J Scott Rankin1, Domenico Mazzitelli2, Theodor J M Fischlein3, Yeong-Hoon Choi4, Diana Aicher5, Lawrence M Wei1, Vinay Badhwar1. 1. Department of Cardiovascular and Thoracic Surgery, West Virginia University Heart and Vascular Institute, Morgantown, WVa. 2. Department of Cardiac Surgery, Klinikum Bogenhausen, Munich, Germany. 3. Department of Cardiovascular Surgery, Klinikum Nürnberg, Paracelsus Medical University, Nürnberg, Germany. 4. Department of Cardiac and Thoracic Surgery, University of Köln, Köln, Germany. 5. Department of Cardiac Surgery, HGZ Bad Bevensen Herzchirurgie, Bad Bevensen, Germany.
Abstract
OBJECTIVE: As bicuspid aortic valve (BAV) repair evolves, more effective annular reduction and stabilization could be advantageous. A geometric annuloplasty ring has been developed, and 2-year regulatory outcomes of a first-in-humans pilot trial are reported. METHODS: A prospective first-in-humans trial of BAV ring annuloplasty was completed in 16 patients. Patient age was 44.4 ± 11.3 (mean ± standard deviation) years, preoperative aortic insufficiency grade was 2.5 ± 1.0, New York Heart Association class 1.8 ± 0.4, and mean systolic gradient 13.4 ± 12.9 mm Hg. Three patients had Sievers type 0 BAV, 11 had type 1, and 2 were type 2. The Dacron-covered titanium rings had circular base geometry with 180° subcommissural posts and were implanted subannularly. Leaflets were reconstructed using plication/cleft closure, creating an effective height of ≥8 mm, even if modest gradients were induced. RESULTS: Mean pre-repair annular diameter was 28.6 ± 3.3 mm, and the average ring diameter was 22.3 ± 1.6 mm. All valves required leaflet plication/reconstruction; pericardium was avoided; and 7 patients had aortic replacement for aneurysms. No early mortalities or major complications occurred. Two patients required early prosthetic valve replacement for technical errors, and all were between 24-38 months' postoperative at follow-up. No late mortalities or valve-related complications occurred, and all patients reverted to New York Heart Association class I. Aortic insufficiency reduction was significant to grade 0.9 ± 0.5 at 2-years (P < .0001). Mean valve gradients were acceptable (13.3 ± 5.0 mm Hg at 2 years; overall P = .11) and tended to fall over time (P < .0001). CONCLUSIONS: Geometric ring annuloplasty was safe and effective for BAV repair. AI reduction was significant, valve gradients were satisfactory, and clinical outcomes were excellent. Geometric ring annuloplasty could simplify and standardize BAV repair.
OBJECTIVE: As bicuspid aortic valve (BAV) repair evolves, more effective annular reduction and stabilization could be advantageous. A geometric annuloplasty ring has been developed, and 2-year regulatory outcomes of a first-in-humans pilot trial are reported. METHODS: A prospective first-in-humans trial of BAV ring annuloplasty was completed in 16 patients. Patient age was 44.4 ± 11.3 (mean ± standard deviation) years, preoperative aortic insufficiency grade was 2.5 ± 1.0, New York Heart Association class 1.8 ± 0.4, and mean systolic gradient 13.4 ± 12.9 mm Hg. Three patients had Sievers type 0 BAV, 11 had type 1, and 2 were type 2. The Dacron-covered titanium rings had circular base geometry with 180° subcommissural posts and were implanted subannularly. Leaflets were reconstructed using plication/cleft closure, creating an effective height of ≥8 mm, even if modest gradients were induced. RESULTS: Mean pre-repair annular diameter was 28.6 ± 3.3 mm, and the average ring diameter was 22.3 ± 1.6 mm. All valves required leaflet plication/reconstruction; pericardium was avoided; and 7 patients had aortic replacement for aneurysms. No early mortalities or major complications occurred. Two patients required early prosthetic valve replacement for technical errors, and all were between 24-38 months' postoperative at follow-up. No late mortalities or valve-related complications occurred, and all patients reverted to New York Heart Association class I. Aortic insufficiency reduction was significant to grade 0.9 ± 0.5 at 2-years (P < .0001). Mean valve gradients were acceptable (13.3 ± 5.0 mm Hg at 2 years; overall P = .11) and tended to fall over time (P < .0001). CONCLUSIONS: Geometric ring annuloplasty was safe and effective for BAV repair. AI reduction was significant, valve gradients were satisfactory, and clinical outcomes were excellent. Geometric ring annuloplasty could simplify and standardize BAV repair.
Outcomes observed over 2 years after BAV repair using geometric ring annuloplasty.BAV repair supported by geometric ring annuloplasty is associated with excellent 2-year outcomes.Techniques now exist to repair most types of BAV disease with potentially improved long-term results as compared with prosthetic valve replacement.See Commentary on page 26.Compared with aortic valve replacement (AVR), aortic valve repair is associated with fewer valve-related complications, and better risk-adjusted long-term survival.3, 4, 5 The development of bicuspid aortic valve (BAV) repair has been especially important, since patients with BAV tend to be younger and are not good candidates for tissue AVR because of early prosthetic degeneration nor mechanical AVR because of life-long anticoagulation requirements. Outcomes with the Ross operation for BAV disease have been satisfactory, but this procedure exposes patients to the long-term complications of 2 operated valves: the pulmonary autograft and the right ventricular conduit.Most would agree that the development of contemporary BAV repair has been a major advance. Yet, problems continue to exist, such as reconstruction of BAVs with relatively equal-sized sinuses, the need to add pericardial tissue that may fail,, late annular dilatation in the absence of root stabilization, and the relative efficacy of annuloplasty techniques.12, 13, 14, 15 To address these issues, an internal geometric annuloplasty ring for use in BAV repair has been developed and tested in a prospective first-in-humans pilot trial. The purpose of this paper is to report the final 2-year regulatory outcomes of that trial as a preliminary to developing a pivotal clinical investigation.
Methods
The BAV ring was computer-machined from a solid block of titanium and covered with a thin layer of polyester to promote endothelialization. It had circular base geometry and two 180° subcommissural posts (Figure 1). The design of the ring was generated from multiple sources. First, the curvature, height, and geometry of the subcommissural posts were derived from computed tomographic angiographic studies of normal trileaflet aortic valves, and these aspects were taken directly from the design of the trileaflet ring. Thus, curvatures, post heights (equal to one radius of the valve base), and the 10° outward flare were the same as the trileaflet ring.
Figure 1
A bicuspid aortic valve annuloplasty ring with circular base geometry and two 180° subcommissural posts.
A bicuspid aortic valve annuloplasty ring with circular base geometry and two 180° subcommissural posts.The second design source consisted of pre- and postoperative computed tomographic angiograms in 10 patients undergoing successful BAV repair with subcommissural annuloplasty. Representative scans from these patients generally showed an elliptical annulus preoperatively, with the long axis of the ellipse in the sinus-to-sinus direction (Figure 2). No matter the preoperative annular configuration, the competent postrepair BAVs tended to assume circular base geometry with closer to 180° commissures. Therefore, the ring was designed with these features.
Figure 2
Computed tomography angiograms of representative BAV repair cases before and after repair. Patient 1 has a Sievers type 0 BAV with nearly equal sinuses and leaflets. Patients 2 to 4 have left-right fusion Sievers type 1 valves with varying degrees of annular enlargement. Patient 5 has left-right fusion with 3 equal sinuses. A 21-mm pink circle is superimposed, as well as red dots in the areas of 180° commissures. At the mid-valve level after successful repair, notice the valve assumes more of a circular base geometry, with symmetrical 180° commissures. These features were incorporated into the ring design.
Computed tomography angiograms of representative BAV repair cases before and after repair. Patient 1 has a Sievers type 0 BAV with nearly equal sinuses and leaflets. Patients 2 to 4 have left-right fusion Sievers type 1 valves with varying degrees of annular enlargement. Patient 5 has left-right fusion with 3 equal sinuses. A 21-mm pink circle is superimposed, as well as red dots in the areas of 180° commissures. At the mid-valve level after successful repair, notice the valve assumes more of a circular base geometry, with symmetrical 180° commissures. These features were incorporated into the ring design.The sample size calculation for this safety and efficacy pilot study was taken from an antecedent trial of trileaflet ring annuloplasty, in which the treatment effect was sufficient to define a statistical benefit with a 90% power and P value of <.05 with 16 patients. Consequently, 16 patients with bicuspid valves were recruited into the study with liberal selection criteria; the only exclusions were active infection and heavy calcification. All protocols were approved by the German Federal Regulatory body (BfArM) and local ethics committees (ClinicalTrials.gov: NCT02071849). Every patient was counseled extensively and provided signed informed consent.Baseline age was 44.4 ± 11.3 (mean ± standard deviation) years, aortic insufficiency (AI) grade was 2.5 ± 1.0, New York Heart Association class was 1.8 ± 0.4, and mean systolic gradient was 13.4 ± 12.9 mm Hg (Table 1). Three patients had Sievers type 0 valves, 11 had Sievers type 1 valves, and 2 had Sievers type 2. Thirteen patients had left-/right-coronary cusp fusion, 1 had right-/noncoronary cusp fusion, and 2 had both (unicuspid valves). Moderate-to-severe AI was present in 12 of the 16 cases, and 4 had mild AI associated with aortic aneurysms. Ascending aortic and/or root aneurysms were replaced in 7 of the 16 patients, using Dacron grafts 7 mm larger than the chosen ring. For bicuspid root aneurysms, remodeling Valsalva grafts were fashioned with 2 approximately equal sinus tongues (Table 1), except for asymmetric root aneurysms, where only the enlarged nonfused sinus was selectively replaced.
Table 1
Baseline and early postoperative data for 16 patients with BAV repair
Patient ID
Age, y
Sex
Annular diameter, mm
Ring size, mm
Leaflets fused—Sievers type
Leaflet procedure
Other procedure
Pre/early postoperative mean gradient, mm Hg
Preoperative AI grade0-4+
Early postoperative AI grade0-4+
Comments
No.
Patient no.
1
01-001
36
M
28
21
LR-1
P
ARR
21/28
3
0
2∗
01-002
59
M
29
21
LR-1
P-C
10/18
3
0/3
Leaflet tear - AVR
3
01-003
33
M
35
25
LR/RN-2
AP
35/27
4
0
4
01-004
29
M
28
21
LR-1
P-C
–/13
3
0/2
? Endocarditis treated medically
5
07-001
47
F
23
19
LR-1
P-C
8/25
1
0
6
01-005
56
F
26
21
LR-1
P
ARR
11/20
1
1
7
07-002
52
M
27
23
LR-1
P-U
6/13
3
0
8∗
01-007
29
M
28
23
LR/RN-2
AP
45/11
2
0/3
Leaflet tear - AVR
9
01-008
47
M
28
23
LR-0
P-AP
ARR
3/13
2
1
10
08-001
49
M
29
23
LR-1
P-C
AA-HA
–/22
4
0
11
07-003
41
M
28
23
LR-0
P
6/17
3
1
12
01-009
25
M
27
21
RN-0
P
AA
13/12
1
0
13
07-005
57
M
25
21
LR-1
P-U
AA
5/10
2
0
14
07-006
48
M
33
25
LR-1
P
ARR
4/6
1
0
15
01-010
43
M
35
23
LR-1
P
Pacer
–/22
3
0
16
07-007
38
M
29
23
LR-1
P-C-U
AF Abl
7/26
4
0
Mean
43.1
88%
28.6
22.3
3 AA
13/17
2.5
0.2/0.7
SD
10.6
M
3.3
1.6
4 ARR
13/7
1.0
0.6/1.1
Data represent the entire 16 patients having BAV repair. Forward slash symbols in gradient data represent preoperative versus early postoperative values. Dashes are missing data. Forward slashes in postoperative AI grades represent early versus later postoperative values, after leaflet tears in 2 and after medically treated endocarditis in 1. M, Male; L, left coronary leaflet; R, right coronary leaflet; AI, aortic insufficiency; P, leaflet plication; ARR, remodeling aortic root replacement; C, cleft closure; AVR, prosthetic aortic valve replacement; N, noncoronary leaflet; AP, autologous pericardial addition to leaflet; F, female; AA, ascending aortic replacement; HA, hemi-arch replacement; U, ultrasonic decalcification; AF Abl, atrial fibrillation ablation; SD, standard deviation.
The 2 patients undergoing early reoperation for technical errors.
Baseline and early postoperative data for 16 patients with BAV repairData represent the entire 16 patients having BAV repair. Forward slash symbols in gradient data represent preoperative versus early postoperative values. Dashes are missing data. Forward slashes in postoperative AI grades represent early versus later postoperative values, after leaflet tears in 2 and after medically treated endocarditis in 1. M, Male; L, left coronary leaflet; R, right coronary leaflet; AI, aortic insufficiency; P, leaflet plication; ARR, remodeling aortic root replacement; C, cleft closure; AVR, prosthetic aortic valve replacement; N, noncoronary leaflet; AP, autologous pericardial addition to leaflet; F, female; AA, ascending aortic replacement; HA, hemi-arch replacement; U, ultrasonic decalcification; AF Abl, atrial fibrillation ablation; SD, standard deviation.The 2 patients undergoing early reoperation for technical errors.Ring implantation technique has been described elsewhere, but to summarize, the required ring diameter was estimated by measurement of nonfused cusp free-edge length (L), using ball sizers in the sinus of Valsalva. Bicuspid ring size was based on the following formula: required ring diameter = L/1.8. In most patients, annular base geometry was usually elliptical, with the sinus-to-sinus diameter being the long axis of the ellipse (Figure 2). The 2 ring posts were sutured into the annuli of the subcommissural triangles (straddling the nonfused leaflet) with transannular horizontal mattress sutures of 4-0 PROLENE (Figure 3) supported with fine supra-annular Dacron pledgets (Video 1). The ring then was passed below the annulus, and 5 to 7 looping mattress sutures were placed deeply through the sinus aspect of both annuli, 2 mm deep to the leaflet–aortic junction. All annular sutures were tied tightly over fine supra-annular Dacron pledgets and fixed laterally to prevent leaflet contact (Figure 3). Burying the ring posts back into the subcommissural spaces and tight apposition of the ring bodies back under the annuli prevented contact between the leaflets and ring Dacron, which could cause leaflet abrasion and repair failure. Remodeling to a circular geometry moved the sinuses centrally and recruited leaflet tissue to the midline for improved coaptation. In addition, the fused annular segment usually was larger, and differential reduction in the fused sinus annulus to the size of the nonfused sinus provided advantages for fused leaflet mobility as a 50%-50% annular and commissural configuration.
Figure 3
Technique of type 1 BAV repair. A, With exposure achieved by 6 commissural and aortic sutures, a cleft in the fused leaflet (yellow arrow) is evident, as well as a redundant and prolapsing nonfused noncoronary leaflet. B, Horizontal mattress sutures bury the ring posts into the subcommissural spaces (red arrow). C, After placing annular sutures around both sinus areas, the sutures are tied over fine Dacron pledgets, and one needle is passed laterally (green arrow) and tied again to laterally fixate annular sutures away from leaflets. D, After repair, the linear closure of the fused leaflet cleft is evident (blue arrow) along with several plications on the non-fused leaflet. Importantly, the effective heights, geometric heights, and free-edge lengths of the 2 leaflets are equal. BAV, Bicuspid aortic valve.
A clinical video of all the techniques employed in this paper. Video available at: https://www.jtcvs.org/article/S2666-2507(20)30023-7/fulltext.Technique of type 1 BAV repair. A, With exposure achieved by 6 commissural and aortic sutures, a cleft in the fused leaflet (yellow arrow) is evident, as well as a redundant and prolapsing nonfused noncoronary leaflet. B, Horizontal mattress sutures bury the ring posts into the subcommissural spaces (red arrow). C, After placing annular sutures around both sinus areas, the sutures are tied over fine Dacron pledgets, and one needle is passed laterally (green arrow) and tied again to laterally fixate annular sutures away from leaflets. D, After repair, the linear closure of the fused leaflet cleft is evident (blue arrow) along with several plications on the non-fused leaflet. Importantly, the effective heights, geometric heights, and free-edge lengths of the 2 leaflets are equal. BAV, Bicuspid aortic valve.Bicuspid leaflet repairs usually were performed using Schäfers' techniques (Video 1). After annuloplasty, type 0 valves underwent plication of both leaflets to correct prolapse. For type 1 valves, the nonfused leaflet was plicated to an effected height of ≥8 mm and used as a reference (Figure 3). When present, the fused leaflet cleft was closed linearly until similar free-edge lengths, geometric heights, and effective heights were achieved for both cusps. For type 2 valves, fused right/noncoronary commissurotomy was performed first, and then the valve was repaired as a type 1. A trade-off existed between effective height and magnitude of gradient. If any question existed, an effective height of ≥8 mm Hg was created, and a greater gradient was tolerated. Glutaraldehyde-fixed autologous pericardium generally was avoided but was needed to augment leaflet defects in 3 patients. Three additional patients required debridement of leaflet calcification using the Cavitron Ultrasonic Aspirator device (Video 1). Only aspirin anticoagulation was administered postoperatively.This pilot first-in-humans study was predefined for a 2-year follow-up of all clinical outcomes and echocardiographic characteristics. The primary endpoint of the trial was all-cause death, and secondary endpoints were AI reduction and mean valve gradients. An echo core laboratory provided all echo readings (MedStar Research Institute, Washington DC), using standard echo criteria and a 0-4+ AI grading scale. Left ventricular dimensions and mass were derived from echocardiographic data using standard techniques. Three patients had missing preoperative valve gradients, and these were imputed to the mean for the rest. Otherwise, missing data were imputed using the last value forward technique. Changes over time in clinical and echocardiographic variables from preoperative to 2 years' postrepair were evaluated with Friedman's nonparametric analysis of variance or 2-tailed paired t tests where appropriate. Analyses and graphics were performed with Prism 7.0 (GraphPad Software, Inc, San Diego, Calif), and a P value of .05 was considered significant.
Results
In the 16 patients with BAV, mean pre-repair annular diameter was 28.6 ± 3.3 mm, and average ring diameter was 22.3 ± 1.6 mm, indicating significant annular dilatation in all cases (Table 1). All valves required leaflet plication and/or cleft closure. Seven had ascending aortic and/or remodeling root replacement for aneurysms. No early or late mortalities or major complications occurred. One patient with pre-existing heart block and syncope required a pacemaker postoperatively. Two patients experienced early leaflet lacerations due to long annular suture tails, with recurrent AI as a technical complication (Table 1). Both were reoperated successfully for prosthetic AVR. Subsequently, a “lateral suture fixation” technique was applied (Figure 3) to avoid this complication, and no more suture-induced leaflet injuries occurred (now to >700 subsequent ring implants).At follow-up, all patients were between 24 and 38 months postoperation, and analyses were performed with and without the 2 leaflet laceration patients included. When included, all their post-AVR missing data were imputed as the worst values observed after repair failure, penalizing this complication. Analyses are presented in Figure 4, excluding the 2 early technical failures because their repairs were lost for follow-up assessment; however, overall results were similar, and P values were no different if they were included. No thromboembolism, strokes, or bleeding occurred. No late mortalities or valve-related complications were observed, and all patients reverted to New York Heart Association class I long term. Serial echocardiograms showed prolonged and stable AI reduction to grade 0.9 ± 0.5 at 2 years (P < .0001). One patient (no. 4) experienced a febrile illness late postoperatively, with leukocytosis, new appearance of central grade 2 AI, but negative cultures. He was treated with prolonged antibiotics for presumed culture-negative endocarditis and maintained a stable grade 2 leak long term without reoperation.
Figure 4
Time course of clinical and echocardiographic variables before and after BAV repair. In this analysis, the early repair failures were omitted, because their repairs were not available for assessment long term. The results, however, changed insignificantly if the 2 repair failures were included. A, Survival was excellent. B, All patients became asymptomatic. C, Average AI grade fell below 1+ and remained low. D, Mean valve gradients increased slightly immediately after annuloplasty, but overall, remained statistically unchanged by analysis of variance (P = .11). BAV, Bicuspid aortic valve; AI, aortic insufficiency; NYHA, New York Heart Association; SEM, standard error of the mean; ANOVA, analysis of variance; Scrn, screening data; Disch, discharge data.
Time course of clinical and echocardiographic variables before and after BAV repair. In this analysis, the early repair failures were omitted, because their repairs were not available for assessment long term. The results, however, changed insignificantly if the 2 repair failures were included. A, Survival was excellent. B, All patients became asymptomatic. C, Average AI grade fell below 1+ and remained low. D, Mean valve gradients increased slightly immediately after annuloplasty, but overall, remained statistically unchanged by analysis of variance (P = .11). BAV, Bicuspid aortic valve; AI, aortic insufficiency; NYHA, New York Heart Association; SEM, standard error of the mean; ANOVA, analysis of variance; Scrn, screening data; Disch, discharge data.Generally, mean valve gradients were acceptable (13.3 ± 5.0 mm Hg at 2 years; overall P = .11 from preoperative). In one third of patients (those with more complex anatomy such as type 2 valves with dysplastic leaflets or patients with decalcified leaflets), discharge mean gradients were 20 to 30 mm Hg echocardiographically (Figure 5). However, gradients fell significantly over the next 2 years (P < .0001), so that the highest gradients approximated 20 mm Hg at late follow-up, presumably as native living leaflets adapted. All patients with significant preoperative AI experienced reductions in left ventricular dimensions and mass after repair (Table 2), even with mean gradients in the 20- to 30-mm Hg range. Patient 1 was an example in whom scarred reoperative leaflets were associated with gradients in the 20s throughout; yet, he became asymptomatic and left ventricular hypertrophy/dilatation recovered. This operation can be viewed at: https://www.ctsnet.org/article/bicuspid-ring-annuloplasty-and-leaflet-reconstruction-after-failed-bicuspid-valve-repair-and. Of all echocardiographic dimensions, calculated left ventricular mass fell the most after repair, decreasing on average by a third after elimination of volume overload (Table 2; P < .002).
Figure 5
Mean valve gradients for individual patients over time after repair. One third of patients had mean gradients in excess of 20 mm Hg at discharge, but they all improved to approximately 20 mm Hg or less at 2-years (P < .001 by analysis of variance of postrepair data only). BAV, Bicuspid aortic valve; ANOVA, analysis of variance; Preop, preoperative; Disch, discharge data.
Table 2
Baseline and 2-year hemodynamic data for 14 patients undergoing BAV repair
Patient ID
Leaflets fused—Sievers type
Preoperative annular diameter, mm
Ring size, mm
Pre/2YAI grade0-4+
Pre/2YLVEDD, cm
Pre/2YLVESD, cm
Pre/2-Y mean valvegradient,mm Hg
Pre/2-YLV mass, g
No.
Patient no.
1
01-001
LR-1
28
21
3/0
5.9/4.9
4.0/3.4
21/21
378/197
3
01-003
LR/RN-2
30
25
4/1
4.7/4.5
2.9/2.9
35/16
309/210
4
01-004
LR-1
28
21
3/1
5.9/5.7
4.1/3.2
–/12
305/273
5
07-001
LR-1
23
19
1/1
4.5/4.5
3.0/2.7
8/16
186/137
6
01-005
LR-1
26
21
1/1
4.4/4.6
2.9/2.3
11/10
139/135
7
07-002
LR-1
27
23
3/0
5.6/5.0
4.6/3.5
6/8
311/186
9
01-008
LR-0
28
23
4/1
6.0/5.3
3.8/3.3
3/10
311/209
10
08-001
LR-1
29
23
4/1
7.3/4.5
–/3.1
–/9
463/198
11
07-003
LR-0
28
23
3/1
5.8/4.8
4.7/2.6
6/16
297/187
12
01-009
RN-0
27
21
1/1
4.2/4.5
2.9/3.0
13/10
157/125
13
07-005
LR-1
25
21
2/1
5.6/6.3
4.0/4.3
5/7
134/163
14
07-006
LR-1
33
25
1/1
5.5/5.2
3.4/3.5
4/9
320/278
15
01-010
LR-1
35
23
3/1
–/4.8
–/3.1
–/13
–/171
16
07-007
LR-1
29
23
4/2
–
–
7/22
–
Mean
28.3
22.3
2.6/0.9
5.5/5.0
3.7/3.1
11/13
276/190
SD
3.0
1.7
1.2/0.5
0.9/0.5
0.7/0.5
10/5
101/47
Data represent the 14 patients achieving the full 2-year follow-up after BAV repair. The forward slash symbols denote preoperative versus 2-year data, and dashes (–) represent missing data that were imputed as described in the Methods. 2Y, 2 Years' postoperative; AI, aortic insufficiency; LV, left ventricular; EDD, end-diastolic diameter; ESD, end-systolic diameter; L, left coronary leaflet; R, right coronary leaflet; N, noncoronary leaflet; SD, standard deviation.
Mean valve gradients for individual patients over time after repair. One third of patients had mean gradients in excess of 20 mm Hg at discharge, but they all improved to approximately 20 mm Hg or less at 2-years (P < .001 by analysis of variance of postrepair data only). BAV, Bicuspid aortic valve; ANOVA, analysis of variance; Preop, preoperative; Disch, discharge data.Baseline and 2-year hemodynamic data for 14 patients undergoing BAV repairData represent the 14 patients achieving the full 2-year follow-up after BAV repair. The forward slash symbols denote preoperative versus 2-year data, and dashes (–) represent missing data that were imputed as described in the Methods. 2Y, 2 Years' postoperative; AI, aortic insufficiency; LV, left ventricular; EDD, end-diastolic diameter; ESD, end-systolic diameter; L, left coronary leaflet; R, right coronary leaflet; N, noncoronary leaflet; SD, standard deviation.
Discussion
The development of a reproducible and stable technique of BAV repair has been a major advance in cardiac surgery.,25, 26, 27 The first patients receiving Schäfers' techniques for leaflet reconstruction are now 20 years' postrepair, and most are doing well. Compared with prosthetic valve replacement, valve-related complications and mortality are improved,1, 2, 3, 4, 5 and the younger patients with BAV defects can live fully active lives without anticoagulation or other sequalae of “prosthetic valve disease.” Yet, imperfections in surgical management remain. Patients with 3 equal-sized sinuses (Figure 2, patient 5) experience greater repair failure/reoperation rates with standard BAV repair, since a 2-leaflet reconstruction can obstruct the valve, among other problems. Failure to stabilize an enlarged annulus, or annuloplasty by Cabrol's technique, both have been associated with greater failure rates.,, The addition of autologous pericardium as a cusp extension (or as hemi-leaflet replacement in Sievers type 2 [unicuspid] valves) has predicted repair failure due to pericardial degeneration. Moreover, other available leaflet tissue substitutes also have performed poorly. Thus, a repair technique that reduces annular diameter more effectively and better recruits native leaflet tissue to midline coaptation (minimizing the need for pericardium) would be useful.In the current study, the major geometric remodeling provided by the BAV annuloplasty ring seemed to accomplish these goals. With ring size objectively based on the free-edge length of the nonfused cusp, average annular diameter was reduced to 22 mm, which recruited more leaflet to the center of the valve, allowed better vertical leaflet coaptation, and minimized the need for pericardial augmentation. Most repairs were accomplished using native leaflet alone, with the advantage of maintaining living tissue, and potentially allowing adaptation and remodeling of native leaflets to the physiologic milieu (Figure 5). Even in patient 1 (a reoperative repair with scarred retracted leaflets; https://www.ctsnet.org/article/bicuspid-ring-annuloplasty-and-leaflet-reconstruction-after-failed-bicuspid-valve-repair-and), enough leaflet was recruited to ensure a long-term competent valve. In addition, regional annular remodeling also was effective, so that a large fused annulus could be reduced to the same size as the nonfused annulus, routinely achieving 180° commissural geometry in a simpler way than plicating the sinus. This annuloplasty approach solves the problem of 3 equal sinuses, and routine 180° root geometry promotes better flow characteristics. This feature could be especially important in repair of unicuspid valves and intermediate-type bicuspid valves, both of which frequently have 3 equal sinuses.In this regulated clinical trial, the endpoints were met, with a zero mortality, low residual AI rates, and acceptable mean valve gradients. The data support the algorithm to first achieve adequate effective height ≥8 mm Hg, even if mean gradients are modestly elevated to 20 to 30 mm Hg. All patients, even those with initially elevated gradients, remained asymptomatic, recovered ventricular function, and, with time, mean valve gradients fell uniformly (P < .0001). The one-third decrease in left ventricular mass after BAV repair was especially impressive. The technique was highly reproducible, applicable to all BAV configurations (Video 1), and could assist in standardizing BAV repair. These satisfactory pilot data justify performance of a pivotal patient study, and an analysis with a larger sample size of 150 patients and a longer follow-up to 10 years currently is being developed.
Conflict of Interest Statement
Drs Rankin, Mazzitelli, Fischlein, Choi, and Wei are consultants for BSE and report personal fees from proctoring clinical cases in Europe and the United States. Dr Rankin reports minor equity and royalty interests in BSE. Neither the company nor the authors had any contact with or influence on the independent regulatory data, and the analysis was performed by an independent statistician. All other authors have nothing to disclose with regard to commercial support.
Authors: William A Zoghbi; Maurice Enriquez-Sarano; Elyse Foster; Paul A Grayburn; Carol D Kraft; Robert A Levine; Petros Nihoyannopoulos; Catherine M Otto; Miguel A Quinones; Harry Rakowski; William J Stewart; Alan Waggoner; Neil J Weissman Journal: J Am Soc Echocardiogr Date: 2003-07 Impact factor: 5.251
Authors: Hunaid A Vohra; Robert N Whistance; Laurent De Kerchove; Prakash Punjabi; Gebrine El Khoury Journal: Eur J Cardiothorac Surg Date: 2013-01-04 Impact factor: 4.191
Authors: Prashanth Vallabhajosyula; Caroline Komlo; Wilson Y Szeto; Tyler J Wallen; Nimesh Desai; Joseph E Bavaria Journal: Ann Thorac Surg Date: 2014-01-10 Impact factor: 4.330
Authors: Domenico Mazzitelli; Christian Nöbauer; J Scott Rankin; Catalin C Badiu; Magdalena Dorfmeister; Philip S Crooke; Anke Wagner; Christian Schreiber; Rüdiger Lange Journal: Eur J Cardiothorac Surg Date: 2013-08-19 Impact factor: 4.191
Authors: Vikas Sharma; Rakesh M Suri; Joseph A Dearani; Harold M Burkhart; Soon J Park; Lyle D Joyce; Zhuo Li; Hartzell V Schaff Journal: J Thorac Cardiovasc Surg Date: 2013-09-29 Impact factor: 5.209
Authors: Hunaid A Vohra; Robert N Whistance; Laurent de Kerchove; David Glineur; Philipe Noirhomme; Gebrine El Khoury Journal: Ann Cardiothorac Surg Date: 2013-01
Authors: Marc W Gerdisch; T Brett Reece; Dominic Emerson; Richard S Downey; Geoffrey B Blossom; Arun Singhal; Joshua N Baker; Theodor J M Fischlein; Vinay Badhwar Journal: JTCVS Tech Date: 2022-06-09