesis size CT computed tomography, PLAX parasternal long axis view, TTE transthoracic echocardiography Scatterplots for the correlation of each calculated diameters with prosthesis size CT scan-based prediction model of aortic prosthesis size 12.89 + 0.335 × d (R2 = 0.6241) 11.104 + 0.378 × d (R2 = 0.6241) 13.068 + 0.341 × d (R2 = 0.6241) 13.275 + 0.315 × d (R2 = 0.5776) 11.213 + 0.371 × d (R2 = 0.5929) 13.724 + 0.308 × d (R2 = 0.5776) 13.626 + 0.309 × d (R2 = 0.5625) 11.593 + 0.363 × d (R2 = 0.5329) 13.910 + 0.310 × d (R2 = 0.5776) CT computed tomography, d corresponding calculated diameter, R2 R-squared, coefficient of determination of the model

Discussion

The major finding of this study is that CT-calculated aortic annulus diameter is more reliable than TTE preoperatively and hence could be used to develop a predictive model of prosthesis size, eventually preventing PPM. Several preventive strategies can be considered when PPM following AVR is likely [3]. A newly developed generation of prosthesis, the stentless bioprosthesis, is a valve designed for supra-annular implantation [9]. The new era of transcatheter AVR (TAVR) also promised to lower the prevalence of PPM, with supporting data from Asian population [10, 11]. Nevertheless, these valves might not be commercially available or too costly for patients in developing countries. Under these circumstances, aortic root enlargement (ARE) is commonly indicated to place a larger valve and has been performed for many years. However, not only does this practice add to operation time and complexity, but could also negatively affect morbidity and mortality in the field of minimally invasive surgical AVR, such as mini-thoracotomy and upper ministernotomy [12]. In addition, ARE is associated with several complications, such as mitral valve prolapse and aorto-left atrial fistula [13]. Therefore, recognizing the need for ARE and adequately determining a patient's risk of PPM are crucial prior to AVR. This study compared preoperative CT scan and TTE measurements of aortic annulus diameter to intraoperative annular sizing. Our results agree with previous literature, which indicated that CT-based calculations were better correlated with operative values. In particular, Kempert et al. demonstrated that utilizing the “effective” diameter on CT scan is preferable to TTE values in patients with oval-shaped annulus [14]. Likewise, Daskevish et al. found that the calculated annular measurements with CT are closer to operative sizing with a Hegar dilator [8]. While suggesting that imaging modalities (CT and TTE) could be fairly accurate in predicting the aortic annulus size, both studies shared similar limitations, i.e. small sample size (26–33 patients) and did not establish predictive models. A recent review by Pibarot et al. proposed an algorithm to predict and prevent PPM by using multidetector CT or 3D TEE but still did not provide an estimated prothesis size preoperatively [3]. This could lead to unplanned ARE during AVR and pose major challenges to inexperienced surgeons. The superiority of CT scan over echocardiography can be explained as follows. TTE is echocardiographer-dependent, and a single dimension on PLAX cannot represent the whole aortic annulus. On the other hand, CT scan provides three different diameters (mean, perimeter-derived and area-derived diameter) that are fairly comparable when calculating the prosthesis size preoperatively. In practice, however, echocardiography is non-invasive and presents as the only option when CT scan is contraindicated or unavailable. Hence, measurements predicted from TTE alone could still be considered while bearing in mind that they might not be as accurate as CT scan-based. Our study showed that CT-calculated algorithms for mechanical and tissue valve shared similar coefficients, thus proving CT-based predictor as applicable for both prothesis types. Overall, we recommend combining all three diameters on CT scan to minimize possible errors during measurements. To our knowledge, this is the first study to propose a quantitative, imaging-based model to predict PPM prior to AVR. Knowing each patient's predictive prosthesis size, surgeons can anticipate and develop suitable strategies (eg using valves with larger EOAi, preparing for ARE or performing TAVR in lieu of surgical AVR) before incision. This would in turn help decrease morbidity caused by unintended procedures and lead to better patient outcome. The main limitations of this study are its retrospective nature, small sample size and that few types of prostheses were used to establish the predictors. As different manufacturers produce protheses with similar size but largely inconsistent dimensions, the two most popular prostheses in our center were chosen to increase the model accuracy (SJM™ Masters Series for mechanical valve and Carpentier-Edwards PERIMOUNT Magna Ease for bioprosthesis). As a result, the suggested algorithm might only apply to these two and further studies using a wider variety of prothesis types are needed.

Conclusions

Preoperative CT scan measurements is efficient in predicting the size of aortic prosthesis. If CT is contraindicated or unavailable, TTE is an alternative imaging method, though its dependence on echocardiographers and inability to represent the full aortic annulus limit its accuracy. Together, these two imaging modalities can be incorporated into a quantitative and straightforward algorithm to predict PPM preoperatively.
  14 in total

Review 1.  Hemodynamic and clinical impact of prosthesis-patient mismatch in the aortic valve position and its prevention.

Authors:  P Pibarot; J G Dumesnil
Journal:  J Am Coll Cardiol       Date:  2000-10       Impact factor: 24.094

2.  Aortic annulus sizing: echocardiographic versus computed tomography derived measurements in comparison with direct surgical sizing.

Authors:  Jörg Kempfert; Arnaud Van Linden; Lukas Lehmkuhl; Ardawan J Rastan; David Holzhey; Johannes Blumenstein; Friedrich W Mohr; Thomas Walther
Journal:  Eur J Cardiothorac Surg       Date:  2012-03-07       Impact factor: 4.191

3.  Preoperative assessment of aortic annulus dimensions: comparison of noninvasive and intraoperative measurement.

Authors:  Alexey Dashkevich; Philipp Blanke; Matthias Siepe; Gregor Pache; Mathias Langer; Christian Schlensak; Friedhelm Beyersdorf
Journal:  Ann Thorac Surg       Date:  2011-03       Impact factor: 4.330

Review 4.  Aortic Root Enlargement Is Safe and Reduces the Incidence of Patient-Prosthesis Mismatch: A Meta-analysis of Early and Late Outcomes.

Authors:  Wanqing Yu; Derrick Y Tam; Rodolfo V Rocha; Ahmad Makhdoum; Maral Ouzounian; Stephen E Fremes
Journal:  Can J Cardiol       Date:  2019-02-13       Impact factor: 5.223

Review 5.  Predictors and Outcomes of Prosthesis-Patient Mismatch After Aortic Valve Replacement.

Authors:  Victor Dayan; Gustavo Vignolo; Gerardo Soca; Juan Jose Paganini; Daniel Brusich; Philippe Pibarot
Journal:  JACC Cardiovasc Imaging       Date:  2016-05-25

6.  Incidence, Predictors, and Clinical Impact of Prosthesis-Patient Mismatch Following Transcatheter Aortic Valve Replacement in Asian Patients: The OCEAN-TAVI Registry.

Authors:  Masaki Miyasaka; Norio Tada; Masataka Taguri; Shigeaki Kato; Yusuke Enta; Tatsushi Otomo; Masaki Hata; Yusuke Watanabe; Toru Naganuma; Motoharu Araki; Futoshi Yamanaka; Shinichi Shirai; Hiroshi Ueno; Kazuki Mizutani; Minoru Tabata; Akihiro Higashimori; Kensuke Takagi; Masanori Yamamoto; Kentaro Hayashida
Journal:  JACC Cardiovasc Interv       Date:  2018-04-23       Impact factor: 11.195

Review 7.  Prosthesis-Patient Mismatch After Transcatheter Aortic Valve Implantation.

Authors:  Hisato Takagi; Takuya Umemoto
Journal:  Ann Thorac Surg       Date:  2016-01-29       Impact factor: 4.330

Review 8.  Recommendations for the imaging assessment of prosthetic heart valves: a report from the European Association of Cardiovascular Imaging endorsed by the Chinese Society of Echocardiography, the Inter-American Society of Echocardiography, and the Brazilian Department of Cardiovascular Imaging.

Authors:  Patrizio Lancellotti; Philippe Pibarot; John Chambers; Thor Edvardsen; Victoria Delgado; Raluca Dulgheru; Mauro Pepi; Bernard Cosyns; Mark R Dweck; Madalina Garbi; Julien Magne; Koen Nieman; Raphael Rosenhek; Anne Bernard; Jorge Lowenstein; Marcelo Luiz Campos Vieira; Arnaldo Rabischoffsky; Rodrigo Hernández Vyhmeister; Xiao Zhou; Yun Zhang; Jose-Luis Zamorano; Gilbert Habib
Journal:  Eur Heart J Cardiovasc Imaging       Date:  2016-05-03       Impact factor: 6.875

Review 9.  Aortic valve replacement and prosthesis-patient mismatch in the era of trans-catheter aortic valve implantation.

Authors:  Shigeki Morita
Journal:  Gen Thorac Cardiovasc Surg       Date:  2016-05-27

10.  Predictors of the size of prosthetic aortic valve and in-hospital mortality in aortic valve replacement.

Authors:  Muhammad Shahzeb Khan; Faizan Imran Bawany; Mudassir Iqbal Dar; Muhammad Umer Ahmed; Mehwish Hussain; Mohammad Hussham Arshad; Asadullah Khan
Journal:  Glob J Health Sci       Date:  2014-04-16