Biomechanical analysis of anterior and posterior instrumentation systems after corpectomy. A calf-spine model.

K R Gurr1, P C McAfee, C M Shih.   

Abstract

UNLABELLED: To simulate the spinal instability that is found clinically after anterior corpectomy for the treatment of a fracture or a neoplasm, twelve fresh calf-spine segments, each containing five motion segments, were destabilized using a complete anterior corpectomy at the third lumbar level and anterior discectomies at the second and third and the third and fourth lumbar levels. Mechanical non-destructive cyclical testing in axial compression, rotation, and flexion was performed on each spinal segment after stabilization was accomplished. The three anterior-stabilization constructs that were compared were: (1) iliac strut grafting, (2) polymethylmethacrylate and anterior Harrington-rod instrumentation (the technique of Siegal and Siegal), and (3) the Kaneda anterior device. After anterior iliac-crest strut grafting, four types of posterior instrumentation were also tested sequentially: (1) Harrington distraction rods, (2) Luque rectangular instrumentation, (3) Cotrel-Dubousset transpedicular instrumentation, and (4) Steffee transpedicular screws and plates. Rotation, torque, axial displacement, and axial loads were measured during loading across the whole spinal segment between the grip points. Using an anterior extensometer, intervertebral displacement at the second, third, and fourth lumbar levels, and thus across the corpectomy defect at the third lumbar level, was recorded "on line" during testing in flexion and axial load. By recording the intervertebral displacement, the efficacy of each spinal construct in minimizing motion across the corpectomy defect could be quantified. The value for one-way analysis of variance for axial intervertebral displacement across the site of the third lumbar corpectomy was F = 10.5, p less than 0.001. The value for one-way analysis of variance for flexural intervertebral displacement across the corpectomy defect was F = 21.1, p less than 0.001. Homogeneous subsets of rigidity for torsional stiffness revealed that the least rigid constructs were iliac grafting alone, Harrington-rod instrumentation, and Luque rectangular instrumentation. The most rigid constructs were the anterior Kaneda device, transpedicular Cotrel-Dubousset instrumentation, and Steffee screws and plates. CLINICAL RELEVANCE: After corpectomy, spinal reconstructive surgery can restore axial, torsional, and flexural rigidity to normal levels. These experimental conclusions applied to the acute restoration of stability, rather than to rigidity after long-term cyclical loading. Using the most rigid anterior system, the Kaneda device, the fixation extended only one vertebral level cephalad and one level caudad to the corpectomy defect.(ABSTRACT TRUNCATED AT 400 WORDS)

Entities:  

Mesh:

Year:  1988        PMID: 3417703

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  18 in total

1.  Thoracolumbar fracture stabilization: comparative biomechanical evaluation of a new video-assisted implantable system.

Authors:  M Schultheiss; E Hartwig; L Kinzl; L Claes; H-J Wilke
Journal:  Eur Spine J       Date:  2003-11-22       Impact factor: 3.134

2.  Influence of screw-cement enhancement on the stability of anterior thoracolumbar fracture stabilization with circumferential instability.

Authors:  Markus Schultheiss; Erich Hartwig; Lutz Claes; Lothar Kinzl; Hans-Joachim Wilke
Journal:  Eur Spine J       Date:  2004-06-22       Impact factor: 3.134

3.  Load-displacement properties of the thoracolumbar calf spine: experimental results and comparison to known human data.

Authors:  H J Wilke; S T Krischak; K H Wenger; L E Claes
Journal:  Eur Spine J       Date:  1997       Impact factor: 3.134

4.  Current concepts on spinal arthrodesis in degenerative disorders of the lumbar spine.

Authors:  Marios G Lykissas; Alexander Aichmair
Journal:  World J Clin Cases       Date:  2013-04-16       Impact factor: 1.337

5.  Surgical treatments for degenerative lumbar scoliosis: a meta analysis.

Authors:  Guohua Wang; Jianzhong Hu; Xiangyang Liu; Yong Cao
Journal:  Eur Spine J       Date:  2015-04-22       Impact factor: 3.134

6.  Risk factors for medical complications after long-level internal fixation in the treatment of adult degenerative scoliosis.

Authors:  Xi-Nuo Zhang; Xiang-Yao Sun; Xiang-Long Meng; Yong Hai
Journal:  Int Orthop       Date:  2018-04-13       Impact factor: 3.075

7.  Short device fixation and early mobilization for burst fractures of the thoracolumbar junction.

Authors:  F de Peretti; I Hovorka; P M Cambas; J M Nasr; C Argenson
Journal:  Eur Spine J       Date:  1996       Impact factor: 3.134

8.  Is sheep lumbar spine a suitable alternative model for human spinal researches? Morphometrical comparison study.

Authors:  Mahmoud Mageed; Dagmar Berner; Henriette Jülke; Christian Hohaus; Walter Brehm; Kerstin Gerlach
Journal:  Lab Anim Res       Date:  2013-12-20

9.  Treatment of acute thoracolumbar burst fractures with kyphoplasty and short pedicle screw fixation: Transpedicular intracorporeal grafting with calcium phosphate: A prospective study.

Authors:  Panagiotis Korovessis; Thomas Repantis; Petsinis George
Journal:  Indian J Orthop       Date:  2007-10       Impact factor: 1.251

10.  Two levels above and one level below pedicle screw fixation for the treatment of unstable thoracolumbar fracture with partial or intact neurology.

Authors:  Hitesh N Modi; Kook Jin Chung; Il Woo Seo; Hoi Soo Yoon; Ji Hyo Hwang; Hong Kyun Kim; Kyu Cheol Noh; Jung Han Yoo
Journal:  J Orthop Surg Res       Date:  2009-07-27       Impact factor: 2.359

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.