Characteristics of x-ray beams in two commercial multidetector computed tomography simulators: Monte Carlo simulations.

Sangroh Kim1, Haijun Song, Benjamin Movsas, Indrin J Chetty.   

Abstract

PURPOSE: As multidetector computed tomography (MDCT) scanning is routinely performed for treatment planning in radiation oncology, understanding the characteristics of the MDCT x-ray beam is essential to accurately estimate patient dose. The purpose of this study is to characterize the x-ray beams of two commercial MDCT simulators widely used in radiation oncology by Monte Carlo (MC) simulations.
METHODS: X-ray tube systems of two wide bore MDCT scanners (GE LightSpeed RT 4 and Philips Brilliance Big Bore) were modeled in the BEAMNRC/EGSNRC MC system. All the tube components were modeled from targets to bowtie filters. To validate our MC models, the authors measured half-value layers (HVL) using aluminum sheets and multifunctional radiation detectors and compared them to those obtained from MC simulations for 120 kVp beams. The authors also compared x-ray spectra obtained from MC simulation to the data provided by manufacturers. Additionally, lateral/axial beam profiles were measured in-air using radiochromic films and compared to the MC results. To understand the scatter effect, the authors also derived the scatter-to-primary energy fluence ratio (SPR) profiles and calculated the total SPR for each CT system with the CT dose index (CTDI) head and body phantoms using the BEAMNRC system.
RESULTS: The authors found that the HVL, x-ray spectrum and beam profiles of the MC simulations agreed well with the manufacturer-specified data within 1%-10% on average for both scanners. The total SPR were ranged from 7.8 to 13.7% for the head phantom and from 10.7 to 18.9% for the body phantom.
CONCLUSIONS: The authors demonstrate the full MC simulations of two commercial MDCT simulators to characterize their x-ray beams. This study may be useful to establish a patient-specific dosimetry for the MDCT systems.

Entities:  

Mesh:

Year:  2012        PMID: 22225302     DOI: 10.1118/1.3670377

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Monte Carlo calculation of patient organ doses from computed tomography.

Authors:  Takeshi Oono; Fujio Araki; Shoya Tsuduki; Keiichi Kawasaki
Journal:  Radiol Phys Technol       Date:  2013-11-30

2.  A measurement-based generalized source model for Monte Carlo dose simulations of CT scans.

Authors:  Xin Ming; Yuanming Feng; Ransheng Liu; Chengwen Yang; Li Zhou; Hezheng Zhai; Jun Deng
Journal:  Phys Med Biol       Date:  2017-01-12       Impact factor: 3.609

3.  Scatter-to-primary ratio in dentomaxillofacial cone-beam CT: effect of field of view and beam energy.

Authors:  Ruben Pauwels; Pisha Pittayapat; Phonkit Sinpitaksakul; Soontra Panmekiate
Journal:  Dentomaxillofac Radiol       Date:  2021-04-29       Impact factor: 2.419

4.  A measurement-based X-ray source model characterization for CT dosimetry computations.

Authors:  Mitchell Sommerville; Yannick Poirier; Mauro Tambasco
Journal:  J Appl Clin Med Phys       Date:  2015-11-08       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.