Multi-analyte calibration and verification of a multi-parameter laser-based pulse oximeter.

Jonas A Pologe1, Donald H Arnold2, Theodore P Delianides3.   

Abstract

Almost since its introduction pulse oximetry was known to overestimate oxygen saturation in cases of carbon monoxide poisoning or elevated methemoglobin (metHb) levels. To eliminate this dangerous behavior some manufacturers have added additional LED emitters to try to increase the number of measured hemoglobin species and to improve measurement accuracy, but have not been very successful. We hypothesized that the use of narrow-band laser light sources would make accurate and precise measurement of the four primary species of hemoglobin possible, even in cases of elevated levels of carboxyhemoglobin (COHb). Calibration and verification studies were performed on a tissue simulator that employed an artificial finger pulsating with whole human blood. This simulator allowed safe generation of 165 different combinations of the levels of oxyhemoglobin (O2Hb), COHb, metHb, and reduced hemoglobin (RHb) for calibration of the laser-based pulse oximeter. A follow-on study used 56 mixed hemoglobin levels for verification and statistical analysis of the performance of this device. This laser-based pulse oximeter measured all four species of hemoglobin accurately and precisely (ARMS ≤ 1.8%) for metHb levels in the clinically normal range. At elevated metHb levels the device continued to provide accurate and precise measurements of metHb and RHb (ARMS ≤ 1.7%). The use of monochromatic laser light sources can create a new generation of highly accurate, multi-parameter, pulse oximeters.
© 2021. The Author(s), under exclusive licence to Springer Nature B.V.

Entities:  

Keywords:  Accuracy; Laser; Monitoring; Photoplethysmography; Pulse oximetry

Mesh:

Substances:

Year:  2021        PMID: 33871764     DOI: 10.1007/s10877-021-00704-1

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  13 in total

1.  Non-invasive carbon monoxide measurement is not accurate.

Authors:  Gerald F O'Malley
Journal:  Ann Emerg Med       Date:  2006-10       Impact factor: 5.721

2.  Carboxyhemoglobin measurement by hospitals: implications for the diagnosis of carbon monoxide poisoning.

Authors:  Neil B Hampson; Karen L Scott; Jennette L Zmaeff
Journal:  J Emerg Med       Date:  2006-07       Impact factor: 1.484

3.  Carbon monoxide intoxication versus myocardial infarction: An easy diagnosis?

Authors:  Gerrit Grieb; David Simons; Andrzej Piatkowski; Ertunc Altiok; Ron J Eppstein; Jürgen Bernhagen; Norbert Pallua
Journal:  Burns       Date:  2011-02-22       Impact factor: 2.744

4.  Clinical Policy: Critical Issues in the Evaluation and Management of Adult Patients Presenting to the Emergency Department With Acute Carbon Monoxide Poisoning.

Authors:  Stephen J Wolf; Gerald E Maloney; Richard D Shih; Bradley D Shy; Michael D Brown
Journal:  Ann Emerg Med       Date:  2017-01       Impact factor: 5.721

5.  Effects of methemoglobinemia on pulse oximetry and mixed venous oximetry.

Authors:  S J Barker; K K Tremper; J Hyatt
Journal:  Anesthesiology       Date:  1989-01       Impact factor: 7.892

6.  Response to: Carbon monoxide poisoning cases presenting with non-specific symptoms by Deniz. Accuracy of triage with RAD-57 pulse CO-oximeter in suspected collective intoxication of a pediatric population.

Authors:  Jacques Bessereau; F Heireche; V Bremond; A Boutin; S Petit; M Tsapis; L Boiron; A Puget; P Garry; M Coulange; F Kerbaul
Journal:  Toxicol Ind Health       Date:  2018-05       Impact factor: 2.273

7.  Performance of the RAD-57 pulse CO-oximeter compared with standard laboratory carboxyhemoglobin measurement.

Authors:  Michael Touger; Adrienne Birnbaum; Jessica Wang; Katherine Chou; Darion Pearson; Polly Bijur
Journal:  Ann Emerg Med       Date:  2010-06-03       Impact factor: 5.721

8.  Unrecognized carbon monoxide poisoning leads to a multiple-casualty incident.

Authors:  Dominik Roth; Mario Krammel; Wolfgang Schreiber; Harald Herkner; Christof Havel; Anton N Laggner
Journal:  J Emerg Med       Date:  2013-06-27       Impact factor: 1.484

9.  The effect of carbon monoxide inhalation on pulse oximetry and transcutaneous PO2.

Authors:  S J Barker; K K Tremper
Journal:  Anesthesiology       Date:  1987-05       Impact factor: 7.892

10.  Accidental carbon monoxide poisoning presenting without a history of exposure: a case report.

Authors:  Luke Bennetto; Louise Powter; Neil J Scolding
Journal:  J Med Case Rep       Date:  2008-04-22
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.