Thomas Bowman1,2, Elisa Gervasoni1, Chiara Arienti1, Stefano Giuseppe Lazzarini1, Stefano Negrini3,4, Simona Crea1,2, Davide Cattaneo1,5, Maria Chiara Carrozza1,2.
Abstract
Wearable devices are used in rehabilitation to provide biofeedback about biomechanical or physiological body parameters to improve outcomes in people with neurological diseases. This is a promising approach that influences motor learning and patients' engagement. Nevertheless, it is not yet clear what the most commonly used sensor configurations are, and it is also not clear which biofeedback components are used for which pathology. To explore these aspects and estimate the effectiveness of wearable device biofeedback rehabilitation on balance and gait, we conducted a systematic review by electronic search on MEDLINE, PubMed, Web of Science, PEDro, and the Cochrane CENTRAL from inception to January 2020. Nineteen randomized controlled trials were included (Parkinson's n = 6; stroke n = 13; mild cognitive impairment n = 1). Wearable devices mostly provided real-time biofeedback during exercise, using biomechanical sensors and a positive reinforcement feedback strategy through auditory or visual modes. Some notable points that could be improved were identified in the included studies; these were helpful in providing practical design rules to maximize the prospective of wearable device biofeedback rehabilitation. Due to the current quality of the literature, it was not possible to achieve firm conclusions about the effectiveness of wearable device biofeedback rehabilitation. However, wearable device biofeedback rehabilitation seems to provide positive effects on dynamic balance and gait for PwND, but higher-quality RCTs with larger sample sizes are needed for stronger conclusions.Entities:
Keywords: Parkinson’s disease; biofeedback; force sensors; gait; inertial measurement unit; mild cognitive impairment; postural balance; rehabilitation; stroke
Mesh:
Year: 2021 PMID: 34063355 PMCID: PMC8156914 DOI: 10.3390/s21103444
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Search strategy example on Pubmed.
| AND | (Humans[mh] OR Adult[mh] OR Nervous System Diseases[mh] OR Gait Disorders, Neurologic[mh] AND (Neurofeedback[mh] OR Feedback, Sensory[mh] OR feedback[tiab] OR Biofeedback[tiab] OR Cues[mh] OR Physical Therapy modalities[mh] OR Rehabilitation[mh] OR Rehab*[tiab] OR Conservative treatment[mh] OR Training[tiab] OR Exercise*[tiab]) |
| (Wearable Electronic Devices[mh] OR wearable[tiab] OR Device[tiab] OR Accelerometry[mh] OR Acceleromet*[tiab] OR gyroscope*[tiab] OR sensor*[tiab] OR shoe*[tiab] OR Insole*[tiab]) | |
| (Walking[mh] OR Walk*[tiab] OR Ambulation[tiab] OR Gait[mh] OR Gait[tiab] OR Postural Balance[mh] OR Balance[tiab] OR Equilibrium[tiab] OR Recovery of function[mh] OR Motor Activity[mh]) | |
| (Randomized controlled trial[pt] OR randomized controlled trials as topic[mh] OR random*[tiab]) |
Figure 1Flow diagram based on PRISMA statement (www.prisma-statement.org, accessed on 1 March 2020).
Characteristics of included studies.
| First Author [Ref] | Aim | Participant Characteristics | Wearable Device | Sensors Type (Short) | Experimental Intervention | Control Intervention | Timing | Balance and Gait Outcome Measure | Evaluation Time Points |
|---|---|---|---|---|---|---|---|---|---|
| Azerpaikan et al. [ | To study the effect of a neurofeedback training on balance problems associated with Parkinson’s disease. | TOT n = 16 patients with PD | Biograph Infiniti Software system (version 5.0), the ProComp differential amplifier (Thought Technology Ltd, Montreal, Quebec) for NeuroFeedback training sessions (FlexComp Infiniti encoder, TT-USB interface unit, fiber optic cable, USB cable) | EEG SENSORS | EG: Neurofeedback training with EEG generator | EG: Sham Neuro Feedback Training using sham EEG generator | D:8 sessions | BBS*, Limit of stability * | PRE, POST |
| Byl et al. [ | To evaluate the effectiveness of supervised gait training with and without visual kinematic feedback | TOT: n = 24 stroke and PD | Wireless joint angle sensors and Smart Shoes | IMU, PRESSURE SENSORS | EG: visual kinematic feedback on the computer screen during progressive and task-oriented balance and gait training activities | CG: balance and gait training activities | D: 12 sessions (6–8 weeks) | Gait Speed; Step Length; TBS, 6MWT, DGI, 5TSS, TUG, BBS, FOG-Q | PRE, POST |
| Carpinella et al. [ | To test the feasibility of a wearable biofeedback system in a typical rehabilitation gym and analyze the effect on balance and gait outcome measures compared to physiotherapy without feedback. | TOT: n = 42 subjects with PD | Gamepad System (IMU, PC and Customized software) | IMU SENSORS | EG: balance and gait functional tailored exercises using Gamepad System | CG: personalized balance and gait exercises defined by the clinical staff | D: 20sessions | BBS *, Gait Speed, UPDRSIII, TUG, ABC, FOGQ, COP ML sway *, COP AP sway, Tele-healthcare Satisfaction Questionnaire-Wearable Technologies | PRE, POST, FU |
| Cha et al. [ | To compare the effectiveness of auditory feedback stimulation from the heel and forefoot areas in terms of ambulatory functional improvements in stroke patients | TOT n = 31 stroke subjects EG1: n = 11 Mean age: 64.6 ± 10.6 Female: n = 3 EG2: n = 10 Mean age: 63.0 ± 4.7 Female: n = 3 CG: n = 10 Mean age: 61.8 ± 9.8 Female: n = 4 | A PedAlert Monitor 120 (PattersonMedical Holdings Inc., | PRESSURE SENSORS | EG1: gait training with active weight bearing on the paretic heel with auditory feedback EG2: gait training with auditory feedback from paretic metatarsals | CG= gait intervention | D: 6 weeks T: 50 min (30 min of conventional therapy + 20 min of gait intervention with or without auditory feedback) F: 3 × week | Gait speed, FGA *, TUG, COP length EO *, COP length EC, COP velocity EO *, COP velocity EC | PRE, POST |
| Cho et al. [ | To examine whether visual biofeedback tracking training can improve gait performance in chronic stroke patients. | TOT n = 10 Stroke subjects | A double-axis electrogoniometer (Biometrics Ltd. Ladysmith, VA) was used to record the instant degrees | ELECTRO-GONIOMETER | EG: visual biofeedback tracking training | CG: not Reported | D:20 sessions (4 week) T: 39 min F: 5 days/week | Motoricity Index, Modified Motor Assessment Scale, Gait Speed. | PRE, POST |
| Choi et al. [ | To compare gait intervention with auditory feedback induced by active weight bearing on the paralyzed side with the effects of the general gait training method | TOT n = 24 stroke subjects EG: n = 12 Mean age: 62.8 ± 4.8 Female: n = 4 CG: n = 12 Mean age: 59.7 ± 10.2 Female: n = 4 | PedAlert Monitor 120, (Patterson Medical Holdings, Inc.). | PRESSURE SENSORS | EG: gait intervention with auditory feedback | CG: general gait training over the ground. | D: 6 weeks T: 50 min (30 min of conventional therapy + 20 mins of gait training with or without bfb) F: 3 × week | Gait Speed (10MWT) *, FGA *, TUG *, COP length EO *, COP length EC* | PRE, POST |
| Cozean et al. [ | To study the effect of EMG Biofeedback and FES as therapies for gait dysfunction in patients with hemiplegia after stroke. | TOT n = 36 | Not reported | EMG SENSORS | EG1: EMG biofeedback during static and dynamic activities EG2: FES during static and dynamic activities EG3: EMG biofeedback+FES during static and dynamic activities | CG: conventional Physical Therapy | D: 6 week T: 30 min F: 3 × Week | ankle angle (swing phase) *, knee angle (swing phase) *, stride length, gait speed* | PRE, POST |
| El-Tamawy et al. [ | To examine the influence of paired proprioceptive cues on gait parameters of individuals with PD. | TOT n = 30 subjects with PD EG: n = 15 Mean age: 61.4 ± 7.28 Female: not specified CG: n = 15 Mean age: 63.2 ± 5.6 Female: not specified | The vibratory device, OPTEC Co. LLtd. | PRESSURE SENSORS | EG: individually designed physiotherapy and traditional gait training plus treadmill training with vibratory stimuli | CG: individually designed physiotherapy and traditional gait training including instructions to walk with long steps. | EG: D: 8 weeks T: 51–70 min F: 3 sess/week, CG: D: 8 weeks T: 45 min F: 3 sess/week | Cadence *, Stride length *, Gait speed *, Walking distance * | PRE, POST |
| Ginis et al. [ | To test the feasibility of CuPID system in the home environment and verify differential effects of CuPID training versus conventional home-based gait intervention | TOT n = 40 subjects with PD EG: n = 22 Mean age: not specified | The CuPiD system consisted of a smartphone (Galaxy S3-mini, | IMU SENSORS | EG: received weekly home visit and patients were instructed to walk with the CUPID system | CG: received weekly home visit by the researcher who gave advice on gait and freezing, and patients were instructed to walk without using the CUPID system | D: 6 weeks T: 30 min F:3 times/week | MiniBEST *, FSST, FES-I, 2MWT, UPDRS III, NFOG-Q, Comfortable gait and Dual task activities (gait speed, stride length, DS time); | PRE, POST, FU |
| Intiso et al. [ | to evaluate the efficacy of electromyographic biofeedback compared with physical therapy. | TOT n = 16 | table Satem PT | EMG SENSORS | EG: EMG Biofeedback and Physical Therapy (standard exercise bobath, facilitation, and inhibition techniques, neurofacilitatory techniques) | CG: Physical therapy (standard exercise bobath, facilitation and inhibition techniques, neurofacilitatory techniques) | D: 2 months T: 60 min F: daily physical therapy (Only EG 30 session of EMG BFB) | Basmajian scale *, Gait speed, step length, ankle angle (swing phase) *, ankle angle (heel contact) | PRE, POST |
| Jonsdottir et al. [ | to assess the efficacy of EMG-BFB applied in a task-oriented approach based on principles of motor learning to increase peak ankle power of the affected leg and gait velocity in patients with chronic mild to moderate hemiparesis | TOT n = 20 | BFB device: (SATEM Mygotron, SATEM srl, Rome, Italy); EMG, system (band-pass filtered at 20 to 950 Hz and then amplified with a gain of 40 000 (50 mVrms range), | EMG SENSORS | EG=Task-oriented gait training with EMG BFB device | CG= conventional physical therapy (at least 15 mins of gait training in each session) | D: 20 session T: 45 min F: 3 × Week | Gait speed *, ankle power peak at push-off *, stride length *, knee flexion peak | PRE, POST, FU |
| Jung et al. [ | to examine the effect of gait training using a cane with an augmented pressure sensor to improve weight bearing on the nonparetic leg in patients with stroke | TOT 22 stroke subjects EG: n = 12 Mean age: 56.4 ± 11.1 Female: n = 4 CG: n = 10 Mean age: 56.3 ± 17.1 Female: n = 3 | An instrumented cane, outfitted with a pressure sensor (CD 210-K200, Dacell Co. Ltd., Korea) connected to an indicator (DN30W, Dacell Co. Ltd., Korea) | PRESSURE SENSORS | EG: gait training with auditory feedback | CG: gait training without auditory feedback | D: 4 weeks T: 60 min (30 min +30 min gait training with or without bfb) F: 5 × week | Peak force cane *, Gait speed *, single support time * | PRE, POST |
| Ki et al. [ | to examine the effects of auditory feedback during gait on the weight bearing of patients with hemiplegia resulting from a stroke. | TOT n = 30 stroke subjects EG: n = 12 Mean age: 55.3 ± 9.2 Female: n = 4 CG: n = 13 Mean age: 60.1 ± 12.3 Female: n = 2 | A pressure gauge Ped-AlertTM120 (ORBITEC, USA) | PRESSURE SENSORS | EG: neurodevelopmental treatment with auditory feedback | CG: neurodevelopmental treatment | D: 4 weeks T: NA F: NA. | TUG *, Stance phase duration, Single support time. | PRE, POST |
| Lee et al [ | to examine the effect of neurofeedack training on brain waves control and gait performed under dual-task conditions. | TOT n = 20 stroke subjects EG: n = 10 Mean age: 53.2 ± 6.46 Female: n = 4 CG: n = 10 Mean age: 54.7 ± 3.77 Female: n = 3 | The Procomp Infiniti system (SA7951 version 5.1, Thought Technology, Canada) was used for neurofeedback training. The QEEG-8 (LXE3208, LAXHA Inc., Korea) system was used to measure brain waves | EEG SENSORS | EG: neurofeedback training | CG: pseudo-neurofeedback training (sham neurofeedback) | D: 8 weeks T: 30 min F: 3 × week | gait speed *, cadence *, stance phase percentage, and plantar foot pressure (dual task) * | PRE, POST |
| Lupo et al. [ | To evaluate the efficacy of training involving the use of a combined biofeedback system versus conventional balance training | TOT n = 15 stroke subjects EG: n = 9 Mean Age: 52.56 ± 13.92 Female: n = 3 CG: n = 6 Mean Age: 65.66 ± 9.64 Female: n = 1 | The RIABLO™ (CoRehab, Trento, Italy) system comprised of several inertial measurement unitsand a force platform connected wirelessly to a computer | IMU SENSORS AND FORCE PLATFORM | EG: balance training with RIABLO biofeedback system using a video interface. | CG: conventional balance training without the use of the RIABLO biofeedback system | D: 10 sessions T: 20 min F:3 times/week | BBS *, RMI, COP length EO *, COP length EC * | PRE, POST, FU |
| Mandel et al. [ | to investigate the efficacy of electromyographic (EMG) versus a novel biofeedback (BFB) approach to improve ankle control and functional gait in stroke patients | TOT n = 37 stroke subjects | Two channels of EM-BFB, a Lamoureux-type parallelogram electrogoniometer was and a computerized | EMG SENSORS and ELECTRO-GONIOMETER | EG1: EMG biofeedback training (computer-generated auditory and visual feedback of calf and pretibial muscle activity during active ankle movements) EG2: EMG biofeedback followed by rhythmic positional biofeedback (computer-generated single-channel feedback of dorsiflexion and plantar flexion) | CG: No training | D: 24 session (EG2 performed 12 session of EMG biofeedback and 12 session of Rythmical positional biofeedback) T: not specified F: 3 × Week | Gait speed *, ROM * | PRE, POST, FU |
| Schwenk et al. [ | To evaluate the feasibility and experience in using the new sensor-based training in a sample of patients with clinically confirmed amnestic MCI | TOT n = 32 subjects with MCI EG: mean age 77.8 ± 6.9 Female: n = 7 CG: mean age 79 ± 10.4 Female: n = 5 | The technology consisted of a 24 inch computer screen, an interactive virtual user interface, and 5 inertial sensors (LegSysTM, BioSensics LLC, MA, USA) | IMU SENSORS AND FORCE PLATFORM | EG: postural balance exercises during standing (ankle point-to-point reaching tasks and virtual obstacles crossing tasks) using biofeedback training | CG: No training | EG: D: 4 weeks T: 45 mins F: 2 sessions/week | CoM area EO *, CoM area EC, CoM sway ML EO *, CoM swayML EC, CoM sway AP EO *, CoM sway AP EC, Gait Speed, Gait stride time variability, FES-I * | PRE, POST |
| Sungkarat et al. [ | To determine whether improved symmetrical weight bearing somatosensory feedback would result in improved gait and balance in people with stroke | TOT n = 35 people with stroke EG: n = 17 Mean Age: 52.12 ± 7.17 years Female: n = 5 CG: n = 18 Mean Age: 53.83 ± 11.18 years Female: n = 6 | Insole Shoe Wedge and Sensors (I-ShoWS) | PRESSURE SENSORS | EG: conventional rehabilitation and gait training with I-ShoWS set-up | CG: conventional rehabilitation and gait training without I-ShoWS set-up | D:15 sessions T: 60 min/session, (30 min gait training and 30 min conventional) F: 5 days/week | Gait speed *, Step Length asymmetry ratio*, Single Support time asymmetry ratio *, Load on paretic leg during stance (% Body Weight) *, | PRE, POST |
| van den Heuvel et al. [ | to investigate the feasibility of visual feedback-based balance training (VFT) and to compare the effects of the training program with conventional training | TOT n = 33 subjects with PD EG: mean age 63.9 ± 6.39 Female: n = 5 CG:mean age 68.8 ± 9.68 Female: n = 8 | Flat-panel LCD monitor connected to a PC (Motek Medical, Amsterdam, The Netherlands), Force plate (Forcelink, Culemborg, and The Netherlands) and Inertial sensors (X sens, Enschede, The Netherlands) | IMU SENSORS | EG: interactive balance games with explicit augmented visual feedback | CG: conventional balance training recommended by the guidelines for physical therapy. | D: 5 weeks T: 60 min (45 min balance workstation) F: 2 sess/week | BBS; Single leg stance test; Gait speed, FES-I, UPDRSIII* | PRE, POST, FU |
* = p ≤ 0.005; 5TSS = 5 Time sit to stand; 2MWT = 2 min Walk Test; 6MWT = 6 min Walk Test; AP = Antero-posterior; BFB = Biofeedback; BBS = Berg balance scale; CG = Control group; COM = Centre of movement; COP = Centre of pressure; D = Duration; DGI = Dynamic Gait Index; EEG = Electroencephalogram; EG = Experimental group; EMG = Electromyography, F = Frequency; FES-I = Fall efficacy scale; FGA = Functional gait assessment; FOG-Q = Freezing of gait questionnaire; FSST = Four Square step test; IMU = Inertial movement unit; MCI = Mild cognitive impairment; ML = Medio-lateral; NFOG-Q = New freezing of gait questionnaire; PD = Parkinson disease; RMI = Rivermead mobility index; ROM = Range of motion; T = time; TBS = Tinetti balance scale; TOT = Total; TUG= Timed up and go; UPDRS = Unified Parkinson disease rating scale.
Figure 2Classification of sensors used in the included studies: biomechanical (IMU, pressure and electrogoniometer) and physiological (EMG and EEG).
Figure 3Sensors placement and number of the studies. (A) Biomechanical sensors: Chest: Carpinella et al., Lupo et al., and Van den Heuvel et al.; Back: Carpinella et al., Schwenk et al.; Thigh: Byl et al., Carpinella et al., Schwenk et al., and Lupo et al.; Shank: Byl et al., Carpinella et al., Lupo et al., and Schwenk et al.; Feet: Byl et al., Choi et al., Cha et al., El-tamawy et al., Ginis et al., Ki et al., and Sugkarat et al.; (B) Physiological sensors: Head: Azerpaikan et al. and Lee et al.; Knee: Cho et al.; M. Tibialis anterior: Cozean et al., Intiso et al., and Mandel et al.; M. Gastrocnemius: Jonsdottir et al., Mandel et al., and Cozean et al.
Biofeedback components.
| Author, Year | Biofeedback | Biofeedback Content | Biofeedback Frequency | Biofeedback Timing | Reinforce |
|---|---|---|---|---|---|
| Azerpaikan, 2014 | Visual | Performance | Constant | Concurrent | Positive |
| Byl, 2015 | Visual | Performance, Result | Constant | Terminal | Positive, Negative |
| Carpinella, 2017 | Auditory, Visual | Performance, Result | Fading | Concurrent, Terminal | Positive, Negative |
| Cha, 2018 | Auditory | Performance | Constant | Concurrent | Positive |
| Cho, 2007 | Visual | Performance | Constant | Concurrent | Not specified |
| Choi, 2019 | Auditory | Performance | Constant | Concurrent | Positive |
| Cozean, 1988 | Auditory, Visual | Performance | Constant | Concurrent | Positive |
| El-Tamawy, 2012 | Vibrotactile | Performance | Constant | Concurrent | Positive |
| Ginis, 2016 | Auditory | Performance | Fading | Concurrent | Positive, Negative |
| Intiso, 1994 | Auditory | Performance | Constant | Concurrent | Positive |
| Jonsdottir, 2010 | Auditory | Performance | Fading | Concurrent | Positive |
| Jung, 2015 | Auditory | Performance | Constant | Concurrent | Negative |
| Ki, 2015 | Auditory | Performance | Constant | Concurrent | Positive |
| Lee, 2015 | Visual | Performance | Constant | Concurrent | Positive |
| Lupo, 2018 | Auditory, Visual | Performance, Result | Constant | Concurrent | Not specified |
| Mandel, 1990 | Auditory, Visual | Performance | Constant | Concurrent | Positive |
| Schwenk, 2016 | Auditory, Visual | Performance | Constant | Concurrent | Positive, Negative |
| Sungkarat, 2011 | Auditory | Performance | Constant | Concurrent | Positive |
| van den Heuvel, 2014 | Visual | Performance, Result | Constant | Concurrent | Positive |
Figure 4Forest plots of studies comparing add-on WDBR versus control. (A) Overall effect of WDBR versus control at postassessment on BBS. (B) Overall effect of WDBR versus control at follow-up assessment on BBS.
Figure 5Forest plots of studies comparing add-on WDBR versus control. (A) Overall effect of WDBR versus control at postassessment on TUG. (B) Subgroup analysis: effect of WDBR (pressure sensors with auditory biofeedback) versus control at postassessment on TUG.
Figure 6Forest plots of studies comparing add-on WDBR versus control. (A) Overall effect of WDBR versus control at postassessment on gait speed. (B) Overall effect of WDBR versus control at follow-up assessment on gait speed. (C) Subgroup analysis: effect of WDBR (pressure sensors with auditory biofeedback) versus control at postassessment on gait speed.
Figure 7Risk of bias graph: review authors judgments about each risk of bias item presented as percentages across all included studies.
Figure 8Risk of bias summary: review authors’ judgements about each risk of bias item for each included study.