| Literature DB >> 34314528 |
Moiz Ahmad1, Dominique Tan2, Sujay Marisetty3.
Abstract
PURPOSE: The global noise (GN) algorithm has been previously introduced as a method for automatic noise measurement in clinical CT images. The accuracy of the GN algorithm has been assessed in abdomen CT examinations, but not in any other body part until now. This work assesses the GN algorithm accuracy in automatic noise measurement in head CT examinations.Entities:
Keywords: automation; computed tomography; quality control
Mesh:
Year: 2021 PMID: 34314528 PMCID: PMC9291315 DOI: 10.1002/mp.15133
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.506
Summary of imaging parameters across scan protocols
| Protocol | N | Scan Mode | Tube current (mA) | CTDIvol (mGy) | Recon. filter | FOV (cm) |
|---|---|---|---|---|---|---|
| GE Protocol 1 | 42 | axial | 300 | 56.8 | Standard | 20.0–25.4 |
| GE Protocol 2 | 7 | helical | 550 | 56.8 | Standard | 20.6–21.3 |
| Siemens Protocol 1 | 30 | (unverified) | 149 or 151 | 43.7 | H40 | 25.0 |
| Siemens Protocol 2 | 20 | (unverified) | 202–211 | 43.7 | H40 | 25.0 |
| Siemens Low dose Protocol 1 | 30 | (unverified) | 37 (approx.) | 10.9 (approx.) | H40 | 25.0 |
| Siemens Low dose Protocol 2 | 20 | (unverified) | 51 (approx.) | 10.9 (approx.) | H40 | H40 |
FIGURE 1Example manual noise measurement using ROI placement in cerebral white matter. The slice location containing all four lateral ventricle horns was used in the manual measurements. Four quadrant ROI locations were used
Summary statistics of reference noise measurements
| Set of examinations | Mean noise (95% CI) | Minimum | Maximum | Standard deviation | Spatial variation |
|---|---|---|---|---|---|
| All | 5.1 HU (4.5–5.8 HU) | 2.9 HU | 10.2 HU | 1.7 HU | 0.5 HU |
| Full dose | 4.0 HU (3.7–4.7 HU) | 2.9 HU | 5.6 HU | 0.4 HU | 0.4 HU |
| Low dose | 7.3 HU (6.6–7.9 HU) | 6.0 HU | 10.2 HU | 0.9 HU | 0.7 HU |
Summary of results for GN algorithm error and bias
| Set of examinations | RMS error | Percent RMS error | Percent bias |
|---|---|---|---|
| All | 0.34 HU (0.24–0.46 HU) | 6.6% (4.7%–9.0%) | +3.9% (+2.6%–+5.1%) |
| Full dose | 0.25 HU (0.16–0.35 HU) | 6.0% (3.9%–8.8%) | +3.1% (+1.7%–+4.6%) |
| Low dose | 0.47 HU (0.25–0.79 HU) | 6.5% (3.5%–10.8%) | +4.6% (+2.8%–+6.4%) |
95% confidence intervals indicated in parentheses.
FIGURE 2Differences between GN and reference noise values shown in a Bland–Altman plot. Data points to the right of the vertical dashed line are from simulated low‐dose examinations from Siemens Protocols 1 and 2
FIGURE 3Boxplot of automatic global noise measurements. * p < 0.05
p test statistics for pairwise differences of mean GN between protocols
| Scan Protocol | GE1 | GE2 | S1 | S2 | SLD1 | SLD2 |
|---|---|---|---|---|---|---|
| GE Protocol 1 (GE1) | — | 0.95 |
| 0.45 |
|
|
| GE Protocol 2 (GE2) | — |
| 0.64 |
|
| |
| Siemens Protocol 1 (S1) | — |
|
|
| ||
| Siemens Protocol 2 (S2) | — |
|
| |||
| Siemens Low‐dose Protocol 1 (SLD1) | — | 0.24 | ||||
| Siemens Low‐dose Protocol 2 (SLD2) | — |
p < 0.05
p < 0.01
p < 0.001
p < 0.0001.
FIGURE 4Scatterplot of GN versus head size