Literature DB >> 33922239

Aortic Annular Sizing Using Novel Software in Three-Dimensional Transesophageal Echocardiography for Transcatheter Aortic Valve Replacement: A Systematic Review and Meta-Analysis.

Chanrith Mork1, Minjie Wei1, Weixi Jiang1,2, Jianli Ren1,2, Haitao Ran1,2.   

Abstract

(1) Background: We performed this study to evaluate the agreement between novel automated software of three-dimensional transesophageal echocardiography (3D-TEE) and multidetector computed tomography (MDCT) for aortic annular measurements of preprocedural transcatheter aortic valve replacement (TAVR); (2)
Methods: PubMed, EMBASE, Web of Science, and Cochrane Library (Wiley) databases were systematically searched for studies that compared 3D-TEE and MDCT as the reference standard for aortic annular measurement of the following parameters: annular area, annular perimeter, area derived-diameter, perimeter derived-diameter, maximum and minimum diameter. Meta-analytic methods were utilized to determine the pooled correlations and mean differences between 3D-TEE and MDCT. Heterogeneity and publication bias were also assessed. Meta-regression analyses were performed based on the potential factors affecting the correlation of aortic annular area; (3)
Results: A total of 889 patients from 10 studies were included in the meta-analysis. Pooled correlation coefficients between 3D-TEE and MDCT of annulus area, perimeter, area derived-diameter, perimeter derived-diameter, maximum and minimum diameter measurements were strong 0.89 (95% CI: 0.84-0.92), 0.88 (95% CI: 0.83-0.92), 0.87 (95% CI: 0.77-0.93), 0.87 (95% CI: 0.77-0.93), 0.79 (95% CI: 0.64-0.87), and 0.75 (95% CI: 0.61-0.84) (Overall p < 0.0001), respectively. Pooled mean differences between 3D-TEE and MDCT of annulus area, perimeter, area derived-diameter, perimeter derived-diameter, maximum and minimum diameter measurements were -20.01 mm2 ((95% CI: -35.37 to -0.64), p = 0.011), -2.31 mm ((95% CI: -3.31 to -1.31), p < 0.0001), -0.22 mm ((95% CI: -0.73 to 0.29), p = 0.40), -0.47 mm ((95% CI: -1.06 to 0.12), p = 0.12), -1.36 mm ((95% CI: -2.43 to -0.30), p = 0.012), and 0.31 mm ((95% CI: -0.15 to 0.77), p = 0.18), respectively. There were no statistically significant associations with the baseline patient characteristics of sex, age, left ventricular ejection fraction, mean transaortic gradient, and aortic valve area to the correlation between 3D-TEE and MDCT for aortic annular area sizing; (4) Conclusions: The present study implies that 3D-TEE using novel software tools, automatically analysis, is feasible to MDCT for annulus sizing in clinical practice.

Entities:  

Keywords:  multidetector computed tomography; novel software; three-dimensional transesophageal echocardiography; transcatheter aortic valve replacement

Year:  2021        PMID: 33922239     DOI: 10.3390/diagnostics11050751

Source DB:  PubMed          Journal:  Diagnostics (Basel)        ISSN: 2075-4418


  66 in total

Review 1.  Multidetector computed tomography in transcatheter aortic valve implantation.

Authors:  Jonathon Leipsic; Ronen Gurvitch; Troy M Labounty; James K Min; David Wood; Mark Johnson; Amr M Ajlan; Namal Wijesinghe; John G Webb
Journal:  JACC Cardiovasc Imaging       Date:  2011-04

Review 2.  Automated Quantification in Echocardiography.

Authors:  Mark T Nolan; Paaladinesh Thavendiranathan
Journal:  JACC Cardiovasc Imaging       Date:  2019-06

3.  2017 ACC Expert Consensus Decision Pathway for Transcatheter Aortic Valve Replacement in the Management of Adults With Aortic Stenosis: A Report of the American College of Cardiology Task Force on Clinical Expert Consensus Documents.

Authors:  Catherine M Otto; Dharam J Kumbhani; Karen P Alexander; John H Calhoon; Milind Y Desai; Sanjay Kaul; James C Lee; Carlos E Ruiz; Christina M Vassileva
Journal:  J Am Coll Cardiol       Date:  2017-01-04       Impact factor: 24.094

4.  Three-dimensional transesophageal echocardiography is an attractive alternative to cardiac multi-detector computed tomography for aortic annular sizing: Systematic review and meta-analysis.

Authors:  Ahmed Elkaryoni; Navin C Nanda; Paramdeep Baweja; Mohammed J Arisha; Harris Zamir; Ahmed Elgebaly; Ahmed Ma Altibi; Rishi Sharma
Journal:  Echocardiography       Date:  2018-10       Impact factor: 1.724

5.  Transcatheter Aortic-Valve Replacement with a Balloon-Expandable Valve in Low-Risk Patients.

Authors:  Michael J Mack; Martin B Leon; Vinod H Thourani; Raj Makkar; Susheel K Kodali; Mark Russo; Samir R Kapadia; S Chris Malaisrie; David J Cohen; Philippe Pibarot; Jonathon Leipsic; Rebecca T Hahn; Philipp Blanke; Mathew R Williams; James M McCabe; David L Brown; Vasilis Babaliaros; Scott Goldman; Wilson Y Szeto; Philippe Genereux; Ashish Pershad; Stuart J Pocock; Maria C Alu; John G Webb; Craig R Smith
Journal:  N Engl J Med       Date:  2019-03-16       Impact factor: 91.245

6.  Measurement of the Aortic Annulus Area and Diameter by Three-Dimensional Transesophageal Echocardiography in Transcatheter Aortic Valve Replacement.

Authors:  Keigo Ebuchi; Kenji Yoshitani; Eiki Kanemaru; Tasuku Fujii; Akito Tsukinaga; Yusuke Shimahara; Yoshiniko Ohnishi
Journal:  J Cardiothorac Vasc Anesth       Date:  2019-04-30       Impact factor: 2.628

7.  Aortic roots assessment by an automated three-dimensional transesophageal echocardiography: an intra-individual comparison.

Authors:  Minghui Zhang; Linyuan Wan; Kun Liu; Weichun Wu; Hui Li; Yuan Wang; Bin Lu; Hao Wang
Journal:  Int J Cardiovasc Imaging       Date:  2019-07-11       Impact factor: 2.357

8.  5-year outcomes of transcatheter aortic valve replacement or surgical aortic valve replacement for high surgical risk patients with aortic stenosis (PARTNER 1): a randomised controlled trial.

Authors:  Michael J Mack; Martin B Leon; Craig R Smith; D Craig Miller; Jeffrey W Moses; E Murat Tuzcu; John G Webb; Pamela S Douglas; William N Anderson; Eugene H Blackstone; Susheel K Kodali; Raj R Makkar; Gregory P Fontana; Samir Kapadia; Joseph Bavaria; Rebecca T Hahn; Vinod H Thourani; Vasilis Babaliaros; Augusto Pichard; Howard C Herrmann; David L Brown; Mathew Williams; Jodi Akin; Michael J Davidson; Lars G Svensson
Journal:  Lancet       Date:  2015-03-15       Impact factor: 79.321

9.  Pre-procedural assessment of aortic annulus dimensions for transcatheter aortic valve replacement: comparison of a non-contrast 3D MRA protocol with contrast-enhanced cardiac dual-source CT angiography.

Authors:  Philipp Ruile; Philipp Blanke; Tobias Krauss; Stephan Dorfs; Bernd Jung; Nikolaus Jander; Jonathon Leipsic; Mathias Langer; Franz-Josef Neumann; Gregor Pache
Journal:  Eur Heart J Cardiovasc Imaging       Date:  2015-07-27       Impact factor: 6.875

10.  2020 ACC/AHA Guideline for the Management of Patients With Valvular Heart Disease: A Report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines.

Authors:  Catherine M Otto; Rick A Nishimura; Robert O Bonow; Blase A Carabello; John P Erwin; Federico Gentile; Hani Jneid; Eric V Krieger; Michael Mack; Christopher McLeod; Patrick T O'Gara; Vera H Rigolin; Thoralf M Sundt; Annemarie Thompson; Christopher Toly
Journal:  J Am Coll Cardiol       Date:  2020-12-17       Impact factor: 24.094

View more
  1 in total

Review 1.  Preoperative TAVR Planning: How to Do It.

Authors:  Rodrigo Petersen Saadi; Ana Paula Tagliari; Eduardo Keller Saadi; Marcelo Haertel Miglioranza; Carisi Anne Polanczyck
Journal:  J Clin Med       Date:  2022-05-05       Impact factor: 4.964

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.