Literature DB >> 3402847

New Jersey low contact stress total ankle replacement: biomechanical rationale and review of 23 cementless cases.

F F Buechel1, M J Pappas, L J Iorio.   

Abstract

A congruent contact, unconstrained, multiaxial ankle replacement has been developed for use without cement. A talar onlay component with a trochlear surface and central fixation fin uses a cylindrical articulating axis that reproduces the lateral talar curvature. A tibial inlay component with a 7 degree anteriorly inclined short fixation stem uses a flat loading plate, recessed anatomically into the distal tibia to distribute tibial loads to the ankle joint. For both components, made of cast cobalt-chromium-molybdenum, a 275-micron pore-size, sintered-bead, porous coating is used to allow tissue ingrowth stabilization. A congruent ultra-high molecular weight polyethylene bearing is inserted between the metallic implants. Its upper surface is flat, whereas its lower surface conforms to the trochlear surface, thereby providing unconstrained, sliding cylindrical motion with low contact stress on the bearing surfaces. Contact pressure and collateral ligaments maintain ankle stability during both static and dynamic loading conditions. Clinically, 23 total ankle arthroplasties were performed in 21 patients. The follow-up period ranged from 24 months to 64 months with a mean of 35.3 months. Diagnoses included rheumatoid arthritis, 6 patients (26.1%); osteoarthritis, 4 patients (17.4%); post-traumatic arthritis, 10 patients (43.5%); avascular necrosis of the talus, 2 patients (8.7%), and painful ankle fusion, 1 patient (4.3%). Pain was the primary reason for surgery in all cases. Postoperatively, 87% of ankles had no pain or, at most, mild pain. Postoperative complications included poor wound healing in four ankles, reflex sympathetic dystrophy in two ankles, deep infection in one ankle, and one bearing subluxation. No ankle replacements were removed and no fusions were performed for failed implants, although one bearing was exchanged without disrupting the metallic elements. In this report, the suggestion is made that total ankle arthroplasty may have an improved application in various arthritis disorders when used with biologic fixation and unconstrained mobile bearings.

Entities:  

Mesh:

Year:  1988        PMID: 3402847     DOI: 10.1177/107110078800800603

Source DB:  PubMed          Journal:  Foot Ankle        ISSN: 0198-0211


  12 in total

1.  Dynamic simulation of the natural and replaced human ankle joint.

Authors:  A Leardini; D Moschella
Journal:  Med Biol Eng Comput       Date:  2002-03       Impact factor: 2.602

2.  Forty-year outcome of ankle "cup" arthroplasty for post-traumatic arthritis.

Authors:  Dawson C Muir; Annunziato Amendola; Charles L Saltzman
Journal:  Iowa Orthop J       Date:  2002

3.  The trapezoid form of the trochlea tali.

Authors:  E Brenner; J Piegger; W Platzer
Journal:  Surg Radiol Anat       Date:  2003-07-24       Impact factor: 1.246

Review 4.  Total ankle replacement.

Authors:  Alexej Barg; Matthias D Wimmer; Martin Wiewiorski; Dieter C Wirtz; Geert I Pagenstert; Victor Valderrabano
Journal:  Dtsch Arztebl Int       Date:  2015-03-13       Impact factor: 5.594

Review 5.  [Modern three-piece total ankle replacement. Frequency and causes of luxation and premature wear of the polyethylene bearing].

Authors:  A H Hoffmann; B Fink
Journal:  Orthopade       Date:  2007-10       Impact factor: 1.087

6.  Radiographic evaluation of INBONE total ankle arthroplasty: a retrospective analysis of 30 cases.

Authors:  Abhijit Datir; Minzhi Xing; Aparna Kakarala; Michael R Terk; Sameh A Labib
Journal:  Skeletal Radiol       Date:  2013-09-12       Impact factor: 2.199

Review 7.  Which are the most frequently used outcome instruments in studies on total ankle arthroplasty?

Authors:  Florian D Naal; Franco M Impellizzeri; Pascal F Rippstein
Journal:  Clin Orthop Relat Res       Date:  2009-08-12       Impact factor: 4.176

8.  [The HINTEGRA ankle: short- and mid-term results].

Authors:  B Hintermann; V Valderrabano; M Knupp; M Horisberger
Journal:  Orthopade       Date:  2006-05       Impact factor: 1.087

9.  Computer-assisted design of the sagittal shapes of a ligament-compatible total ankle replacement.

Authors:  A Leardini; F Catani; S Giannini; J J O'Connor
Journal:  Med Biol Eng Comput       Date:  2001-03       Impact factor: 3.079

10.  Clinical Radiographical Outcomes and Complications after a Brand-New Total Ankle Replacement Design through an Anterior Approach: A Retrospective at a Short-Term Follow Up.

Authors:  Massimiliano Mosca; Silvio Caravelli; Emanuele Vocale; Simone Massimi; Davide Censoni; Marco Di Ponte; Mario Fuiano; Stefano Zaffagnini
Journal:  J Clin Med       Date:  2021-05-23       Impact factor: 4.241

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.