Keigo Ebuchi1, Kenji Yoshitani2, Eiki Kanemaru1, Tasuku Fujii1, Akito Tsukinaga1, Yusuke Shimahara3, Yoshiniko Ohnishi1. 1. Department of Anesthesiology, National Cerebral and Cardiovascular Center, Osaka, Japan. 2. Department of Anesthesiology, National Cerebral and Cardiovascular Center, Osaka, Japan. Electronic address: ykenji@kfz.biglobe.ne.jp. 3. Department of Cardiovascular Surgery, National Cerebral and Cardiovascular Center, Osaka, Japan.
Abstract
OBJECTIVES: Sizing of the aortic valve is crucial for transcatheter aortic valve replacement (TAVR). Multidetector computed tomography (MDCT) is used for sizing. Recently, three-dimensional transesophageal echocardiography (3DTEE) has enabled accurate measurement of the aortic annulus area and diameter in cases that are difficult to measure. The authors compared measurements of aortic annulus areas and diameters acquired by MDCT and 3DTEE. DESIGN: Retrospective observational study. SETTING: Single national center. PARTICIPANTS: Sixty-eight patients who underwent TAVR replacement between September 2015 and March 2017. INTERVENTIONS: None. MEASUREMENTS AND MAIN RESULTS: The authors extracted and compared preoperative measurements of the aortic annulus area, as well as the long- and short-axis diameter, measured by MDCT and 3DTEE. There was no significant difference in the aortic annulus area (409 ± 74 v 414 ± 70 mm2, p = 0.15) or short-axis diameter (20.4 ± 2.0 v 20.6 ± 1.9 mm, p = 0.103) between 3DTEE and MDCT, but the long-axis diameter differed significantly (25.0 ± 2.4 v 25.8 ± 2.0 mm, p < 0.001), respectively. Prosthesis sizes based on 3DTEE and MDCT were the same, except in 3 patients who could not stay still during MDCT measurement; in those cases, prosthesis sizes based on 3DTEE were adopted. CONCLUSIONS: Measurements of the aortic annulus area and diameter in TAVR were similar between 3DTEE and MDCT. Patients who have difficulty remaining still during MDCT measurement because of dementia should have their prostheses sized based on 3DTEE measurements.
OBJECTIVES: Sizing of the aortic valve is crucial for transcatheter aortic valve replacement (TAVR). Multidetector computed tomography (MDCT) is used for sizing. Recently, three-dimensional transesophageal echocardiography (3DTEE) has enabled accurate measurement of the aortic annulus area and diameter in cases that are difficult to measure. The authors compared measurements of aortic annulus areas and diameters acquired by MDCT and 3DTEE. DESIGN: Retrospective observational study. SETTING: Single national center. PARTICIPANTS: Sixty-eight patients who underwent TAVR replacement between September 2015 and March 2017. INTERVENTIONS: None. MEASUREMENTS AND MAIN RESULTS: The authors extracted and compared preoperative measurements of the aortic annulus area, as well as the long- and short-axis diameter, measured by MDCT and 3DTEE. There was no significant difference in the aortic annulus area (409 ± 74 v 414 ± 70 mm2, p = 0.15) or short-axis diameter (20.4 ± 2.0 v 20.6 ± 1.9 mm, p = 0.103) between 3DTEE and MDCT, but the long-axis diameter differed significantly (25.0 ± 2.4 v 25.8 ± 2.0 mm, p < 0.001), respectively. Prosthesis sizes based on 3DTEE and MDCT were the same, except in 3 patients who could not stay still during MDCT measurement; in those cases, prosthesis sizes based on 3DTEE were adopted. CONCLUSIONS: Measurements of the aortic annulus area and diameter in TAVR were similar between 3DTEE and MDCT. Patients who have difficulty remaining still during MDCT measurement because of dementia should have their prostheses sized based on 3DTEE measurements.