Influence of lens position as detected by an anterior segment analysis system on postoperative refraction in cataract surgery.

Jia-Ju Zhang1, Jian-Qing Li1, Chen Li1, Yi-Hong Cao1, Pei-Rong Lu1.   

Abstract

AIM: To predict postoperative intraocular lens (IOL) position using the Sirius anterior segment analysis system and investigate the effect of lens position and IOL type on postoperative refraction.
METHODS: A total of 97 patients (102 eyes) were enrolled in the final analysis. An anterior segment biometry measurement was performed preoperatively with Sirius and Lenstar. The results of predicted lens position (PLP) and IOL power were automatically calculated by the software used by the instruments. Effective lens position (ELP) was measured manually using Sirius 3mo postoperatively. Pearson's correlation analysis and linear regression analysis were used to determine the correlation of lens position to other parameters.
RESULTS: PLP and ELP were positively correlated to axial length (AL; r=0.42, P<0.0001 and r=0.49, P<0.0001, respectively). There was a weak correlation between the peLP (ELP-PLP) and the prediction error of spherical refraction (peSR; r=0.34, P<0.0001). The peLP of Softec HD IOL differed statistically from those of both the TECNIS ZCB00 and Sensor AR40E IOLs. Multiple linear regression was used to obtain the prediction formula: ELP=0.66+0.63×[aqueous depth (AQD)+0.6LT] (r=0.61, P<0.0001), and a new variable (AQD+0.6 LT) was found to have the strongest correlation with ELP.
CONCLUSION: The Sirius anterior segment analysis system is helpful to predict ELP, which reduces postoperative refraction error. International Journal of Ophthalmology Press.

Entities:  

Keywords:  anterior segment analysis system; intraocular lens; lens position; postoperative refraction

Year:  2021        PMID: 34282384      PMCID: PMC8243177          DOI: 10.18240/ijo.2021.07.07

Source DB:  PubMed          Journal:  Int J Ophthalmol        ISSN: 2222-3959            Impact factor:   1.779


  37 in total

1.  Clinical results using the Holladay 2 intraocular lens power formula.

Authors:  K J Hoffer
Journal:  J Cataract Refract Surg       Date:  2000-08       Impact factor: 3.351

Review 2.  Optical coherence biometry.

Authors:  Wolfgang Haigis
Journal:  Dev Ophthalmol       Date:  2002

3.  Prediction of the effective postoperative (intraocular lens) anterior chamber depth.

Authors:  Thomas Olsen
Journal:  J Cataract Refract Surg       Date:  2006-03       Impact factor: 3.351

Review 4.  Calculation of intraocular lens power: a review.

Authors:  Thomas Olsen
Journal:  Acta Ophthalmol Scand       Date:  2007-04-02

5.  Predicting the postoperative intraocular lens position using continuous intraoperative optical coherence tomography measurements.

Authors:  Nino Hirnschall; Sahand Amir-Asgari; Sophie Maedel; Oliver Findl
Journal:  Invest Ophthalmol Vis Sci       Date:  2013-08-05       Impact factor: 4.799

6.  Using continuous intraoperative optical coherence tomography measurements of the aphakic eye for intraocular lens power calculation.

Authors:  Nino Hirnschall; Sverker Norrby; Maria Weber; Sophie Maedel; Sahand Amir-Asgari; Oliver Findl
Journal:  Br J Ophthalmol       Date:  2014-02-11       Impact factor: 4.638

7.  Prediction of Postoperative Intraocular Lens Position with Angle-to-Angle Depth Using Anterior Segment Optical Coherence Tomography.

Authors:  So Goto; Naoyuki Maeda; Shizuka Koh; Kazuhiko Ohnuma; Kenichi Hayashi; Ikko Iehisa; Toru Noda; Kohji Nishida
Journal:  Ophthalmology       Date:  2016-10-18       Impact factor: 12.079

8.  Effects of intraocular lens position errors on postoperative refractive error.

Authors:  P Erickson
Journal:  J Cataract Refract Surg       Date:  1990-05       Impact factor: 3.351

9.  Comparison of two biometers: A swept-source optical coherence tomography and an optical low-coherence reflectometry biometer.

Authors:  Hussam El Chehab; Emilie Agard; Corinne Dot
Journal:  Eur J Ophthalmol       Date:  2018-10-07       Impact factor: 2.597

10.  Comparison of agreement and efficiency of a swept source-optical coherence tomography device and an optical low-coherence reflectometry device for biometry measurements during cataract evaluation.

Authors:  Samuel F Passi; Atalie C Thompson; Preeya K Gupta
Journal:  Clin Ophthalmol       Date:  2018-11-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.