Victoria Eley1,2, Rebecca Christensen3,4, Louis Guy3,4, Kerstin Wyssusek3,4, Anita Pelecanos5, Benjamin Dodd4,6, Michael Stowasser4,7, Andre van Zundert3,4.
Abstract
BACKGROUND: Measuring blood pressure in patients with obesity is challenging. The ClearSight™ finger cuff (FC) uses the vascular unloading technique to provide continuous non-invasive blood pressure measurements. We aimed to test the agreement of the FC with invasive radial arterial monitoring (INV) in patients with obesity.Entities:
Keywords: ClearSight™; Invasive blood pressure; Non-invasive blood pressure; Obesity; Vascular unloading
Mesh:
Year: 2021 PMID: 34006231 PMCID: PMC8130355 DOI: 10.1186/s12871-021-01374-x
Source DB: PubMed Journal: BMC Anesthesiol ISSN: 1471-2253 Impact factor: 2.217
Fig. 1Recruitment flowchart
Characteristics of 30 patients presenting for elective bariatric surgery
| Characteristic | Description |
|---|---|
| Weight, (kg) median (IQR) | 146 (133 to 151) |
| BMI category (kg/m2) n (%) | |
| 45.0–49.9 | 14 (47) |
| 50.0–54.9 | 7 (23) |
| 55.0–59.9 | 5 (17) |
| ≥ 60.0 | 4 (13) |
| Diagnosis of hypertension n (%) | 10 (33) |
| Ethnicity n (%) | |
| Caucasian | 28 (93) |
| Aboriginal or Torres Strait Islander | 2 (7) |
| Type of laparoscopic surgery n (%) | |
| Sleeve gastrectomy | 10 (33) |
| Roux-en-y gastric bypass | 10 (33) |
| Single anastomosis gastric bypass | 5 (17) |
| Remove gastric band and convert to gastric bypass | 4 (13) |
| Removal of gastric band | 1 (3) |
| Right mid-arm circumference (cm), mean ± SD (range) | 47.2 ± 6.3 (35–61) |
| Left mid-arm circumference (cm), mean ± SD (range) | 46.4 ± 5.7 (38–58) |
| Right arm conicity index, mean ± SD (range) | 14.9 ± 5.0 (4.9–24.3) |
| Left arm conicity index, mean ± SD (range) | 15.3 ± 4.4 (8.5–22.7) |
| Vasopressor total dose | |
| Metaraminol (mg), median (IQR) ( | 3.5 (2.0 to 6.2) |
| Phenylephrine (mcg), median (IQR) ( | 955 (450 to 2210) |
| Ephedrine (mg), individual value presented | 6 |
Comparison of values for mean arterial pressure, systolic blood pressure and diastolic blood pressure obtained from the finger cuff compared with invasive arterial line. Summary of the calculated bias and 95% limits of agreement with 95% CIs calculated and used to create the modified Bland-Altman plots shown in Fig. 2, INV – FC
| Bias ± SD | Lower LOA (95% CI) mmHg | Upper LOA (95% CI) mmHg | |
|---|---|---|---|
| Mean arterial pressure | 5.2 ± 10.9 | -16.0 (-22.0 to -11.9) | 26.5 (22.3 to 32.4) |
| Systolic blood pressure | 14.3 ± 14.1 | -13.4 (-20.6 to -8.2) | 42.0 (36.9 to 49.2) |
| Diastolic blood pressure | 2.6 ± 10.8 | -18.6 (-24.4 to -14.4) | 23.8 (19.7 to 29.6) |
FC finger cuff measurements, INV invasive measurements, LOA limits of agreement
Fig. 2a Modified Bland-Altman plot for mean arterial pressure (MAP). The plot shows the agreement between measurements from invasive radial arterial monitoring (MAP-INV) and the finger cuff (MAP-FC). b Modified Bland-Altman plot for systolic blood pressure (SBP). The plot shows the agreement between measurements from invasive radial arterial monitoring (SBP-INV) and the finger cuff (SBP-FC). c Modified Bland-Altman plot for diastolic blood pressure (DBP). The plot shows the agreement between measurements from invasive radial arterial monitoring (DBP-INV) and the finger cuff (DBP-FC). Participants’ multiple measurements are presented individually as participant number. The middle horizontal line indicates the bias, the bottom and top lines are the 95% limits of agreement and shaded regions represent the 95% CIs for the lower and upper limits of agreement
Fig. 3a Error grid analysis for systolic blood pressure (SBP). The figure shows the error grid for the test method (ClearSight™ finger cuff, FC) FC-SBP compared with the reference method (invasive radial arterial monitoring, INV) INV-SBP. b Error grid analysis for mean arterial pressure (MAP). The figure shows the error grid for the test method (ClearSight™ finger cuff, FC) FC-MAP compared with the reference method (invasive radial arterial monitoring, INV) INV-MAP
Fig. 4Four-quadrant plots of consecutive differences across time in a SBP, b MAP and c DBP, comparing measurements obtained from the ClearSight™ finger cuff (FC) and invasive radial arterial monitoring (INV). Grey squares indicate +/− 5 mmHg exclusion zone. Top left and bottom right quadrants indicate discordant pairs