Rimjhim Kanaujia1, Arnab Ghosh2, Ritin Mohindra3, Vidhi Singla3, Kapil Goyal2, Rajendra Gudisa2, Vikrant Sharma2, Lalit Mohan2, Navpreet Kaur2, Gursimran Kaur Mohi2, Ishani Bora2, Radha Kanta Ratho2, Roop Kishor Soni3, Ashish Bhalla3, Mini P Singh4.
Abstract
PURPOSE: To determine the utility, diagnostic accuracy, sensitivity, specificity, and negative predictive value of the laboratory based Covid-19 antigen detection test (Coris Bio- Concept, Gembloux, Belgium) for the diagnosis of SARS-CoV-2 in a tertiary care hospital among symptomatic and asymptomatic patients.Entities:
Keywords: Point-of-care; Rapid antigen detection; SARS-CoV-2
Mesh:
Substances:
Year: 2021 PMID: 34294504 PMCID: PMC8289730 DOI: 10.1016/j.ijmmb.2021.07.003
Source DB: PubMed Journal: Indian J Med Microbiol ISSN: 0255-0857 Impact factor: 0.985
Demographic features of 484 patients for which both RT-PCR and RAT was done.
| COVID-19 Result by Real-Time PCR | COVID-19 result by Ag Respi-Strip CORIS | |||
|---|---|---|---|---|
| Positive (n) | Negative(n) | |||
| Male | Positive | 84 | 28 | |
| Negative | 1 | 148 | ||
| Female | Positive (n) | 52 | 25 | |
| Negative (n) | 1 | 145 | ||
| Mean | 35.5 | |||
| Median | 32 | |||
Clinical Characteristics of the 825 Patients in relation to COVID-19 result by Ag Respi-Strip CORIS.
| COVID-19 result by Ag Respi-Strip CORIS | ||||
|---|---|---|---|---|
| Positive | Negative | p-value | ||
| Present | 115 | 518 | ||
| Absent | 23 | 169 | ||
| Present | 93 | 268 | ||
| Absent | 45 | 419 | ||
| Present | 55 | 198 | ||
| Absent | 83 | 489 | ||
| Present | 3 | 17 | .83 | |
| Absent | 135 | 670 | ||
| Present | 53 | 256 | .80 | |
| Absent | 85 | 431 | ||
| Present | 1 | 15 | .25 | |
| Absent | 137 | 672 | ||
| Present | 1 | 3 | .65 | |
| Absent | 137 | 684 | ||
| Present | 9 | 18 | ||
| Absent | 129 | 669 | ||
| Present | 2 | 2 | .07 | |
| Absent | 136 | 685 | ||
| Present | 7 | 25 | .42 | |
| Absent | 131 | 662 | ||
| Present | 29 | 49 | ||
| Absent | 109 | 638 | ||
| Present | 0 | 56 | NA | |
| Absent | 138 | 631 | ||
| Present | 0 | 1 | NA | |
| Absent | 139 | 685 | ||
Sensitivity, specificity, positive, and negative predictive values of COVID-19 Ag respi-strip CORIS for detection of SARS-CoV-2 in nasopharyngeal samples.
| COVID-19 Ag Respi-Strip CORIS | Positive Samples on RT-PCR (N = 189) | Negative Samples on RT-PCR (N = 295) | Total | Positive | Negative | Sensitivity | Specificity | Accuracy |
|---|---|---|---|---|---|---|---|---|
| Positive | 136 | 2 | 138 | 98.55% | 71.96% | 88.64% | ||
| Negative | 53 | 293 | 346 | 84.68% | 99.32% |
Fig. 1Median Ct value of antigen positive and negative samples. (19.6 vs 25.6) (P < 0.0001 (95%CI:4.8503 to 7.1497)).
Fig. 2Receiver operating characteristic (ROC) curve analysis. The rapid antigen test achieved an area under the ROC curve (AUC) value of 0.835.