Comparison of two biometers: A swept-source optical coherence tomography and an optical low-coherence reflectometry biometer.

Hussam El Chehab1, Emilie Agard1,2, Corinne Dot1,2.   

Abstract

PURPOSE: The technological improvements in optical biometers have made cataract surgery evolve to the level of refractive surgery. This study assessed the correlation and agreement between a biometer using swept-source optical coherence tomography technology (IOLMaster 700; Carl Zeiss Meditec, Jena, Germany) and a biometer using optical low-coherence reflectometry (Lenstar; Haag Streit, Koeniz, Switzerland). PROCEDURES: Optical biometric measurements were taken with the IOLMaster 700 and the Lenstar on 129 eyes of 129 patients referred for cataract surgery. We compared biometric data (axial length, mean keratometry, central pachymetry, anterior chamber depth, lens thickness, and horizontal white-to-white measurement) and the emmetropic intraocular lens power calculated with SRK/T and Haigis formulas. Correlation and agreement for these parameters were evaluated.
RESULTS: Only anterior chamber depth measurements (3.05 ± 0.07 vs 3.07 ± 0.07, p < 0.001) and the white-to-white measurement (11.97 ± 0.07 vs 12.06 ± 0.07, p < 0.001) differed significantly between the machines. Correlations and agreements were excellent, with perfect agreement for axial length measurement (r = 1, intraclass correlation coefficient = 1). Emmetropic intraocular lens power did not differ between devices with SRK/T formula (20.94 ± 0.51 vs 20.92 ± 0.50, p = 0.51). With Haigis formula, there was a statistically significant difference (21.04 ± 0.52 vs 20.84 ± 0.52, p < 0.001).
CONCLUSION: This study suggests that there is a very high correlation and agreement for the biometric measurements and calculation of intraocular lens implant power with the SRK/T formula.

Entities:  

Keywords:  Biometry; IOLMaster 700; Lenstar; cataract surgery

Mesh:

Year:  2018        PMID: 30295063     DOI: 10.1177/1120672118802918

Source DB:  PubMed          Journal:  Eur J Ophthalmol        ISSN: 1120-6721            Impact factor:   2.597


  2 in total

1.  Influence of lens position as detected by an anterior segment analysis system on postoperative refraction in cataract surgery.

Authors:  Jia-Ju Zhang; Jian-Qing Li; Chen Li; Yi-Hong Cao; Pei-Rong Lu
Journal:  Int J Ophthalmol       Date:  2021-07-18       Impact factor: 1.779

2.  Comparison of ocular parameters of two biometric measurement devices in highly myopic eyes.

Authors:  Xiao-Xiao Guo; Ran You; Shan-Shan Li; Xiu-Fen Yang; Lu Zhao; Fan Zhang; Yan-Ling Wang; Xi Chen
Journal:  Int J Ophthalmol       Date:  2019-10-18       Impact factor: 1.779

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.