A comparison of symmetry and flatness measurements in small electron fields by different dosimeters in electron beam radiotherapy.

Mohamad Reza Bayatiani1, Fatemeh Fallahi2, Akbar Aliasgharzadeh2, Mahdi Ghorbani3, Benyamin Khajetash4, Fatemeh Seif1.   

Abstract

BACKGROUND: Symmetry and flatness are two quantities which should be evaluated in the commissioning and quality control of an electron beam in electron beam radiotherapy. The aim of this study is to compare symmetry and flatness obtained using three different dosimeters for various small and large fields in electron beam radiotherapy with linac.
MATERIALS AND METHODS: Beam profile measurements were performed in a PTW water phantom for 10, 15 and 18 MeV electron beams of an Elekta Precise linac for small and large beams (1.5 × 1.5 cm2 to 20 × 20 cm2 field sizes). A Diode E detector and Semiflex-3D and Advanced Markus ionization chambers were used for dosimetry.
RESULTS: Based on the obtained results, there are minor differences between the responses from different dosimeters (Diode E detector and Semiflex-3D and Advanced Markus ionization chambers) in measurement of symmetry and flatness for the electron beams. The symmetry and flatness values increase with increasing field size and electron beam energy for small and large field sizes, while the increases are minor in some cases.
CONCLUSIONS: The results indicate that the differences between the symmetry and flatness values obtained from the three dosimeter types are not practically important.
© 2021 Greater Poland Cancer Centre.

Entities:  

Keywords:  Advanced Markus; Diode E; Semiflex-3D; electron beam; flatness; radiotherapy; small field dosimetry; symmetry

Year:  2021        PMID: 33948302      PMCID: PMC8086714          DOI: 10.5603/RPOR.a2021.0009

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  13 in total

1.  Clinical impact of the detector size effect in 3D-CRT.

Authors:  Feliciano García-Vicente; Maria José Béjar; Leopoldo Pérez; Juan J Torres
Journal:  Radiother Oncol       Date:  2004-11-25       Impact factor: 6.280

2.  Clinical electron-beam dosimetry: report of AAPM Radiation Therapy Committee Task Group No. 25.

Authors:  F M Khan; K P Doppke; K R Hogstrom; G J Kutcher; R Nath; S C Prasad; J A Purdy; M Rozenfeld; B L Werner
Journal:  Med Phys       Date:  1991 Jan-Feb       Impact factor: 4.071

3.  Recommendations for clinical electron beam dosimetry: supplement to the recommendations of Task Group 25.

Authors:  Bruce J Gerbi; John A Antolak; F Christopher Deibel; David S Followill; Michael G Herman; Patrick D Higgins; M Saiful Huq; Dimitris N Mihailidis; Ellen D Yorke; Kenneth R Hogstrom; Faiz M Khan
Journal:  Med Phys       Date:  2009-07       Impact factor: 4.071

4.  Evaluation of a diode detector array for use as a linear accelerator QC device.

Authors:  R J Watts
Journal:  Med Phys       Date:  1998-02       Impact factor: 4.071

5.  Rapid acceptance testing of modern linac using on-board MV and kV imaging systems.

Authors:  Sridhar Yaddanapudi; Bin Cai; Taylor Harry; Steven Dolly; Baozhou Sun; Hua Li; Keith Stinson; Camille Noel; Lakshmi Santanam; Todd Pawlicki; Sasa Mutic; S Murty Goddu
Journal:  Med Phys       Date:  2017-05-26       Impact factor: 4.071

6.  On-site audits to investigate the quality of radiation physics of radiation therapy institutions in the Republic of Korea.

Authors:  Jong Min Park; So-Yeon Park; Minsoo Chun; Sang-Tae Kim
Journal:  Phys Med       Date:  2017-08-04       Impact factor: 2.685

7.  Changes in electron beam dosimetry with a new scattering foil-applicator system on a CL2100C.

Authors:  E E Klein; D A Low; J A Purdy
Journal:  Int J Radiat Oncol Biol Phys       Date:  1995-05-15       Impact factor: 7.038

8.  Monitoring Daily QA 3 constancy for routine quality assurance on linear accelerators.

Authors:  Diana Binny; Craig M Lancaster; Tanya Kairn; Jamie V Trapp; Scott B Crowe
Journal:  Phys Med       Date:  2016-11-10       Impact factor: 2.685

9.  Commissioning an Elekta Versa HD linear accelerator.

Authors:  Ganesh Narayanasamy; Daniel Saenz; Wilbert Cruz; Chul S Ha; Niko Papanikolaou; Sotirios Stathakis
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

10.  Visual Analysis of the Daily QA Results of Photon and Electron Beams of a Trilogy Linac over a Five-year Period.

Authors:  Maria F Chan; Qiongge Li; Xiaoli Tang; Xiang Li; Jingdong Li; Grace Tang; Margie A Hunt; Joseph O Deasy
Journal:  Int J Med Phys Clin Eng Radiat Oncol       Date:  2015-11-09
View more
  2 in total

1.  Nontarget and Out-of-Field Doses from Electron Beam Radiotherapy.

Authors:  Natalia Matuszak; Marta Kruszyna-Mochalska; Agnieszka Skrobala; Adam Ryczkowski; Piotr Romanski; Igor Piotrowski; Katarzyna Kulcenty; Wiktoria Maria Suchorska; Julian Malicki
Journal:  Life (Basel)       Date:  2022-06-08

2.  Evaluation of Surface Dose and Commissioning of Compensator-Based Total Body Irradiation.

Authors:  Bharath Pandu; D Khanna; P Mohandass; Hima Ninan; Rajadurai Elavarasan; Saro Jacob; Goutham Sunny
Journal:  J Med Phys       Date:  2022-08-05
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.