| Literature DB >> 34319493 |
Sven Knecht1, Christian Sticherling1, Laurent Roten2, Patrick Badertscher1, Laurève Chollet2, Thomas Küffer2, Florian Spies1, Antonio Madaffari2, Aline Mühl2, Samuel H Baldinger2, Helge Servatius2, Stefan Osswald1, Tobias Reichlin2, Michael Kühne3.
Abstract
PURPOSE: The aim was to report procedural and technical differences of a novel cryoballoon (NCB) ablation catheter for pulmonary vein isolation (PVI) compared to the standard cryoballoon (SCB) catheter.Entities:
Keywords: Atrial fibrillation; Cryoballoon ablation; Technical specification
Mesh:
Year: 2021 PMID: 34319493 PMCID: PMC9399022 DOI: 10.1007/s10840-021-01035-6
Source DB: PubMed Journal: J Interv Card Electrophysiol ISSN: 1383-875X Impact factor: 1.759
Patient characteristics
| Standard CB | Novel CB | ||
|---|---|---|---|
| Age (years) | 66 ± 9 | 65 ± 11 | 0.675 |
| Men | 26 (65) | 26 (65) | 1.000 |
| Paroxysmal AF | 28 (70) | 23 (58) | 0.352 |
| BMI (kg/m2) | 27 ± 4 | 28 ± 7 | 0.897 |
| Left atrial size (mm) | 40 ± 6 | 40 ± 6 | 0.973 |
| Left atrial volume index (ml/m2) | 41 ± 13 | 36 ± 12 | 0.188 |
| Left ventricular ejection fraction (%) | 58 ± 7 | 57 ± 12 | 0.899 |
| Hypertension | 20 (50) | 20 (50) | 1.000 |
| CAD | 7 (18) | 11 (28) | 0.422 |
| Heart failure | 1 (3) | 5 (13) | 0.201 |
Continuous variables are shown mean ± standard deviation and categorical parameters as numbers and percentage
AF, atrial fibrillation; BMI, body mass index; IQR, interquartile range; CAD, coronary artery disease
Fig. 1Dissected NCB (left) and SCB catheter (right). The white lines represent the position of the pole of the balloon and the dash-dotted line highlights the position of the thermocouple (TC). The gray arrows represent the position of the injection coils and the white arrows the position of the gas outflow
Technical specifications
| Standard CB | Novel CB | |
|---|---|---|
| General specifications | ||
| Shaft diameter (F) | 10.5 | 11.8 |
| Balloon design | Double-layer | Double-layer |
| Balloon size (mm) | 23 & 28 | 28 |
| Tip length (mm) | 8 | 5 (short tip) or 12 (long tip) |
| Handle | Slider switch and push button for balloon elongation | Tension nob and a slider switch |
| Handle procedure status | None | LED color code Green: ready Blue: inflation/thaw Flashing blue: ablation |
| N2O injection | 8-hole coil | 8-hole coil |
| N2O fluid flow during freeze (sccm) | 7200 (28 mm) | 7800 |
| Pressure during freeze (psi) | 530–600 | < 525, constant after approx. 12 s |
| Automatic termination | None | Temperature < − 70 °C |
| Sheath outer diameter (F) | 15.0 | 15.9 |
| Sheath stopcock | yes | no |
| Measured specifications | ||
| Location of injection coil (mm) | 3.5 (from pole of balloon) | 2.5 (from pole of balloon) |
| Location of TC (mm) | 25 (from tip) | 25 (from tip) |
| 15 (from coil) | 18 (from coil) | |
| Location of gas outflow (mm) | 10 (proximal of TC) | 5 (proximal of TC) |
CB, cryoballoon; TC, thermocouple
Fig. 2Exemplary temporal evolution of the ice cap formation (border highlighted) for the SCB (top row) and NCB (bottom row) at 60, 120, 180, and 240 s. The black dash-dotted line represents the middle of the CB (equator), the black dashed line represents the minimal coverage, and the white dashed line represents the maximal coverage. Note the slower increase of the maximal coverage with the NCB
Procedural data
| Standard CB | Novel CB | ||
|---|---|---|---|
| Procedure time (min) | 65 ± 17 (62) | 84 ± 29 (84) | 0.003 |
| LA dwell time (min) | 47 ± 22 (43) | 57 ± 25 (54) | 0.053 |
| Net ablation time (s) | 1064 ± 296 (955) | 1086 ± 369 (975) | 0.832 |
| Fluoroscopy time (min) | 20 ± 7 (15) | 25 ± 36 (18) | 0.083 |
| Single-shot isolation LSPV | 31 (78) | 33 (83) | 0.781 |
| Nadir temperature LSPV (°C) | − 48 ± 7 (47) | − 61 ± 6 (− 61) | < 0.001 |
| Temperature at isolation LSPV (°C) | − 35 ± 10 (− 37) | − 50 ± 10 (− 50) | < 0.001 |
| Time to isolation LSPV (s) | 45 ± 23 (37) | 53 ± 25 (45) | 0.131 |
| Single-shot isolation LIPV | 27 (68) | 32 (80) | 0.310 |
| Nadir temperature LIPV (°C) | − 44 ± 5 (− 45) | − 56 ± 5 (− 56) | < 0.001 |
| Temperature at isolation LIPV (°C) | − 23 ± 38 (− 35) | − 43 ± 22 (− 47) | < 0.001 |
| Time to isolation LIPV (s) | 47 ± 42 (43) | 55 ± 34 (40) | 0.561 |
| Single-shot isolation RSPV | 30 (75) | 25 (63) | 0.335 |
| Nadir temperature RSPV (°C) | − 47 ± 6 (− 47) | − 60 ± 7 (− 61) | < 0.001 |
| Temperature at isolation RSPV (°C) | − 34 ± 8 (− 35) | − 41 ± 16 (− 43) | 0.028 |
| Time to isolation RSPV (s) | 52 ± 42 (37) | 45 ± 37 (30) | 0.506 |
| Single-shot isolation RIPV | 25 (63) | 26 (65) | 0.815 |
| Nadir temperature RIPV (°C) | − 47 ± 5 (− 47) | − 59 ± 6 (− 58) | < 0.001 |
| Temperature at isolation RIPV (°C) | − 37 ± 8 (38) | − 46 ± 8 (− 49) | < 0.001 |
| Time to isolation RIPV (s) | 62 ± 44 (45) | 55 ± 37 (37) | 0.647 |
| Interrupted freezing cycles | 1.3 ± 1.5 (1) | 1.2 ± 1.1 (1) | 0.551 |
Continuous variables are shown mean ± standard deviation (median) and categorical parameters as numbers and percentage
IQR, interquartile range; LSPV, left superior pulmonary vein; LIPV, left inferior pulmonary vein; RSPV, right superior pulmonary vein; RIPV, right inferior pulmonary vein; LA, left atrial