Performance of the RAD-57 pulse CO-oximeter compared with standard laboratory carboxyhemoglobin measurement.

Michael Touger1, Adrienne Birnbaum, Jessica Wang, Katherine Chou, Darion Pearson, Polly Bijur.   

Abstract

STUDY
OBJECTIVE: We assess agreement between carboxyhemoglobin levels measured by the Rad-57 signal extraction pulse CO-oximeter (RAD), a Food and Drug Administration-approved device for noninvasive bedside measurement, and standard laboratory arterial or venous measurement in a sample of emergency department (ED) patients with suspected carbon monoxide poisoning.
METHODS: The study was a cross-sectional cohort design using a convenience sample of adult and pediatric ED patients in a Level I trauma, burn, and hyperbaric oxygen referral center. Measurement of RAD carboxyhemoglobin was performed simultaneously with blood sampling for laboratory determination of carboxyhemoglobin level. The difference between the measures for each patient was calculated as laboratory carboxyhemoglobin minus carboxyhemoglobin from the carbon monoxide oximeter. The limits of agreement from a Bland-Altman analysis are calculated as the mean of the differences between methods ±1.96 SDs above and below the mean.
RESULTS: Median laboratory percentage carboxyhemoglobin level was 2.3% (interquartile range 1 to 8.5; range 0% to 38%). The mean difference between laboratory carboxyhemoglobin values and RAD values was 1.4% carboxyhemoglobin (95% confidence interval [CI] 0.2% to 2.6%). The limits of agreement of differences of measurement made with the 2 devices were -11.6% and 14.4% carboxyhemoglobin. This range exceeded the value of ±5% carboxyhemoglobin defined a priori as clinically acceptable. RAD correctly identified 11 of 23 patients with laboratory values greater than 15% carboxyhemoglobin (sensitivity 48%; 95% CI 27% to 69%). There was one case of a laboratory carboxyhemoglobin level less than 15%, in which the RAD device gave a result greater than 15% (specificity of RAD 96/97=99%; 95% CI 94% to 100%).
CONCLUSION: In the range of carboxyhemoglobin values measured in this sample, the level of agreement observed suggests RAD measurement may not be used interchangeably with standard laboratory measurement.
Copyright © 2010 American College of Emergency Physicians. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20605259     DOI: 10.1016/j.annemergmed.2010.03.041

Source DB:  PubMed          Journal:  Ann Emerg Med        ISSN: 0196-0644            Impact factor:   5.721


  15 in total

Review 1.  The Diagnosis and Treatment of Carbon Monoxide Poisoning.

Authors:  Lars Eichhorn; Marcus Thudium; Björn Jüttner
Journal:  Dtsch Arztebl Int       Date:  2018-12-24       Impact factor: 5.594

Review 2.  Anesthesia-Related Carbon Monoxide Exposure: Toxicity and Potential Therapy.

Authors:  Richard J Levy
Journal:  Anesth Analg       Date:  2016-09       Impact factor: 5.108

Review 3.  [Carbon monoxide intoxication-New aspects and current guideline-based recommendations].

Authors:  A Fichtner; L Eichhorn
Journal:  Anaesthesiologie       Date:  2022-06-23

4.  Accuracy of carboxyhemoglobin detection by pulse CO-oximetry during hypoxemia.

Authors:  John R Feiner; Mark D Rollins; Jeffrey W Sall; Helge Eilers; Paul Au; Philip E Bickler
Journal:  Anesth Analg       Date:  2013-03-11       Impact factor: 5.108

5.  Screening by pulse CO-oximetry for environmental tobacco smoke exposure in preanesthetic children.

Authors:  Kathryn Cardwell; Zhaoxing Pan; Rebecca Boucher; Jeannie Zuk; Robert H Friesen
Journal:  Paediatr Anaesth       Date:  2012-05-15       Impact factor: 2.556

6.  Multi-analyte calibration and verification of a multi-parameter laser-based pulse oximeter.

Authors:  Jonas A Pologe; Donald H Arnold; Theodore P Delianides
Journal:  J Clin Monit Comput       Date:  2021-04-19       Impact factor: 2.502

Review 7.  Pulse oximetry.

Authors:  Amal Jubran
Journal:  Crit Care       Date:  2015-07-16       Impact factor: 9.097

8.  Utility of the Measurement of Carboxyhemoglobin Level at the Site of Acute Carbon Monoxide Poisoning in Rural Areas.

Authors:  Makoto Onodera; Yasuhisa Fujino; Satoshi Kikuchi; Masayuki Sato; Kiyofumi Mori; Takaaki Beppu; Yoshihiro Inoue
Journal:  Scientifica (Cairo)       Date:  2016-04-30

Review 9.  Pulse oximetry: fundamentals and technology update.

Authors:  Meir Nitzan; Ayal Romem; Robert Koppel
Journal:  Med Devices (Auckl)       Date:  2014-07-08

10.  Methaemoglobin and COHb in patients with malaria.

Authors:  Thomas Hänscheid; Tom Gresnigt; Sascha Löhr; Arnaud Flamen; Thomas Zoller; José Melo-Cristino; Martin P Grobusch
Journal:  Malar J       Date:  2014-07-23       Impact factor: 2.979