Literature DB >> 30661246

Accuracy of detecting vertical root fractures in non-root filled teeth using cone beam computed tomography: effect of voxel size and fracture width.

X L Guo1, G Li1, J Q Zheng2, R H Ma1, F C Liu3, F S Yuan2, P J Lyu2, Y J Guo4, S Yin1.   

Abstract

AIM: To investigate simultaneously the effect of voxel size and fracture width on the accuracy of detecting vertical root fractures (VRFs) in non-root filled teeth when using cone beam computed tomography.
METHODOLOGY: Fifty-one of 161 extracted human permanent teeth (16 anterior teeth, 132 premolars and 13 mandibular molars) were selected randomly for VRF induction with two fracture widths. All teeth were scanned with four CBCT units at different voxel sizes provided by the units. Three observers classified the presence or absence of VRF using a 5-point scale. Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV) and area under the ROC curve (AUC) were calculated. AUCs amongst voxel sizes and between the fracture widths were compared using the Z test. Intra- and inter-observer agreement was assessed using weighted Cohen kappa.
RESULTS: For the NewTom VGi and ProMax 3D Mid CBCT unit, no significant differences were found amongst voxel sizes for the AUCs, irrespective of the fracture width (P > 0.05). There were significant differences between images scanned with voxel size 250 and 160 μm (P = 0.02), and images scanned with voxel size 250 and 80 μm for AUCs in the narrow VRF group for the 3D Accuitomo 170 unit (P = 0.03). For i-CAT FLX, significant differences were found between the voxel protocols of 300 μm and of the other three voxel sizes for AUC, sensitivity and NPV (P < 0.05). Significant differences between the wide and the narrow VRF groups for AUCs were found for 3D Accuitomo 170 (P = 0.01) and ProMax 3D Mid (P < 0.01).
CONCLUSIONS: Cone beam computed tomography was accurate for detecting VRF in non-root filled teeth. Fracture width had an effect on the detection of VRF. The effect of the voxel size on the detection of VRF depended on the CBCT unit used.
© 2019 International Endodontic Journal. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  area under the ROC curve; cone beam computed tomography; fracture width; non-root filled teeth; vertical root fracture; voxel size

Mesh:

Year:  2019        PMID: 30661246     DOI: 10.1111/iej.13076

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  7 in total

1.  Determination of a cone-beam CT low-dose protocol for root fracture diagnosis in non-endodontically treated anterior maxillary teeth.

Authors:  Fedil Andraws Yalda; Chrysoula Theodorakou; Rosalyn J Clarkson; Jonathan Davies; Lee Feinberg; Anita Sengupta; Keith Horner
Journal:  Dentomaxillofac Radiol       Date:  2021-09-08       Impact factor: 2.419

2.  Scatter-to-primary ratio in dentomaxillofacial cone-beam CT: effect of field of view and beam energy.

Authors:  Ruben Pauwels; Pisha Pittayapat; Phonkit Sinpitaksakul; Soontra Panmekiate
Journal:  Dentomaxillofac Radiol       Date:  2021-04-29       Impact factor: 2.419

3.  In vivo detection of vertical root fractures in endodontically treated teeth: Accuracy of cone-beam computed tomography and assessment of potential predictor variables.

Authors:  Marcela Quintero-Álvarez; Leslie-Melissa Bolaños-Alzate; Paula-Andrea Villa-Machado; Felipe-Augusto Restrepo-Restrepo; Sergio-Iván Tobón-Arroyave
Journal:  J Clin Exp Dent       Date:  2021-02-01

4.  An investigation into dose optimisation for imaging root canal anatomy using cone beam CT.

Authors:  Margarete B McGuigan; Christie Theodorakou; Henry F Duncan; Jonathan Davies; Anita Sengupta; Keith Horner
Journal:  Dentomaxillofac Radiol       Date:  2020-06-22       Impact factor: 2.419

5.  Influence of size of field of view (FOV), position within the FOV, and scanning mode on the detection of root fracture and observer's perception of artifacts in CBCT images.

Authors:  Martina Gerlane de Oliveira Pinto; Saulo L Sousa Melo; Fernanda Clotilde Mariz Suassuna; Luiz Eduardo Marinho; José Bruno da Silva Leite; Andre Ulisses Dantas Batista; Patrícia Meira Bento; Daniela Pita Melo
Journal:  Dentomaxillofac Radiol       Date:  2021-02-10       Impact factor: 3.525

Review 6.  Diagnosis of cracked teeth using cone-beam computed tomography: literature review and clinical experience.

Authors:  Antian Gao; Dantong Cao; Zitong Lin
Journal:  Dentomaxillofac Radiol       Date:  2021-04-29       Impact factor: 3.525

7.  Establishment of Narrow Root Fracture Models Using Modified Temperature Cycling Method and Diagnosis Using Different Cone-Beam CT Units.

Authors:  Jiahao Liang; Ziyang Hu; Dantong Cao; Ya Cao; Xin Xie; Antian Gao; Zhiyong Wang; Zitong Lin
Journal:  Scanning       Date:  2022-07-14       Impact factor: 1.750

  7 in total