| Literature DB >> 33937389 |
Gordon B Drummond1, Darius Fischer2, Margaret Lees3, Andrew Bates2, Janek Mann2, D K Arvind2.
Abstract
BACKGROUND: Automatic measurement of respiratory rate in general hospital patients is difficult. Patient movement degrades the signal and variation of the breathing cycle means that accurate observation for ≥60 s is needed for adequate precision.Entities:
Year: 2021 PMID: 33937389 PMCID: PMC8071973 DOI: 10.1183/23120541.00681-2020
Source DB: PubMed Journal: ERJ Open Res ISSN: 2312-0541
Initial features considered for testing in the model
| Normalised as | |
| Extreme values are less likely to be reliable | |
| A large range of rates within the sampling window would be less trustworthy | |
| Mean and median of frequency differences between one breath and the next within the window | |
| A measure of large changes in axis signal, often associated with posture changes | |
| Activity level is computed as the length of the vector differences using the | |
| A large standard deviation indicates probable posture change | |
| A small number of detected breaths suggests the signal is unreliable | |
| Greater variation in the size of breaths in a window suggests unstable breathing | |
| Large differences between breath amplitudes suggest unstable breathing | |
| Times between successive peaks in the window should be similar |
These are derived from the waveform of the Respeck signal, the accelerometer signals and the breath timing measurements.
Windows used in the study
| 21 055 | 2223 | |
| 603 (3) | 29 (1) | |
| 618 (3) | 67 (3) | |
| 4844 (23) | 363 (16) | |
| 575 (3) | 94 (4) | |
| 14 415 (68) | 1670 (75) | |
Data are presented as n (%) unless otherwise stated.
FIGURE 1An example of a section of recording from a single subject. a) Nasal cannula pressure. A downward deflection (pressure decrease) indicates inspiration. b) Accelerometer forces. These are smoothed and combined into the Respeck signal. The arrow indicates the start of a patient movement, seen in one of the axis measures, that causes a large deflection of the Respeck signal. +ve: positive; −ve: negative.
FIGURE 2Effects of applying selection process to the test data. The six test patients are identified by letters, A–F. The test patients are ordered according to proportion of retained measurement windows. Patient G has a low proportion of windows retained.
FIGURE 3Bland and Altman plots of nasal−Respeck differences a) before and b) after selection of suitable signals, plotted in relation to the mean respiratory rate . The data are from all six validation subjects: each dot indicates one pair of compared values, nasal and Respeck measures, for a single observation window in one validation subject's data. Plots show bias (solid line) and limits of agreement (outer dashed lines) surrounded by their 95% confidence ranges (shaded). To allow comparable scales to be used in both panels, 31 of the 1670 data points from the “before selection” panel are not plotted.
FIGURE 4Plots of rate estimates from two representative patients shown for a 17-min time period (patient records are from A and G in the test sample, see figure 2). These are patients with extreme proportions of measurement windows that were chosen as accurate, from the whole window dataset for that patient. Even with a 10% acceptance rate, in patient G, these samples are sufficient to give a reasonable measure of the overall respiratory rate over the time period shown.