Evaluation of the Ohmeda 3700 pulse oximeter.

A R Warley1, J H Mitchell, J R Stradling.   

Abstract

Arterial oxygen saturation values (Sao2) from 60% to 98% were measured by the Ohmeda 3700 pulse oximeter with the three types of probe available and compared with values of oxygen saturation estimated from direct arterial sampling (arterial oxygen and carbon dioxide tensions and pH) on 65 occasions. The response time of the oximeter was measured after a sudden rise in inspired oxygen concentration. Artefact rejection was assessed by arterial compression proximal to the probe site, and by simultaneous recordings of overnight Sao2 on opposite hands. The ability to recreate patterns of oscillating Sao2 from the data stored in the oximeter was also investigated. With the best probe system the oximeter measured Sao2, relative to arterial values estimated from Pao2, with a mean (SD) difference of -0.4% (1.8%). The response time was comparable with those of previous oximeters. It was not possible to generate artefactual dips in excess of 2% Sao2, and the dual overnight recordings rarely showed even small dips on one tracing alone. The stored data can recreate oscillating Sao2 signals with wavelengths down to about 35 seconds, but not below. The Ohmeda 3700 pulse oximeter appears to be suitable for unattended overnight recordings of Sao2.

Entities:  

Mesh:

Substances:

Year:  1987        PMID: 3424271      PMCID: PMC461017          DOI: 10.1136/thx.42.11.892

Source DB:  PubMed          Journal:  Thorax        ISSN: 0040-6376            Impact factor:   9.139


  6 in total

1.  Accuracy sensitivity to carboxyhemoglobin, and speed of response of the Hewlett-Packard 47201A ear oximeter.

Authors:  N J Douglas; H M Brash; P K Wraith; P M Calverley; R J Leggett; L McElderry; D C Flenley
Journal:  Am Rev Respir Dis       Date:  1979-02

2.  A clinical method for assessing the ventilatory response to hypoxia.

Authors:  A S Rebuck; E J Campbell
Journal:  Am Rev Respir Dis       Date:  1974-03

3.  Range of accuracy of two wavelength oximetry.

Authors:  K R Chapman; F L Liu; R M Watson; A S Rebuck
Journal:  Chest       Date:  1986-04       Impact factor: 9.410

4.  Taste theories and the chemical stimulation of single papillae.

Authors:  G Von Békésy
Journal:  J Appl Physiol       Date:  1966-01       Impact factor: 3.531

5.  The accuracy and response characteristics of a simplified ear oximeter.

Authors:  K R Chapman; A D'Urzo; A S Rebuck
Journal:  Chest       Date:  1983-06       Impact factor: 9.410

6.  Evaluation of Biox IIA ear oximeter.

Authors:  P M Tweeddale; N J Douglas
Journal:  Thorax       Date:  1985-11       Impact factor: 9.139

  6 in total
  20 in total

1.  Predictors and prevalence of obstructive sleep apnoea and snoring in 1001 middle aged men.

Authors:  J R Stradling; J H Crosby
Journal:  Thorax       Date:  1991-02       Impact factor: 9.139

Review 2.  Pulse oximetry.

Authors:  J F Kelleher
Journal:  J Clin Monit       Date:  1989-01

3.  Ventricular arrhythmia, Cheyne-Stokes respiration, and death: observations from patients with defibrillators.

Authors:  A D Staniforth; S C Sporton; M J Early; J A Wedzicha; A W Nathan; R J Schilling
Journal:  Heart       Date:  2005-04-06       Impact factor: 5.994

4.  Relation between systemic hypertension and sleep hypoxaemia or snoring: analysis in 748 men drawn from general practice.

Authors:  J R Stradling; J H Crosby
Journal:  BMJ       Date:  1990-01-13

5.  Oxygen desaturation and breathlessness during corridor walking in chronic obstructive pulmonary disease: effect of oxitropium bromide.

Authors:  D P Spence; J G Hay; J Carter; M G Pearson; P M Calverley
Journal:  Thorax       Date:  1993-11       Impact factor: 9.139

6.  Nocturnal oxygen desaturation and spirometric parameters in adults with cystic fibrosis.

Authors:  M N Pond; S P Conway
Journal:  Thorax       Date:  1995-05       Impact factor: 9.139

7.  Neck circumference and other clinical features in the diagnosis of the obstructive sleep apnoea syndrome.

Authors:  R J Davies; N J Ali; J R Stradling
Journal:  Thorax       Date:  1992-02       Impact factor: 9.139

8.  Non-invasive beat to beat arterial blood pressure during non-REM sleep in obstructive sleep apnoea and snoring.

Authors:  R J Davies; J Crosby; K Vardi-Visy; M Clarke; J R Stradling
Journal:  Thorax       Date:  1994-04       Impact factor: 9.139

9.  Value of nocturnal oxygen saturation as a screening test for sleep apnoea.

Authors:  B G Cooper; D Veale; C J Griffiths; G J Gibson
Journal:  Thorax       Date:  1991-08       Impact factor: 9.139

10.  Changes in nocturnal oximetry after treatment of exacerbations in cystic fibrosis.

Authors:  M B Allen; A F Mellon; E J Simmonds; R L Page; J M Littlewood
Journal:  Arch Dis Child       Date:  1993-08       Impact factor: 3.791

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.