Literature DB >> 34066152

Water as a Blood Model for Determination of CO2 Removal Performance of Membrane Oxygenators.

Benjamin Lukitsch1, Raffael Koller1, Paul Ecker1,2, Martin Elenkov2, Christoph Janeczek2, Markus Pekovits1,2, Bahram Haddadi1, Christian Jordan1, Margit Gfoehler2, Michael Harasek1.   

Abstract

CO2 removal via membrane oxygenators has become an important and reliable clinical technique. Nevertheless, oxygenators must be further optimized to increase CO2 removal performance and to reduce severe side effects. Here, in vitro tests with water can significantly reduce costs and effort during development. However, they must be able to reasonably represent the CO2 removal performance observed with blood. In this study, the deviation between the CO2 removal rate determined in vivo with porcine blood from that determined in vitro with water is quantified. The magnitude of this deviation (approx. 10%) is consistent with results reported in the literature. To better understand the remaining difference in CO2 removal rate and in order to assess the application limits of in vitro water tests, CFD simulations were conducted. They allow to quantify and investigate the influences of the differing fluid properties of blood and water on the CO2 removal rate. The CFD results indicate that the main CO2 transport resistance, the diffusional boundary layer, behaves generally differently in blood and water. Hence, studies of the CO2 boundary layer should be preferably conducted with blood. In contrast, water tests can be considered suitable for reliable determination of the total CO2 removal performance of oxygenators.

Entities:  

Keywords:  carbon dioxide (CO2) removal; computational fluid dynamic simulations; in vitro tests; in vivo tests; membrane modeling; membrane oxygenation; oxygenator development

Year:  2021        PMID: 34066152      PMCID: PMC8151077          DOI: 10.3390/membranes11050356

Source DB:  PubMed          Journal:  Membranes (Basel)        ISSN: 2077-0375


1. Introduction

Blood oxygenators, also known as artificial lungs, are needed to supplement respiratory function during cardiopulmonary bypass or to support patients with respiratory failure. A prerequisite for sufficient gas exchange is a large gas-exchange surface and efficient contact between the blood and membrane. To provide a large gas-exchange surface at the lowest possible priming volume, hollow fiber membrane packings are used [1]. In modern oxygenators, blood flows on the shell side of the packing while the fiber lumen are purged with O2. The transmembrane transfer of the respiratory gases (CO2 and O2) is hereby facilitated by the partial pressure difference between the blood and gas phase [2]. Although great efforts have been made to improve the biocompatibility of oxygenator circuits [3], serious side effects occur due to contact of blood with the artificial polymer surfaces. These side effects ultimately include reduced platelet function and survival as well as prolonged bleeding times after perfusion [4]. Consequently, the optimization of oxygenators aims for increasing gas exchange efficiency while reducing the membrane surface and blood priming volume [5]. For reviewing and studying optimized oxygenator designs, experimental methods are the most reliable. However, the use of blood significantly increases the effort and costs in many ways: The use of blood is accompanied by animal suffering. The use of blood is not permitted in all technical laboratories. Test circuits get contaminated and must be rebuilt due to blood deposits. Additional logistical challenges due to limited durability of blood. Ethics committee approval must be obtained. Water is a cheap, save and easy-to-handle substitute. However, water tests must be able to reasonably represent the CO2 removal performance observed with blood. Here, the difference in physical and chemical properties between blood and waster must be considered. While O2 solubility of blood differs significantly from that of water due to the binding of O2 to hemoglobin [6], the CO2 solubility in water and blood is subject to similar mechanisms. CO2 first dissolves physically and then reacts to carbonic acid (H2CO3), which dissociates to bicarbonate (HCO3−) [7]: In order to shift the reaction equilibrium toward the production of bicarbonate, blood buffers the pH decrease, which is caused by the dissociation reaction. Hence, the total CO2 capacity of blood exceeds that of water. Examining the solubility curves of blood and water at clinically relevant CO2 partial pressures (40 to 100 mmHg) shows that the CO2 capacity of blood is up to 11 times higher (Figure 1a). Additionally, the reaction from CO2 to carbonic acid is accelerated by the catalytic enzyme carbonic anhydrase located in the red blood cells [7].
Figure 1

CO2 solubility behavior of water (Henry [9]) and blood (Loeppky et al. [10]): (a) CO2 concentration in dependency of CO2 partial pressure; (b) slope of CO2 concentration in dependency of CO2 partial pressure.

However, the effect of the buffer system and the enzymatic catalyst on the overall CO2 removal of oxygenators is limited because of two reasons. First, the slope of the solubility curve of blood and water at relevant venous levels is comparable. This is of importance, as the CO2 removal is primarily dependent on the concentration difference. At CO2 partial pressures from 40 to 100 mmHg, the slope of the solubility curve of blood is only four times higher than the slope of water (Figure 1b). Second, outside the red blood cells and in the absence of the catalyst carbonic anhydrase, the reaction from physically dissolved CO2 to carbonic acid and vice versa is slow, compared to the short residence time of blood in an oxygenator. Consequently, mostly physically dissolved CO2 is removed at the membrane [8]. Compared to water, the diffusion of CO2 components in blood is hindered by the presence of proteins [7]. This leads to an approximately 2.6 times higher diffusion rate for physically dissolved CO2 in water than in blood plasma [11]. Furthermore, blood as a suspension of blood plasma and red blood cells shows shear thinning behavior. The shear thinning behavior hereby differs between different animal species [12] and depends, among other parameters, on the hematocrit [13]. At high shear rates, the whole blood viscosity converges toward its minimum (Figure 2). For human blood, this minimum is approximately 3.5 mPa s. In contrast, water is a Newtonian fluid and has a viscosity of 0.69 mPa s at 37 °C [14]. The different rheological behavior of blood and water (Figure 2) has two opposing influences on CO2 separation. On the one hand, lower viscosities result in more turbulent flow conditions. This could promote additional mixing. On the other hand, higher viscosities produce higher shear stress. Due to the higher shear stress, thinner boundary layers can be expected. Both additional mixing at lower viscosities (water) and thinner boundary layers at higher viscosities (blood) would result in an increase in CO2 removal.
Figure 2

Rheologic behavior of blood and water at 37 °C. For representation of whole blood viscosity, the power law viscosity model was used for human [15] and porcine blood [16].

To conclude, blood and water differ in CO2 solubility, CO2 diffusion rate and viscosity. These important parameters have partially opposing effects on the CO2 removal of oxygenators. This makes it difficult to evaluate the suitability and limitations of water as a blood substitute without using in-depth studies. In recent research, multiple in vitro studies were conducted, using water as a blood substitute to determine the CO2 removal of an oxygenator. Hout et al. [17] investigated the dependency of CO2 removal on the sweep flow rate. This was done for two different oxygenator models. In order to be able to use water as a blood substitute, the gas exchange rate was normalized by the maximum CO2 removal rate of the respective oxygenator. By doing so, the measured CO2 removal rates can be made independent from the specific mass transfer characteristics of an oxygenator [18], which, in turn, are dependent on the used test fluid (blood/water). Hattler et al. [19] tested the CO2 and O2 transfer performance of a gas exchange catheter in vitro with water, and in vivo, using calves as a large animal model. However, the suitability of water is only discussed for the O2 exchange. Additionally, the in vitro and in vivo results cannot be compared directly due to the different hydrodynamic conditions. Consequently, a representative deviation of the CO2 removal rates determined with blood and water cannot be calculated. Svitek et al. [11] proposed a Sherwood model, allowing to predict the CO2 removal with blood, based on experiments with water, using an adapted diffusivity. To validate the model, in vitro tests with blood and water were conducted. Although the experimental results show comparable CO2 removal rates for blood and water they are not compared directly by the authors. The suitability of water for prediction of the CO2 removal of oxygenators is only discussed in the scope of the proposed conversion of water to blood data. Tabesh et al. [20] determined the CO2 removal rate of oxygenators via in vitro tests with porcine blood and water. To reduce the difference between the two CO2 removal rates of blood and water, N2 was blended into the saturation stream. This allowed to reduce the maximum deviation to 5%. Yet, the reason for the good fit between blood and water is not analyzed and discussed. Mihelc et al. [21] and Jeffries et al. [22] reported that the CO2 removal measured for an intracorporal membrane catheter correlates well (within 10% deviation) when comparing in vitro trials with water to in vivo trials with calves. The good agreement is attributed to the opposing effects on the CO2 removal rate induced by the different viscosities and CO2 solubilities of blood and water. An in-depth examination of this phenomenon is not conducted. Furthermore, they explicitly limit the suitability of water to their devices and consider the good correlation between water and blood to be fortuitous. A recommendation to use water for the in vitro determination of the CO2 removal rate is not given. To summarize, water is commonly used as a blood substitute for the in vitro determination of the CO2 removal rate of oxygenators. Its suitability for determining the total CO2 removal rate has been confirmed by multiple independent research groups but has never been main focus of their published research. The CO2 solubility, CO2 diffusion rate and viscosity of water differ significantly from those of blood. Due to the complex interactions of these parameters, the reason for the suitability of water as a blood substitute for the determination of CO2 removal in oxygenators remains unclear. The scope of this research is to give a detailed comparison of CO2 removal rates gained from in vitro tests using water with data from in vivo tests using pigs as large animal models. To better understand the contribution of the differing CO2 solubility, CO2 diffusion and viscosity on the CO2 exchange, the CO2 concentration polarization in the boundary layer attached to the membrane is studied. It represents the main CO2 transport resistance and, therefore, characterizes the CO2-removal performance of oxygenators [2]. As the boundary layer cannot be resolved experimentally, computational fluid dynamic (CFD) simulations were conducted to investigate and compare the behavior of the boundary layer for blood and water. By comprising experimental and CFD results, this work aims to evaluate the suitability and limitations of water as a blood substitute for the determination of the CO2 removal rate of oxygenators.

2. Materials and Methods

2.1. In Vivo and In Vitro Tests

The in vivo and in vitro results displayed in this study represent a secondary analysis of previously published data [8,16]. In vivo tests using pigs as large animal models were conducted to validate a CFD model that allows to predict the CO2 removal rate of oxygenators [16]. In vitro tests with water were conducted to evaluate the accuracy of CO2 removal prediction via blood gas analyzer measurements [8]. A comparison of the CO2 removal rate of water and blood, as presented in this study, was not conducted in the previous publications. In both test series, in vivo and in vitro, the CO2 removal rate of an oxygenator prototype was measured. The separation rate was determined at three blood/water flow rates (1000, 1300, and 1600 mL/min). This corresponds to the flow rate range of the oxygenator prototype, which was designed for the partial separation of metabolic CO2 production. For each of the three flow rates, three clinically relevant, pathologically elevated CO2 partial pressures levels (50, 70, and 100 mmHg) were investigated. This equals a total of nine measurement points. For each measurement point, three repetitions were performed. The CO2 removal rate was determined by measuring the flow rate (Defender 510, Bios DryCal, Mesa Laboratories, Inc., Lakewood, CO, USA) and CO2 concentration (BINOS 100 M, Emerson, St Louis, MO, USA) of the sweep gas flow exiting the prototype oxygenator. The examined prototype oxygenator (Figure 3) had a membrane surface of 0.06 m2 provided by commercial polymethylpentene (PMP) hollow fibers with an outer diameter and wall thickness of 380 and 90 µm, respectively. The schematic structure of the tests can be seen in Figure 4. Further experimental details can be found in our preceding publications (in vivo [16] and in vitro tests [8]).
Figure 3

Prototype oxygenator: (a) Principle flow guidance; (b) slice of fiber potting to illustrate fiber arrangement—by Lukitsch et al. [16] (CC BY 4.0).

Figure 4

Schemes of the test setups showing the prototype oxygenator, blood pump, pressure sensors (PR), flow rate sensors (FR), sample ports for the blood gas analyzer (BG) and CO2 concentration sensor (QR): (a) scheme of in vivo loop—by Lukitsch et al. [16] (CC BY 4.0); (b) scheme of in vitro loop—by Lukitsch et al. [8] (CC BY 4.0).

2.2. Computational Fluid Dynamic Simulations

All CFD simulations were conducted with OpenFOAM® 4.1 (ESI Group, Paris, France). The simulations were run on server nodes equipped with 32 core CPUs (16 cores in two physical modules, EPYC 7351, AMD, Santa Clara, CA, USA). For the CFD CO2 transport simulations, the geometry of the prototype oxygenator fiber packing was simplified to reduce the computational effort. Velocity boundary conditions of this reduced geometry were determined by means of an upscaling method [16] that uses CFD flow simulations of a complete or representative part of a hollow fiber module and samples the velocities within this packing. The velocity samples are, afterward, used to calculate an average velocity, which can be used to set the inlet velocity boundary condition of the reduced geometry. In doing so, the method allows to model the flow conditions in the reduced geometry to be representative for the flow regime within the complete fiber packing. CO2 transport simulations of the reduced geometry are then capable to give an accurate prediction of the average transmembrane flux. In Figure 5, the workflow of this upscaling method is illustrated. The upscaling method has been validated for a similar application. In the investigated case, the oxygenator CO2 removal rate predicted by CFD simulations of the reduced and complete geometry deviated by approx. 10% [16]. This deviation is comparable to the deviation between the experimentally determined CO2 removal rate and the CO2 removal rate determined by the CFD CO2 transport simulations of the reduced geometry (Section 3).
Figure 5

Workflow of the upscaling method for CFD based prediction of oxygenator CO2 removal performance—by Lukitsch et al. [16] (CC BY 4.0).

The CFD simulations were performed to extend the experimentally determined data. The experimental measurements of the nine measurement points were performed under steady-state conditions to allow for the three measurement repetitions. Therefore, only the steady-state flow and mass transport problem was considered in the CFD simulations.

2.2.1. Flow Simulation of the Complete Prototype Oxygenator (Macro Scale)

The velocity distribution of water in the complete prototype oxygenator was computed by solving the finite volume formulation of the steady incompressible Navier–Stokes equations, i.e., conservation of mass and momentum. Here, U, p, ρ and µ denote the velocity field, pressure field, fluid density and dynamic viscosity. The conservation of mass and momentum were solved with simpleFoam, the OpenFOAM implementation of the Semi-Implicit Method for Pressure Linked Equations (SIMPLE) algorithm. The computational mesh (Figure 6) was produced with Gambit 2.4.6 (ANSYS, Canonsburg, PA, USA), contained 32 Mio. hexahedron cells and was adapted directly from previous studies [16].
Figure 6

Geometry of CFD flow simulations: (a) Cross section of the prototype oxygenator, inside view; (b) cross section of the fiber packing with close up on the boundary layer mesh of the fibers.

At maximum flow rates, the Reynolds number at the shell side inlet of the prototype oxygenator (Reinlet) is elevated, suggesting a transitional flow regime. Reinlet was calculated using the diameter of the inlet pipe (4 mm) and average velocity in the inlet pipe at the maximum water flow rate (1600 mL/min). In contrast, the Reynolds number within the packing (Repacking) is low and indicates a laminar flow regime. Repacking was calculated using the fiber spacing of the packing (200 µm) and average radial velocity within the packing at the maximum water flow rate (1600 mL/min). Average radial velocity was determined with the CFD flow simulations. Due to the low Reynolds number in the region of interest (membrane packing), we expect limited influence of emerging turbulence on the CO2 removal performance. This assumption is supported by the experimental data. The laminar simulations allow for a reasonable prediction of the dependency of the CO2 removal rate on the blood/water flow rate (Section 3.1.2). The transport equations (Equations (2) and (3)) were discretized using second order schemes (Van Leer [23]). At the beginning of the inlet pipe, uniform velocities were set corresponding to inlet flow rates of 1000, 1300 and 1600 mL/min. A no-slip velocity boundary condition was applied to all walls, including the membrane surfaces. A fixed uniform value of 0 Pa for the relative pressure was set at the end of the outlet pipe. All remaining boundary conditions for velocity and pressure were set to zero gradient (Neumann conditions). All boundary conditions of the flow simulations are summarized in Table 1. Kinematic viscosity of water was set to 6.96 × 10−7 m2/s (37 °C) [14].
Table 1

Boundary conditions of the flow simulations.

BoundaryVelocityPressure
Inletuniform constant valuezero gradient 1
Outletzero gradient 1uniform constant value
Membraneno-slipzero gradient 1
Wallsno-slipzero gradient1

1 Equal to Neumann boundary condition.

The setup of the flow simulations computing the blood distribution in the complete oxygenator prototype is analogous to water flow simulations and described in detail in [8].

2.2.2. CO2 Transport Simulations of the Simplified Packing (Micro Scale)

Details regarding the elaboration of the reduced geometry as well as the design and generation of the mesh can be taken from previous studies [8]. The reduced geometry consists of eight non-staggered fibers (Figure 7) representing the eight fiber mat layers built into the prototype oxygenator (Figure 3b). As the flow simulations show that the fibers are positioned mostly in cross flow mode, the inlet velocity was set to be perpendicular to the membrane packing (Figure 7a). The computational mesh, including only the shell side of the packing, counts 32000 hexahedron cells. To adequately resolve the boundary layer, 20 successively refining cell layers were applied to the membrane surface. In this refinement region, the cell thickness ratio between the most outer cell layer and the most inner cell layer (attached to the membrane) is 5 to 1 (Figure 7b). The cell thickness of the most inner cell equals 0.7 µm.
Figure 7

Simplified geometry of CFD CO2 transport simulations. (a) Fibers representing the eight fiber layers of the membrane packing positioned in a non-staggered arrangement, (b) close up of the computational mesh.

CO2 transport simulations have been performed using an inhouse solver, membraneFoam [24]. It is based on the open-source code OpenFOAM® v4.1 (ESI Group, Paris, France) and implemented as a multi-region solver where the single regions are separated by a membrane. The solver balances the transport equations of velocity, pressure, density, energy and mass fraction for each single region separately. However, the regions are interlinked via the transmembrane transport, which is implemented as a volumetric source term in all of the transport equations. Transmembrane transport is calculated for all cells attached to the membrane. It is computed based on the membrane area of the cell (A) and the permeance (P). As the driving force, the partial pressure difference between the computational cell and an adjacent cell in the neighboring region at the other side of the membrane is utilized. If the membrane area of these two cells does not fully overlap (non-conformal mesh between the regions), partial pressure (pCO) can be interpolated from cells that are close by, adding further flexibility to the design of the mesh. A detailed description of the solver implementation as well as the mathematical formulation of the governing equations are provided by Haddadi et al. [24]. Previous findings indicated only a slight increase in CO2 partial pressure (pCO) on the sweep gas side [25]. Consequently, in this work, membraneFoam was used in a single region mode, setting the CO2 partial pressure on the sweep gas side constantly to 0 mmHg and reducing the computation of transmembrane CO2 transport (JCO) to the following: To solve the governing equations (transport equation of velocity, pressure, density, energy and mass fraction) the Pressure Implicit Method for Pressure-Linked Equations (PIMPLE) was used. They were discretized, applying second order schemes (linear, upwind). The inlet velocity was calculated using the upscaling method [8] and applied uniformly to the inlet patch. The inlet velocities for the reduced geometry corresponding to the blood and water flow rates off the complete geometry (prototype oxygenator) are summarized in Table 2.
Table 2

Inlet velocities of reduced geometry determined with up-scaling method [8].

Total Flow Rate (mL/min)Inlet Velocity Blood (m/s) 1Inlet Velocity Water (m/s) 1
10000.0150.0148
13000.0240.0192
16000.0310.0237

1 Inlet velocity of reduced geometry representative for the total flow rate in prototype oxygenator.

The mass fractions of water and CO2 at the inlet were chosen to correspond to the CO2 partial pressures of 50, 70 and 100 mmHg, which were investigated in vitro and in vivo. On all walls and the outlet, Neumann conditions were used for the CO2 mass fraction. Symmetry conditions were applied to the sides of the geometry to account for the influence of adjacent fibers. The remaining velocity and pressure boundary conditions were set analogous to the flow simulation. All boundary conditions of the CO2 transport simulations are summarized in Table 3.
Table 3

Boundary conditions of the CO2 transport simulations.

BoundaryVelocityPressureCO2
Inletuniform constant valuezero gradient 1uniform constant value
Outletzero gradient 1uniform constant valuezero gradient 1
Membraneuniform constant valuezero gradient 1zero gradient 1,2
Sidessymmetrysymmetrysymmetry

1 Equal to Neumann boundary condition. 2 Transmembrane transport applied as source term in the cell center.

CO2 was treated as a single species. The CO2 partial pressure was computed based on the mass fraction provided by the transport equation and by the use of the Henry’s model. The CO2 solubility (αCO) was set to 8.27 × 10−4 mL CO2 /mL/mmHg [9]. For the diffusion coefficient of CO2 (DCO) in water, a value of 2.38 × 10−9 m2/s was used [26]. Dynamic viscosity (µ) was set independently from CO2 concentration to 6.91 × 10−4 Pa s (at 37 °C) [14]. For the density (ρ) of water, a value of 993.33 kg/m3 (at 37 °C) was used [14]. The setup of the CO2 transport simulation of blood is analogous to the water CO2 transport simulation and described in detail in [8]. To enable a better overview, the material data used for blood and water are compared in Table 4.
Table 4

Comparison of parameter values for blood and water used in CO2 transport simulations.

SymbolDescriptionValue Blood [8]Value WaterUnit
αCO2CO2 solubility8.77 × 10−3 (1)8.27 × 10−4mL CO2/mL/mmHg
DCO2CO2 diffusion coefficient5.05 × 10−10 (2)2.38 × 10−9m2/s
µdynamic viscosity2.38 × 10−3 (3)6.91 × 10−4Pa s
ρdensity at 37 °C1.05 × 1039.93 × 102kg/m3
PCO2CO2 permeance of used fibers157275GPU

1 At 70 mmHg. 2 Diffusion of total CO2 (bicarbonate + physically dissolved CO2). 3 At shear rates > 400 1/s.

Before the trials, the permeances (P) of the prototype oxygenators were measured. The CO2 permeance (PCO) of unused fibers amounts to approx. 730 GPU. Follow-up tests showed that after in vitro or in vivo trials, the CO2 permeance of the prototype oxygenator is decreased considerably (50–150 GPU, 0 h of follow-up testing, Figure 8). During these follow-up tests, the permeances increased successively (to values of 90–380 GPU, 24 h of follow-up testing), most probably because of drying mechanisms induced by the sample gas flow (Figure 8). This observation is consistent with condensate water detected on the lumen side shortly after the beginning of the in vitro or in vivo trials. Both observations (successive increase in permeance and condensate on lumen side) additionally indicate wetting of the pores within the membrane. Furthermore, water residues also remain on the shell side of the membrane packing due to rinsing of the prototype oxygenator after the trials (porcine blood test) or due the trials themselves (water tests).
Figure 8

Increase in CO2 permeance during the follow up gas permeation measurements. Continuous lines: selected values for CFD CO2 transport simulations.

The recorded wetting of the lumen, the membrane pores and the shell reduce the permeances measured in the follow-up tests. However, the permeances set in the CO2 transport simulations must only account for the wetting of the lumen and pores, as the shell side (equivalent with blood/water side) CO2 transport resistance is resolved by the transport equations themselves. Thus, permeances required for CO2 transport simulations cannot be accurately measured for two reasons. First, the transport resistance of the wetted lumen and pores cannot be measured separately from the transport resistance present in the shell. Second, the drying of the shell results in the drying of the lumen and membrane pores to an uncertain extent, affecting the measured permeances. Consequently, the permeances used in the simulations were chosen in order to fit the transmembrane CO2 flux of the simulation and experiments at a pCO of 70 mmHg and a water/blood flow rate of 1300 mL/min. To keep the deviations between simulations and experiments below 10%, different permeances had to be chosen for blood and water. CO2 permeance with water (275 GPU) is 1.8 times higher than CO2 permeance with blood (157 GPU). This could be explained by the blood residues adding an additional transport resistance. The identified permeances lie within the measured range of pure gas permeances determined for the used membranes in the follow-up tests. A comparison of the permeances used in the CFD CO2 transport simulations and the permeances measured in the follow-up tests is given in Figure 8.

3. Results and Discussion

3.1. In Vivo and In Vitro Tests

In the following section, the CO2 removal rate of the prototype oxygenator determined in vivo with porcine blood and in vitro with water is compared. Additionally, the experimental results are used to validate the CFD simulations of CO2 transport in porcine blood and water.

3.1.1. CO2 Removal Rate of In Vivo Porcine Blood and In Vitro Water Tests

Figure 9 summarizes the results of the in vivo tests with porcine blood and the in vitro tests with water. It shows the average CO2 removal of the prototype oxygenator for the nine measurement points (Section 2.1). The CO2 removal determined with porcine blood is generally higher than the CO2 removal of water. This is discussed in detail in Section 3.2.3. In both tests, CO2 removal increases with higher CO2 partial pressure (pCO) or higher blood/water flow rate. The dependency of CO2 removal on the flow rate increases with a higher pCO. This can be explained by an increased availability of physically dissolved CO2 at a higher pCO and associated higher pH levels [7]. Since physically dissolved CO2 can be regarded as the main CO2 component removed via the membrane surface [8], higher availability of physically dissolved CO2 offers a larger potential of CO2 removal increase via flow rate increase.
Figure 9

Comparison of experimentally determined CO2 removal. Red squares: in vivo studies with porcine blood; blue circles: in vitro studies with water.

The average deviation (ε) of the CO2 removal rate determined in vitro with water (JCO,water) from that determined in vivo with blood (JCO,blood) was calculated using the following equation: Average deviation is shown for the nine measurement points in Table 5. For pCO ≥ 70 mmHg, ε is homogenously distributed and equals approximately 9%. At pCO of 50 mmHg, ε shows a dependency on the flow rate. The deviation is highest at 1000 mL/min (21.8%) and decreases monotonously to a value of 2.1% at a flow rate of 1600 mL/min.
Table 5

Deviation of the in vitro CO2 removal with water to the in vivo CO2 removal with blood (ε).

Flow Rate 1Deviation (ε) of In Vitro CO2 Removal (Water) to In Vivo CO2 Removal (Blood)
(mL/min)pCO2: 50 mmHgpCO2: 70 mmHgpCO2: 100 mmHg
100021.8%8.6%9.7%
13009.9%10.6%6.7%
16002.1%7.9%9.6%

1 Flow rate of blood or water on prototype oxygenator shell side.

However, the elevated ε at a pCO of 50 mmHg and flow rate of 1000 mL/min (21.8%) could be caused by increased measurement errors, as at this measurement point, ε is two times the average and twice as large as the second largest deviation (Table 5). Furthermore, the dependency of ε on the flow rate is caused by the negligible, small dependency of the CO2 removal rate on the blood flow rate, measured in the in vivo trials at a pCO of 50 mmHg. This small, recorded dependency can only be justified physically to a limited extent since at a pCO of 50 mmHg, a dependency of the CO2 separation rate on the flow rate is predicted by both in vitro water tests and CFD CO2 transport simulations of blood (Section 3.1.2). Furthermore, Sherwood correlations also suggest a similar dependency [27]. In summary, this indicates increased measurement errors at a pCO level of 50 mmHg.

3.1.2. Validation of CFD CO2 Transport Simulations for Porcine Blood and Water

To validate the CFD CO2 transport simulations, the numerically predicted CO2 removal rate is compared with the experimentally determined CO2 removal rate. Figure 10 shows a comparison of the experimental and computational results for porcine blood. In general, the results agree well. Average deviation of the numerically from the experimentally determined separation rate is 6%. However, a stronger deviation has to be noted at higher pCO (100 mmHg) and higher flow rates (1600 mL/min). This can be attributed partly to an increased standard deviation (σ) of pCO during the in vivo trials at a level of 100 mmHg (σ = 7.8 mmHg), compared to the other two pCO levels (pCO: 50 mmHg, σ = 2.7 mmHg; pCO: 70 mmHg, σ = 3.8 mmHg). The dependence of the CO2 removal rate on the blood flow rate can be adequately described by the upscaling method used.
Figure 10

Comparison of CO2 removal determined with experiments and CFD simulations for blood.

Figure 11 compares the experimentally and numerically determined CO2 removal rate for water. The results are in satisfactory agreement. The deviation of the experimentally and numerically determined CO2 removal rate is, on average, 3%. The dependence of CO2 removal on pCO and water flow rate can be adequately described by the CFD model and the used upscaling method.
Figure 11

Comparison of CO2 removal determined with experiments and CFD simulations for water.

The good agreement between the experimental and numerical results for both porcine blood and water suggests that the CFD models are suitable for detailed studies of the boundary layer. The latter determines the CO2 removal performance of an oxygenator since it can be considered the main transport resistance of the respiratory gas exchange.

3.2. Computational Fluid Dynamic—CO2 Transport Simulations

In the following section, the results of the CO2 transport simulations of porcine blood and water are summarized. Furthermore, the behavior of the boundary layer at different flow rates and CO2 partial pressures is discussed. Finally, the influence of the CO2 diffusion coefficient, CO2 permeance, CO2 solubility and viscosity on the deviation of the CO2 removal of porcine blood and water is evaluated.

3.2.1. Flow and CO2 Partial Pressure Distribution in the Simplified Packing

Figure 12 compares the velocity distribution in the simplified fiber packing for porcine blood and water. Flow distribution is presented for an inlet velocity of 0.02 m/s (approx. 1300 mL/min). Porcine blood flow shows stronger velocity gradients and higher maximum flow rates (0.10 m/s) than water (0.08 m/s). Additionally, the velocity field indicates laminar flow profiles free of wakes. In comparison, the water flow distribution shows small wakes downstream of the fibers. However, in these areas, only low velocities (approx. 0.01 m/s) occur. In general, the flow behaves similarly at all fibers.
Figure 12

Flow distribution within the simplified packing (a) porcine blood (b) water. Bottom half of contour plot shows direction of velocity vectors.

In Figure 13, the CO2 partial pressure (pCO) contour plot of blood and water are compared for a uniform inlet pCO of 70 mmHg. The CO2 partial pressure distribution in water shows a larger area with decreased pCO than blood. This is most pronounced downstream of the fibers where the small wakes are positioned. This indicates that these wakes produce additional mixing. Furthermore, the higher CO2 diffusion rates of water could contribute to CO2 depletion in wider areas of the flow.
Figure 13

Distribution of CO2 partial pressure within the simplified packing (a) porcine blood (b) water.

3.2.2. Boundary Layer Study

As can be seen in Figure 12 and Figure 13, flow and pCO distribution depend only slightly on fiber position. They deviate strongest for the first fiber, as it is the only fiber which is not positioned in the slip-stream of another fiber. This can be also seen in Figure 14. It shows the pCO profiles perpendicular to the main flow direction and the membrane wall for all eight fibers. The presented data were computed for a pCO of 70 mmHg and an inlet velocity of 0.02 m/s (approx. 1300 mL/min). The sample lines of these profiles and the fiber positions are illustrated in Figure 7a.
Figure 14

CFD boundary profiles of CO2 partial pressure at different fiber positions (sampling line illustrated in Figure 7a) for an inlet pCO of 70 mmHg and inlet velocity of 0.02 m/s (approx. 1300 mL/min), fiber 1—first fiber at inlet, fiber 8—last fiber at outlet: (a) porcine blood, (b) water.

The pCO profiles of blood (Figure 14a) depend less on fiber position than the pCO profiles of water (Figure 14b). This could be attributed to the higher Re of the water flow, which promotes additional mixing. The latter increases the effects of the upstream fibers on the pCO distribution of downstream fibers. In general, the first fiber shows the thinnest boundary layer. The boundary layer becomes thicker the further downstream the fiber is positioned. For the relatively homogenous flow distribution observed in this simplified packing, the average of all eight fibers gives a reasonable representation of the single pCO profiles of the individual fiber positions (Figure 14). Consequently, in the following graphs, only the average of all eight fiber profiles are shown in order to maintain clarity. Figure 15 compares the average pCO boundary layer profiles for the three different inlet pCO levels (50, 70, 100 mmHg) at an inlet velocity of 0.02 m/s (approx. 1300 mL/min). With decreasing distance to the hollow fiber membrane, the pCO decreases slowly at first, but drops steeply in the last section toward the fiber. This steep drop starts at comparable positions independent from the inlet pCO. Hence, the pCO gradient of this section deviates strongly between the different investigated inlet CO2 partial pressures. The gradient of pCO equals, on average, 3.5 mmHg/µm for an inlet pCO of 100 mmHg and 1.7 mmHg/µm for an inlet pCO of 50 mmHg.
Figure 15

Boundary profiles of CO2 partial pressure (pCO2) and normalized CO2 partial pressure (pCO′) (sample line illustrated in Figure 7a) at different inlet pCO and an inlet velocity of 0.02 m/s (approx. 1300 mL/min): (a) porcine blood (b) water (normalized profiles are overlapping).

Additionally, the normalized pCO profiles (pCO(x)) are given. They are calculated by dividing the pCO at any point in the packing (pCO(x)) by the maximum pCO in the bulk flow (pCO2,max), which is equal to the pCO at the inlet (pCO,inlet—Equation (6). While for water the normalized pCO profiles are very similar for all three inlet pCO (overlapping of all three dimensionless profiles, Figure 15b), the normalized pCO profiles for blood deviate slightly (Figure 15a). However, pCO is capable of giving a reasonable representation of the pCO boundary layers at different inlet CO2 partial pressures. Figure 16 compares the dependence of the pCO boundary layer profiles on the inlet velocities of the simplified packing for porcine blood and water. Five inlet velocities (0.005, 0.01, 0.02, 0.03, 0.1 m/s) were simulated. With decreasing inlet velocities, the boundary layers become thicker and the pCO gradients smaller. The principal shape of the profiles differs depending on whether higher or lower flow velocities are present. At higher inlet velocities (≥0.1 m/s), there is a low decrease in pCO with decreasing fiber distance followed by a steep decrease. At lower inlet velocities (≤ 0.005m/s), pCO decreases gradually with fiber distance. In general, water shows thicker pCO boundary layers than blood and a more pronounced dependence of the pCO boundary layer on the inlet velocity.
Figure 16

Boundary profiles (sampling line illustrated in Figure 7a) of normalized CO2 partial pressure (pCO2/pCO2,max) for different inlet CO2 partial pressures: (a) porcine blood, (b) water.

Figure 17 compares the boundary layer thickness (δ) for blood and water at different velocities. The end of the boundary layer was defined at the fiber distance (x) where 99% of bulk flow pCO was reached (Equation (7)) [28]. The fiber distance and pCO were taken along the sample line, illustrated in Figure 7a. For pCO, average profiles of all eight fibers were used.
Figure 17

Dependency of boundary layer thickness on: (a) maximum velocity between fibers; (b) Reynolds number computed with maximum velocity between fibers.

In Figure 17a, δ is plotted over the maximum velocity between fibers. The boundary layer thickness of porcine blood is smaller and, in contrast to water, dependent on inlet pCO. At elevated velocities, this dependency becomes less. In Figure 17b, δ is plotted over the Reynolds number (Re). Re was calculated Equation (8) using the fiber diameter (dfiber), maximum velocity between fibers (umax), and, due to elevated shear rates > 400 s−1 in larger parts of the geometry, Newtonian kinematic viscosity (ν) of the respective fluids (porcine blood or water, Figure 2). In general, water shows a higher Re than porcine blood due to its lower viscosity. The boundary layer thickness deviates more strongly when comparing porcine blood and water at the same Re (Figure 17b) than when comparing at the same maximum velocities (Figure 17a). This indicates that at similar velocities, additional mixing, promoted by lower viscosity of water (higher Re), does reduce the difference in the boundary layer thickness between blood and water. The boundary layer thickness of porcine blood and water at the same maximum shear stress agree reasonably (Figure 18a), confirming that boundary layer thickness is mainly dependent on shear stress [29]. However, when comparing the CO2 removal rates of porcine blood and water at same shear stresses (Figure 18b), the deviation is stronger (min. deviation 20% for max shear stress > 1 Pa) than when compared at the same blood/water flow rates (approx. 10%, Figure 9). Consequently, the boundary layer thickness and CO2 removal rate cannot be matched simultaneously.
Figure 18

Influence of maximum shear stress on: (a) boundary layer thickness; (b) specific CO2 removal rate.

The differences between the boundary layer mass transfer characteristics of porcine blood and water can be confirmed by establishing dimensionless Sherwood correlations. The mass transfer analogy for crossflow within hollow fiber membrane packings is usually expressed by the dimensionless Sherwood (Sh) Reynold (Re) and Schmidt (Sc) number, as well as the empirical parameters a and b [27]: To account for the diffusion of bicarbonate and physically dissolved CO2 in blood, the definitions of Sh and Sc proposed by Federspiel et al. [11] were used. Sh is calculated using the CO2 mass transfer coefficient (kCO), fiber diameter (dfiber), CO2 solubility in blood (αCO, Table 4), and facilitated diffusion of CO2 in blood (Df). Sc number is defined by the Newtonian kinematic viscosity of blood (νblood) and effective diffusion of CO2 in blood (Deff). The Sh and Sc numbers of water can be calculated analogously. Here, facilitated and effective diffusivity are replaced with the diffusivity of CO2 in water (Table 4). The CO2 mass transfer coefficient was determined with the numerically predicted average CO2 flux (jCO) and the CO2 partial pressure difference between the blood/water bulk flow and gas side (ΔpCO): The Re number was calculated according to Equation (10). The definitions of Df and Deff are given in [11]. Comparison of the Sherwood analogies (Figure 19) shows different correlations for porcine blood and water. While parameter a of porcine blood (0.47) and water (0.51) is comparable, parameter b deviates by a factor of 2 (b porcine blood: 0.21, b water: 0.42). This leads to agreement between the two correlations only for Re numbers of about 1. The varying values of Sh/Sc0.33 at different pCO for porcine blood are caused by the non-linear dependency of the CO2 partial pressure and CO2 concentration (cCO) and the definition of kCO, which uses pCO instead of cCO as the driving force (Equation (14)). Based on the available data, CO2 mass transfer within the boundary layer of porcine blood and water does not exhibit similar characteristics.
Figure 19

Sherwood correlations for porcine blood and water determined via the CFD results.

3.2.3. Influences of Fluid Properties on Boundary Layer Thickness

In further CFD studies, the dependency of the boundary layer thickness on important fluid and material parameters was examined. Therefore, the original CFD model for water was adapted by singularly modifying the CO2 diffusion coefficient (CO2 diff. coeff.), CO2 permeance, CO2 solubility and the viscosity model to the properties of blood. The boundary layer thicknesses, determined with these modified CFD models, are compared in Figure 20 for a flow rate of 1300 mL/min and CO2 partial pressures of 50 and 100 mmHg. As reference, the boundary layer thicknesses determined with the original CFD models are entered. A summary of the model and parameter values can be found in Table 4.
Figure 20

Influence of CO2 diffusion, CO2 permeance, CO2 solubility, and viscosity model on the boundary layer thickness.

Modification of the water CFD model by changing individual fluid and material parameters to those of blood resulted in a reduction in boundary layer thickness in all of the studied cases (Figure 20). The strongest dependency of boundary layer thickness was detected for adaption of the CO2 diffusion coefficient from a value of 2.38 × 10−9 m2/s (CO2 in water) to 6.96 × 10−10 m2/s (CO2 in blood). By this adaption, the boundary layer thickness reduced to 57 µm, which is comparable to porcine blood with a boundary layer thickness of 55 µm at 100 mmHg inlet pCO. The reduction in boundary layer thickness induced by CO2 permeance, CO2 solubility model and viscosity is less pronounced. The boundary layer thickness determined with these simulations is similar and approximately 68 µm. Of all model adaptions, only the CO2 solubility model of blood introduces a dependency of the boundary layer thickness on the inlet pCO. This is probably due to the binding of CO2 in multiple components (physically dissolved, bicarbonate and carbaminohemoglobin) and the resulting nonlinear dependency of the CO2 partial pressure from the CO2 concentration (Figure 1a).

3.2.4. Influences of Fluid Properties on CO2 Removal Rate

Additional CFD studies were conducted to examine the influence of the CO2 diffusion coefficient (CO2 diff. coeff.), CO2 permeance, CO2 solubility and the viscosity model on the specific CO2 removal rate. To do so, the original CFD model for water was adapted by singularly modifying the fluid and material parameters to those of blood. Figure 21 compares the specific CO2 removal rate at a flow rate of 1300 mL/min and a CO2 partial pressure of 70 mmHg, determined with the modified and original CFD models for blood and water. As can be seen in Figure 21, the CO2 removal rates of blood and water are comparable despite the different fluid and material properties (Table 4). This is due to their opposing effects on the CO2 removal rate. While the higher CO2 solubility of blood allows higher CO2 removal rates, the slower CO2 diffusion in blood, the lower permeance of membranes contacted with blood and the higher blood viscosity lower the CO2 removal rate. The strongest effect on the CO2 removal rate is caused by the difference in CO2 solubility between blood and water (121% increase), followed by the difference in CO2 diffusion rate (53% decrease) and the difference in CO2 permeance (18% decrease). Based on the CFD results, the rheologic difference between blood and water has the smallest effect on the CO2 removal rate (10% decrease).
Figure 21

Influence of CO2 diffusion, CO2 permeance CO2 solubility, and viscosity model on the CO2 removal rate.

4. Conclusions

The CO2 removal rate of a prototype oxygenator was measured in vivo using porcine blood and in vitro using water. In general, the CO2 removal rates of porcine blood and water are comparable despite differing fluid properties (CO2 diffusion, CO2 solubility and viscosity). The deviation of the CO2 removal rate determined with porcine blood from that determined with water amounts to approximately 10%. This deviation agrees well with the data found in recent literature [20,21,22]. Based on the results of the in vivo and in vitro tests, the CFD CO2 transport simulations were validated. Besides differences in the fluid properties of blood and water, our experimental and simulation data indicate that additionally higher CO2 permeances are available during tests with water, probably due to the absence of blood residues on the membrane surface. The influences of the differing fluid properties (CO2 diffusion, CO2 solubility and viscosity) as well as membrane material properties (CO2 permeance) on CO2 removal were quantified, utilizing the developed CFD models. The difference in CO2 solubility between blood and water has the strongest effect on the CO2 removal rate, followed by the difference in the CO2 diffusion rate, difference in CO2 permeance and difference in viscosity. The CFD simulations also allow to resolve and study the main CO2 transport resistance—the diffusional boundary layer attached to the membrane surface. The simulations show that the pCO boundary layer in water is, in general, thicker than that in blood. The CFD results indicate that the thicker boundary layer in water can be mainly attributed to the higher diffusion coefficient of CO2 in water (2.38 × 10−9 m2/s) than in blood (6.96 × 10−10 m2/s). Furthermore, the CFD model suggests that the pCO boundary layer thickness in blood is dependent on bulk pCO. This is probably due to the binding of CO2 in multiple components (physically dissolved, bicarbonate and carbaminohemoglobin) and the resulting nonlinear dependency of the CO2 partial pressure from the CO2 concentration (CO2 solubility). Additionally, CFD simulations indicate that the pCO boundary layer thicknesses of porcine blood and water are in good agreement when compared at same shear stresses. However, the CO2 removal rates of porcine blood and water deviate stronger at the same shear stresses (>20%) than at the same blood/water flow rates (approx. 10%). Consequently, the boundary layer thickness and CO2 removal rate cannot be matched simultaneously. Differences in the CO2 mass transfer characteristics of the boundary layer can be confirmed when comparing Sherwood correlations established for porcine blood and water. To conclude, the boundary layers of blood and water behave, in general, differently. Studies of blood oxygenators aiming to investigate the boundary layer or measures for boundary layer reduction should, therefore, rely on blood tests. However, the determination of the total CO2 removal rate of an oxygenator using water as a blood substitute should be possible with reasonable accuracy. Deviation of the CO2 removal rate determined with water to the CO2 removal rate determined with blood should lie within 10%. This level of accuracy could potentially differ for flow conditions other than those studied here (crossflow, Re within the packing approx. 20). Nevertheless, the results of this work provide a basis for future optimization of the CO2 removal performance of oxygenators, using in vitro tests with water.
  15 in total

1.  Evaluation of a respiratory assist catheter that uses an impeller within a hollow fiber membrane bundle.

Authors:  Kevin M Mihelc; Brian J Frankowski; Samuel C Lieber; Nathan D Moore; Brack G Hattler; William J Federspiel
Journal:  ASAIO J       Date:  2009 Nov-Dec       Impact factor: 2.872

2.  Sweep gas flowrate and CO2 exchange in artificial lungs.

Authors:  W J Federspiel; B G Hattler
Journal:  Artif Organs       Date:  1996-09       Impact factor: 3.094

Review 3.  Carbon dioxide transport and carbonic anhydrase in blood and muscle.

Authors:  C Geers; G Gros
Journal:  Physiol Rev       Date:  2000-04       Impact factor: 37.312

Review 4.  Theory of oxygen transport to tissue.

Authors:  A S Popel
Journal:  Crit Rev Biomed Eng       Date:  1989

5.  Transient adhesion of platelets in pump-oxygenator systems: influence of SMA and nitric oxide treatments.

Authors:  J Li; M K Sly; R Chao; A Constantinescu; P V Kulkarni; F H Wians; M E Jessen; R C Eberhart
Journal:  J Biomater Sci Polym Ed       Date:  1999       Impact factor: 3.517

6.  Whole blood viscosity, plasma viscosity and erythrocyte aggregation in nine mammalian species: reference values and comparison of data.

Authors:  U Windberger; A Bartholovitsch; R Plasenzotti; K J Korak; G Heinze
Journal:  Exp Physiol       Date:  2003-05       Impact factor: 2.969

7.  Effect of impeller design and spacing on gas exchange in a percutaneous respiratory assist catheter.

Authors:  R Garrett Jeffries; Brian J Frankowski; Greg W Burgreen; William J Federspiel
Journal:  Artif Organs       Date:  2014-04-22       Impact factor: 3.094

8.  Validation of a model for flow-dependent carbon dioxide exchange in artificial lungs.

Authors:  M S Hout; B G Hattler; W J Federspiel
Journal:  Artif Organs       Date:  2000-02       Impact factor: 3.094

9.  Quantitative description of whole blood CO2 dissociation curve and Haldane effect.

Authors:  J A Loeppky; U C Luft; E R Fletcher
Journal:  Respir Physiol       Date:  1983-02

10.  Suitable CO2 Solubility Models for Determination of the CO2 Removal Performance of Oxygenators.

Authors:  Benjamin Lukitsch; Paul Ecker; Martin Elenkov; Christoph Janeczek; Christian Jordan; Claus G Krenn; Roman Ullrich; Margit Gfoehler; Michael Harasek
Journal:  Bioengineering (Basel)       Date:  2021-03-02