A respiratory gas exchange catheter: in vitro and in vivo tests in large animals.

Brack G Hattler1, Laura W Lund, Joseph Golob, Heide Russian, Michael F Lann, Thomas L Merrill, Brian Frankowski, William J Federspiel.   

Abstract

OBJECTIVES: Acute respiratory failure is associated with a mortality of 40% to 50%, despite advanced ventilator support and extracorporeal membrane oxygenation. A respiratory gas exchange catheter (the Hattler Catheter) has been developed as an oxygenator and carbon dioxide removal device for placement in the vena cava and right atrium in the treatment of acute respiratory failure to improve survival.
METHODS: Differing from a previously clinically tested intravenous gas exchange device (ie, IVOX), the Hattler Catheter incorporates a small, pulsating balloon surrounded by hollow fibers. The pulsating balloon redirects blood toward the fibers, enhances red cell contact with the membrane, and significantly improves gas exchange so that smaller catheter devices are still efficient on insertion and can be inserted through the jugular or femoral vein. Devices were tested in mock circulatory loops and in short-term (8 hours) and long-term (4 days) experiments in calves to study the effect of various sized balloons and the anatomic location of the device in the venous system as a function of hemodynamics and gas exchange.
RESULTS: In vitro performance in water demonstrates an oxygen delivery (Vo(2)) of 140 +/- 8.9 mL. min(-1). m(-2) and a carbon dioxide removal (Vco(2)) of 240 +/- 6.1 mL. min(-1). m(-2). Acute in vivo experiments demonstrate a maximum carbon dioxide consumption of 378 +/- 11.2 mL. min(-1). m(-2). Devices positioned in the right atrium had an average carbon dioxide exchange of 305 mL. min(-1). m(-2), whereas in the inferior vena cava position carbon dioxide exchange was 255 mL. min(-1). m(-2). Devices have been tested long term in calves, with gas exchange rates maintained over this time interval (carbon dioxide consumption, 265 +/- 35 mL. min(-1). m(-2)). Plasma-free hemoglobin levels at the end of 4 days have been 4.8 +/- 3.2 mg/dL. Hemodynamic measurements, including a decrease in cardiac outputs and increased mean pressure decreases across the device become significant only with the larger balloon (40-mL) devices (P <.05, 40-mL vs 13-mL devices). Autopsies show no end-organ damage. The device linearly increases its carbon dioxide output with progressive hypercapnea, predicting its ability to meet tidal volume reduction in the therapy of respiratory failure.
CONCLUSIONS: Progress has been made toward developing an intravenous gas exchange catheter to provide temporary pulmonary support for patients in acute respiratory failure.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12202869     DOI: 10.1067/mtc.2002.123811

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  15 in total

1.  Evaluation of local gas exchange in a pulsating respiratory support catheter.

Authors:  Heide J Eash; Brian J Frankowski; Brack G Hattler; William J Federspiel
Journal:  ASAIO J       Date:  2005 Mar-Apr       Impact factor: 2.872

2.  Flow visualization study of a pulsating respiratory assist catheter.

Authors:  Stephanus G Budilarto; Brian J Frankowski; Brack G Hattler; William J Federspiel
Journal:  ASAIO J       Date:  2005 Nov-Dec       Impact factor: 2.872

Review 3.  Artificial lung basics: fundamental challenges, alternative designs and future innovations.

Authors:  Heather Nolan; Dongfang Wang; Joseph B Zwischenberger
Journal:  Organogenesis       Date:  2011-01-01       Impact factor: 2.500

Review 4.  The evolution of patient selection criteria and indications for extracorporeal life support in pediatric cardiopulmonary failure: next time, let's not eat the bones.

Authors:  Joseph R Custer
Journal:  Organogenesis       Date:  2011-01-01       Impact factor: 2.500

5.  Blood biocompatibility assessment of an intravenous gas exchange device.

Authors:  Trevor A Snyder; Heide J Eash; Kenneth N Litwak; Brian J Frankowski; Brack G Hattler; William J Federspiel; William R Wagner
Journal:  Artif Organs       Date:  2006-09       Impact factor: 3.094

6.  Towards improved artificial lungs through biocatalysis.

Authors:  Joel L Kaar; Heung-Il Oh; Alan J Russell; William J Federspiel
Journal:  Biomaterials       Date:  2007-03-21       Impact factor: 12.479

Review 7.  Extracorporeal carbon dioxide removal (ECCO2R) in respiratory deficiency and current investigations on its improvement: a review.

Authors:  Hany Hazfiza Manap; Ahmad Khairi Abdul Wahab
Journal:  J Artif Organs       Date:  2016-05-18       Impact factor: 1.731

8.  Acute in vivo testing of a respiratory assist catheter: implants in calves versus sheep.

Authors:  Heide J Eash; Brian J Frankowski; Kenneth Litwak; William R Wagner; Brack G Hattler; William J Federspiel
Journal:  ASAIO J       Date:  2003 Jul-Aug       Impact factor: 2.872

9.  Evaluation of fiber bundle rotation for enhancing gas exchange in a respiratory assist catheter.

Authors:  Heide J Eash; Kevin M Mihelc; Brian J Frankowski; Brack G Hattler; William J Federspiel
Journal:  ASAIO J       Date:  2007 May-Jun       Impact factor: 2.872

10.  Development of a novel intravascular oxygenator catheter: Oxygen mass transfer properties across nonporous hollow fiber membranes.

Authors:  Stewart Farling; Tobias L Straube; Travis P Vesel; Nick Bottenus; Bruce Klitzman; Ira M Cheifetz; Marc A Deshusses
Journal:  Biotechnol Bioeng       Date:  2020-10-07       Impact factor: 4.530