Development of an L-band resonator optimized for fast scan EPR imaging of the mouse head.

Alexandre Samouilov1, Denis Komarov1, Sergey Petryakov1, Arkadiy Iosilevich1, Jay L Zweier1.   

Abstract

PURPOSE: To develop a novel resonator for high-quality fast scan electron paramagnetic resonance (EPR) and EPR/NMR co-imaging of the head and brain of mice at 1.25 GHz.
METHODS: Resonator dimensions were scaled to fit the mouse head with maximum filling factor. A single-loop 6-gap resonator of 20 mm diameter and 20 mm length was constructed. High resonator stability was achieved utilizing a fixed position double coupling loop. Symmetrical mutually inverted connections rendered it insensitive to field modulation and fast scan. Coupling adjustment was provided by a parallel-connected variable capacitor located at the feeding line at λ/4 distance. To minimize radiation loss, the shield around the resonator was supplemented with a planar conductive disc that focuses return magnetic flux.
RESULTS: Coupling of the resonator loaded with the mouse head was efficient and easy. This resonator enabled high-quality in vivo 3D EPR imaging of the mouse head following intravenous infusion of nitroxide probes. With this resonator and rapid scan EPR system, 4 ms scans were acquired in forward and reverse directions so that images with 2-scan 3,136 projections were acquired in 25 s. Head images were achieved with resolutions of 0.4 mm, enabling visualization of probe localization and uptake across the blood-brain barrier.
CONCLUSIONS: This resonator design provides good sensitivity, high stability, and B1 field homogeneity for in vivo fast scan EPR of the mouse head and brain, enabling faster measurements and higher resolution imaging of probe uptake, localization, and metabolism than previously possible.
© 2021 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  electron paramagnetic resonance imaging; instrument development; microwave resonator; mouse head imaging; paramagnetic probes

Mesh:

Year:  2021        PMID: 33938574      PMCID: PMC8295191          DOI: 10.1002/mrm.28821

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   3.737


  43 in total

1.  Noninvasive measurement of anatomic structure and intraluminal oxygenation in the gastrointestinal tract of living mice with spatial and spectral EPR imaging.

Authors:  G He; R A Shankar; M Chzhan; A Samouilov; P Kuppusamy; J L Zweier
Journal:  Proc Natl Acad Sci U S A       Date:  1999-04-13       Impact factor: 11.205

2.  A bridged loop-gap S-band surface resonator for topical EPR spectroscopy.

Authors:  S Petryakov; M Chzhan; A Samouilov; G He; P Kuppusamy; J L Zweier
Journal:  J Magn Reson       Date:  2001-07       Impact factor: 2.229

3.  Frequency dependence of EPR signal intensity, 250 MHz to 9.1 GHz.

Authors:  George A Rinard; Richard W Quine; Sandra S Eaton; Gareth R Eaton
Journal:  J Magn Reson       Date:  2002-05       Impact factor: 2.229

4.  Evaluation of Mitochondrial Oxidative Stress in the Brain of a Transgenic Mouse Model of Alzheimer's Disease by in vitro Electron Paramagnetic Resonance Spectroscopy.

Authors:  Tatsuo Manabe; Akihiro Matsumura; Kazuki Yokokawa; Taro Saito; Mai Fujikura; Naotoshi Iwahara; Takashi Matsushita; Syuuichirou Suzuki; Shin Hisahara; Jun Kawamata; Hiromi Suzuki; Miho C Emoto; Hirotada G Fujii; Shun Shimohama
Journal:  J Alzheimers Dis       Date:  2019       Impact factor: 4.472

5.  A 750-MHz Electronically Tunable Resonator Using Microstrip Line Couplers for Electron Paramagnetic Resonance Imaging of a Mouse Tumor-Bearing Leg.

Authors:  Tatsuya Amida; Ririko Nakaoka; Denis A Komarov; Kumiko Yamamoto; Osamu Inanami; Shingo Matsumoto; Hiroshi Hirata
Journal:  IEEE Trans Biomed Eng       Date:  2017-08-23       Impact factor: 4.538

6.  Development of a fast-scan EPR imaging system for highly accelerated free radical imaging.

Authors:  Alexandre Samouilov; Rizwan Ahmad; James Boslett; Xiaoping Liu; Sergey Petryakov; Jay L Zweier
Journal:  Magn Reson Med       Date:  2019-04-25       Impact factor: 4.668

7.  Design and testing of a 750MHz CW-EPR digital console for small animal imaging.

Authors:  Hideo Sato-Akaba; Miho C Emoto; Hiroshi Hirata; Hirotada G Fujii
Journal:  J Magn Reson       Date:  2017-09-20       Impact factor: 2.229

8.  Comparative studies with EPR and MRI on the in vivo tissue redox status estimation using redox-sensitive nitroxyl probes: influence of the choice of the region of interest.

Authors:  Ken-Ichiro Matsumoto; James B Mitchell; Murali C Krishna
Journal:  Free Radic Res       Date:  2018-01-31

9.  In vivo measurement and imaging of tumor oxygenation using coembedded paramagnetic particulates.

Authors:  Govindasamy Ilangovan; Anna Bratasz; Haiquan Li; Petra Schmalbrock; Jay L Zweier; Periannan Kuppusamy
Journal:  Magn Reson Med       Date:  2004-09       Impact factor: 4.668

10.  Nitroxide free radical clearance in the live rat monitored by radio-frequency CW-EPR and PEDRI.

Authors:  M Alecci; I Seimenis; S J McCallum; D J Lurie; M A Foster
Journal:  Phys Med Biol       Date:  1998-07       Impact factor: 3.609

View more
  1 in total

1.  High fidelity triangular sweep of the magnetic field for millisecond scan EPR imaging.

Authors:  Denis A Komarov; Alexandre Samouilov; Hiroshi Hirata; Jay L Zweier
Journal:  J Magn Reson       Date:  2021-06-09       Impact factor: 2.734

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.