Literature DB >> 34046219

In-vivo dose measurements with MOSFET dosimeters during MV portal imaging.

Sathish Kumar1, Rabi Raja Singh1, Henry Finlay Godson1, Retna Ponmalar1, Paul Ravindran1, Sunil Dutt Sharma2, Subhashini John1.   

Abstract

BACKGROUND: The purpose of this study was to investigate the feasibility of MOSFET dosimeter in measuring eye dose during 2D MV portal imaging for setup verification in radiotherapy.
MATERIALS AND METHODS: The in-vivo dose measurements were performed by placing the dosimeters over the eyes of 30 brain patients during the acquisition of portal images in linear accelerator by delivering 1 MU with the field sizes of 10 × 10 cm2 and 15 × 15 cm2.
RESULTS: The mean doses received by the left and right eyes of 10 out of 30 patients when both eyes were completely inside the anterior portal field were found to be 2.56 ± 0.2 cGy and 2.75 ± 0.2, respectively. Similarly, for next 10 patients out of the same 30 patients the mean doses to left and right eyes when both eyes were completely out of the anterior portal fields were found to be 0.13 ± 0.02 cGy and 0.17 ± 0.02 cGy, respectively. The mean doses to ipsilateral and contralateral eye for the last 10 patients when one eye was inside the anterior portal field were found to be 3.28 ± 0.2 cGy and 0.36 ± 0.1 cGy, respectively.
CONCLUSION: The promising results obtained during 2D MV portal imaging using MOSFET have shown that this dosimeter is well suitable for assessing low doses during imaging thereby enabling to optimize the imaging procedure using the dosimetric data obtained. In addition, the documentation of the dose received by the patient during imaging procedure is possible with the help of an in-built software in conjunction with the MOSFET reader module.
© 2021 Greater Poland Cancer Centre.

Entities:  

Keywords:  MOSFET; MV portal imaging; in vivo dosimetry; setup verification

Year:  2021        PMID: 34046219      PMCID: PMC8149133          DOI: 10.5603/RPOR.a2021.0016

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  16 in total

1.  Comparison of entrance and exit dose measurements using ionization chambers and silicon diodes.

Authors:  S Heukelom; J H Lanson; B J Mijnheer
Journal:  Phys Med Biol       Date:  1991-01       Impact factor: 3.609

Review 2.  Emergent technologies for 3-dimensional image-guided radiation delivery.

Authors:  David A Jaffray
Journal:  Semin Radiat Oncol       Date:  2005-07       Impact factor: 5.934

Review 3.  Clinical use of electronic portal imaging.

Authors:  Michael G Herman
Journal:  Semin Radiat Oncol       Date:  2005-07       Impact factor: 5.934

4.  The management of imaging dose during image-guided radiotherapy: report of the AAPM Task Group 75.

Authors:  Martin J Murphy; James Balter; Stephen Balter; Jose A BenComo; Indra J Das; Steve B Jiang; C M Ma; Gustavo H Olivera; Raymond F Rodebaugh; Kenneth J Ruchala; Hiroki Shirato; Fang-Fang Yin
Journal:  Med Phys       Date:  2007-10       Impact factor: 4.071

5.  A free program for calculating EUD-based NTCP and TCP in external beam radiotherapy.

Authors:  Hiram A Gay; Andrzej Niemierko
Journal:  Phys Med       Date:  2007-09-07       Impact factor: 2.685

Review 6.  TCP and NTCP: a basic introduction.

Authors:  Michael Baumann; Cordula Petersen
Journal:  Rays       Date:  2005 Apr-Jun

7.  Multinomial model-based formulations of TCP and NTCP for radiotherapy treatment planning.

Authors:  R Keinj; T Bastogne; P Vallois
Journal:  J Theor Biol       Date:  2011-03-30       Impact factor: 2.691

8.  Comprehensive QA for radiation oncology: report of AAPM Radiation Therapy Committee Task Group 40.

Authors:  G J Kutcher; L Coia; M Gillin; W F Hanson; S Leibel; R J Morton; J R Palta; J A Purdy; L E Reinstein; G K Svensson
Journal:  Med Phys       Date:  1994-04       Impact factor: 4.071

9.  Phantom and in-vivo measurements of dose exposure by image-guided radiotherapy (IGRT): MV portal images vs. kV portal images vs. cone-beam CT.

Authors:  Cornelia Walter; Judit Boda-Heggemann; Hansjörg Wertz; Iris Loeb; Angelika Rahn; Frank Lohr; Frederik Wenz
Journal:  Radiother Oncol       Date:  2007-11-26       Impact factor: 6.280

10.  In vivo diode dosimetry for routine quality assurance in IMRT.

Authors:  P D Higgins; P Alaei; B J Gerbi; K E Dusenbery
Journal:  Med Phys       Date:  2003-12       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.