Literature DB >> 31046130

Acute outcome after a single cryoballoon ablation: Comparison between Arctic Front Advance and Arctic Front Advance PRO.

Massimo Moltrasio1, Rita Sicuso1, Gaetano M Fassini1, Stefania I Riva1, Fabrizio Tundo1, Antonio Dello Russo1, Michela Casella1, Benedetta Majocchi1, Martina Zucchetti1, Selene Cellucci1, Claudio Tondo1,2.   

Abstract

BACKGROUND: The novel fourth-generation cryoballoon (CB4) potentially allows for enhanced catheter maneuverability and more frequent capture of pulmonary vein (PV) potentials which can be used to monitor real-time PV isolation (PVI). The aim of our study is to compare the acute procedural endpoints between the CB4 and second-generation cryoballoon (CB2).
METHODS: A single-center retrospective chart review was used to examine 50 consecutive patients with drug-refractory atrial fibrillation undergoing CB4-based PVI. Procedural data and acute success of these patients were compared to 50 propensity-matched controls who underwent cryoballoon ablation procedure using CB2.
RESULTS: Procedures performed with the CB4 showed significant shorter fluoroscopy time (14.8 ± 5.5 vs 18.0 ± 6.5 minutes, P = .04), shorter procedure time (58.3 ± 15.7 vs 65.3 ± 21 minutes, P = .13), and shorter total ablation time (10.8 ± 1.5 vs 13.8 ± 1.9 minutes, P = .42). The real-time PVI visualization rate was 33.3% in the CB2 group and 74.7% in the CB4 group (P < .001). CB4 was correlated to significant increase of acute real-time recordings with regard to all the single PV (left superior PV: 58% vs 84%, P = .02; left inferior PV: 26% vs 71%, P = .001; right superior PV 29% vs 61%, P = .01; and right inferior PV 19% vs 58%, P = .002).
CONCLUSION: The CB4 was more often able to capture real-time recordings of PV potentials and the subsequent acute PV isolation.
© 2019 Wiley Periodicals, Inc.

Entities:  

Keywords:  atrial fibrillation; catheter ablation; cryoablation; cryoballoon catheter; pulmonary vein isolation

Mesh:

Year:  2019        PMID: 31046130     DOI: 10.1111/pace.13718

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

1.  Comparison of procedural efficacy and biophysical parameters between two competing cryoballoon technologies for pulmonary vein isolation: Insights from an initial multicenter experience.

Authors:  Sing-Chien Yap; Ante Anic; Toni Breskovic; Annika Haas; Rohit E Bhagwandien; Zrinka Jurisic; Tamas Szili-Torok; Armin Luik
Journal:  J Cardiovasc Electrophysiol       Date:  2021-02-01

2.  Technical and procedural comparison of two different cryoballoon ablation systems in patients with atrial fibrillation.

Authors:  Sven Knecht; Christian Sticherling; Laurent Roten; Patrick Badertscher; Laurève Chollet; Thomas Küffer; Florian Spies; Antonio Madaffari; Aline Mühl; Samuel H Baldinger; Helge Servatius; Stefan Osswald; Tobias Reichlin; Michael Kühne
Journal:  J Interv Card Electrophysiol       Date:  2021-07-28       Impact factor: 1.759

  2 in total