| Literature DB >> 34122566 |
Weijuan Jiang1, Yi Chen1, Zhe Ji1, Yuliang Jiang1, Bin Qiu1, Haitao Sun1, Junjie Wang1.
Abstract
PURPOSE: To investigate the accuracy and dosimetry of 3D-printing non-coplanar template (3D-PNCT)-assisted computed tomography (CT)-guided iodine-125 seed implantation (125I-SI) for recurrent chest wall cancer (rCWC).Entities:
Keywords: 125I; 3D printing; brachytherapy; chest wall cancer; seed implantation
Year: 2021 PMID: 34122566 PMCID: PMC8170518 DOI: 10.5114/jcb.2021.106250
Source DB: PubMed Journal: J Contemp Brachytherapy ISSN: 2081-2841
Characteristics of the 19 patients
| Items | Number/value | % | |
|---|---|---|---|
| Gender | |||
| Male | 13 | 68.4 | |
| Female | 6 | 31.6 | |
| Median age (years) | 58 (20-74) | ||
| Primary cancer | |||
| Esophagus | 4 | 21.0 | |
| Breast | 4 | 21.0 | |
| Cervix | 1 | 5.3 | |
| Lung | 5 | 26.2 | |
| Liver | 1 | 5.3 | |
| Nasopharynx | 1 | 5.3 | |
| Oropharynx | 1 | 5.3 | |
| Stomach | 1 | 5.3 | |
| Chest wall | 1 | 5.3 | |
| Pathology | |||
| Squamous cell carcinoma | 9 | 47.4 | |
| Adenocarcinoma | 4 | 21.0 | |
| Hepatocellular carcinoma | 1 | 5.3 | |
| Invasive ductal carcinoma | 4 | 21.0 | |
| Synovial sarcoma | 1 | 5.3 | |
| Pain | 8 | 42.1 | |
| Previous radiotherapy | 16 | 84.2 | |
| EBRT once | 14 | ||
| Median dose (Gy) | 60 (50-112) | ||
| EBRT twice | 2 | ||
| Median dose (Gy) | 60 (50-76) | ||
| 125I-SI site | |||
| Chest wall | 18 | 81.8 | |
| Cervicothoracic junction | 4 | 18.2 | |
| Median GTV for 125I-SI | 27.4 cm3 (7.9-238.4 cm3) | ||
EBRT – external beam radiotherapy, 125I-SI – 125I seed implantation, GTV – gross tumor volume
Fig. 1A) Digital modeling of individualized 3D-printing non-coplanar templates (3DPNCT); B) 3D-PNCT included anatomical information of the target area, needles pathway and entrance, position lines, etc.
Fig. 23D-printing non-coplanar templates (3D-PNCT) was aligned to target area of the patient’s body according to the laser lines marks and position lines. Two or three anchoring needles (pink) and seed needles (yellow) were inserted
Fig. 3A) Pre-plan, B) intra-operative, and C) post-plan validations. The color lines represent different isodose as shown in the graphical representation (blue, 5500 cGy; yellow, 7700 cGy; pink, 9900 cGy; brownness, 11000 cGy)
Dosimetric parameters of 125I seeds implantation (mean ± SD)
| Pre-plan | Intra-operative | Post-plan | ||||
|---|---|---|---|---|---|---|
| Number of needles | 17.19 ±9.32 | 17.29 ±9.22 | 17.29 ±9.22 | 0.629 | 0.629 | 0.999 |
| Number of seeds | 74.14 ±43.64 | 77.52 ±45.08 | 79.00 ±48.62 | 0.061 | 0.444 | 0.721 |
| D90 | 157.74 ±24.23 | 155.01 ±27.94 | 151.71 ±33.62 | 0.287 | 0.228 | 0.432 |
| D100 | 85.36 ±34.09 | 74.74 ±30.39 | 70.46 ±23.48 | 0.164 | 0.067 | 0.466 |
| V100 | 0.93 ±0.04 | 0.92 ±0.04 | 0.90 ±0.07 | 0.150 | 0.068 | 0.110 |
| V150 | 0.64 ±0.16 | 0.64 ±0.19 | 0.64 ±0.16 | 0.845 | 0.984 | 0.871 |
| V200 | 0.35 ±0.17 | 0.37 ±0.17 | 0.37 ±0.18 | 0.151 | 0.382 | 0.957 |
| CI | 0.57 ±0.16 | 0.53 ±0.16 | 0.52 ±0.15 | 0.006 | 0.007 | 0.175 |
| EI | 0.73 ±0.55 | 0.81 ±0.59 | 0.79 ±0.53 | 0.040 | 0.096 | 0.731 |
| HI | 0.31 ±0.15 | 0.29 ±0.14 | 0.30 ±0.14 | 0.074 | 0.504 | 0.560 |
D90/D100 – prescribed dose delivered to 90%/100% of the gross tumor volume, V100 – V150, and V200 – gross tumor volume receiving 100%, 150%, and 200% of the prescribed dose, respectively, CI – conformation index, EI – external volume index, HI – homogeneity index, *p – pre-plan vs. intra-operative, **p – pre-plan vs. post-plan, ***p – intra-operative vs. post-plan
Numerical rating scale (NRS) for the 8 pain- ful patients
| Patient no. | Pre-seed implantation | Post-seed implantation |
|---|---|---|
| 1 | 2 | 2 |
| 2 | 4 | 1 |
| 3 | 4 | 0 |
| 4 | 4 | 1 |
| 5 | 3 | 0 |
| 6 | 3 | 1 |
| 7 | 4 | 2 |
| 8 | 4 | 0 |