Apoorva Mittal1, Shalini Verma2, Gopishankar Natanasabapathi3, Pratik Kumar1, Akhilesh K Verma2. 1. Department of Medical Physics, Dr. B. R. A. Institute Rotary Cancer Hospital, All India Institute of Medical Sciences, New Delhi 110029, India. 2. Department of Chemistry, University of Delhi, Delhi 110007, India. 3. Department of Radiotherapy, Dr. B. R. A. Institute Rotary Cancer Hospital, All India Institute of Medical Sciences, New Delhi 110029, India.
Abstract
Blood and its cellular components are irradiated by ionizing radiation before transfusion to prevent the proliferation of viable T lymphocytes which cause transfusion associated-graft versus host disease. The immunodeficient patients undergoing chemotherapy for various malignancies are at risk of this disease. The international guidelines for blood transfusion recommend a minimum radiation exposure of 25 Gray (Gy) to the midplane of the blood bag, while a minimum dose of 15 Gy and a maximum dose of 50 Gy should be given to each portion of the blood bag. Therefore, precise dosimetry of the blood irradiator is essential to ensure the adequate irradiation of the blood components. The paper presents the fabrication of diacetylene-based colorimetric film dosimeters for the verification of irradiated doses. The diacetylene analogues are synthesized by tailoring them with different amide-based headgroups followed by their coating to develop colorimetric film dosimeters. Among all the synthesized diacetylene analogues, aminofluorene-substituted diacetylene exhibits the most significant color transition from white to blue color at a minimum γ radiation dose of 5 Gy. The quantitative study of color change is performed by the digitization of the scanned images of film dosimeters. The digital image processing of the developed film dosimeters facilitates rapid dose measurement which enables their facile implementation and promising application in routine blood irradiator dosimetry.
Blood and its cellular components are irradiated by ionizing radiation before transfusion to prevent the proliferation of viable T lymphocytes which cause transfusion associated-graft versus host disease. The immunodeficientpatients undergoing chemotherapy for various malignancies are at risk of this disease. The international guidelines for blood transfusion recommend a minimum radiation exposure of 25 Gray (Gy) to the midplane of the blood bag, while a minimum dose of 15 Gy and a maximum dose of 50 Gy should be given to each portion of the blood bag. Therefore, precise dosimetry of the blood irradiator is essential to ensure the adequate irradiation of the blood components. The paper presents the fabrication of diacetylene-based colorimetric film dosimeters for the verification of irradiated doses. The diacetylene analogues are synthesized by tailoring them with different amide-based headgroups followed by their coating to develop colorimetric film dosimeters. Among all the synthesized diacetylene analogues, aminofluorene-substituted diacetylene exhibits the most significant color transition from white to blue color at a minimum γ radiation dose of 5 Gy. The quantitative study of color change is performed by the digitization of the scanned images of film dosimeters. The digital image processing of the developed film dosimeters facilitates rapid dose measurement which enables their facile implementation and promising application in routine blood irradiator dosimetry.
Cancerpatients undergoing
chemotherapy for the treatment of various
malignancies are given blood or blood components transfusion.[1−3] Because of their immune deficiency and treatment with aggressive
immunosuppressive agents these patients are at a high risk of developing
transfusion associated-graft versus host disease (TA-GvHD). TA-GvHD
is a rare complication of transfusion with a high mortality rate that
results from the engraftment of the residual T-lymphocytes present
in cellular blood components.[4−6] Currently, the only accepted methodology
to inhibit TA-GvHD is prophylactic irradiation of whole blood and
its cellular components by ionizing radiation like γ and X-rays.[1,7−9] Irradiation of blood bags is performed by using free-standing,
dedicated blood irradiators that emit γ rays or X-rays. The
irradiation of blood cellular components by ionizing radiation inactivates
proliferative T lymphocytes while leaving platelets, granulocytes,
erythrocytes, and other blood components functional and viable, which
in turn abrogates TA-GvHD. The blood bags are typically placed inside
the blood irradiator in a cylindrical canister that is positioned
on a rotating turntable (Figure S1). Although
the role of turntable rotation is to maximize uniform radiation dose
exposure over the entire blood bag placed inside the canister but
the irregular shape of the blood bag and the geometry of the radiation
field can cause an inevitable possibility of inhomogeneous radiation
exposure. According to the recommendations of World Health Organization
(WHO) and American Association of Blood Banks (AABB),[10,11] a dose of 25 Gray (Gy) should be given to the central area of the
blood bag with no portion of the bag receiving less than 15 Gy or
more than 50 Gy. Based on these guidelines, the regular verification
of dose and dose distribution throughout the canister of the blood
irradiators must be conducted with a suitable dosimeter to avoid over
or under radiation exposure of the blood bags. The radiation doses
must be measured with dosimeters that can be placed throughout the
canister under routine irradiation conditions with blood bags or equivalent
medium. Till date, dosimetry of blood irradiators using various commercially
available dosimeters like thermoluminescent dosimeters, diode detectors,
and gel-based dosimeters has been reported.[7,12−14] However, these dosimeters prove to be cumbersome
and labor-intensive as compared to films that can be cut into any
size and pasted on the blood bags to measure the absorbed radiation
doses. The ease of handling of the films allows the possibility of
obtaining relative radiation dose profiles along the different axes
of the cylindrical geometry of the blood irradiator canister. Therefore,
film dosimetry is relatively simple, efficient, and less time-consuming.
The commercially available Gafchromic film[15,16] (International Specialty Products, Inc., Wayne, NJ) has also been
used for this purpose but it has several limitations like UV sensitivity,
temperature sensitivity, and post irradiation instability. Also, its
high cost restricts its use in routine dosimetry practice in the blood
irradiator.[17−26] In this context, we have attempted to develop novel diacetylene
(DA)-based colorimetric film dosimeters for the measurement of radiation
dose and dose distribution in blood irradiation.Conjugated
systems like DAs are a promising class of colorimetric
materials.[27] The highly ordered backbones
of DAs with the potential to customize the side groups make them suitable
for numerous applications.[28] Upon exposure
to radiation, they undergo topochemical polymerization reaction resulting
in the formation of polydiacetylenes (PDAs). The existence of extensively
delocalized π-electron networks and conformational restrictions
present along the main chain causes a distinct color change because
of various stimuli.[29] There are specific
geometrical parameters and optimal packing orientation of the DA monomers,
which are responsible for the successful topochemical polymerization.[30] The interactions between the head groups and
side chains prominently influence the overall conformation of the
DA, thus causing optical changes. Therefore, the optical and colorimetric
properties of DAs can be altered to a great extent by the introduction
of various side groups.[31,32]The stimulus-induced
color transitions in DA derivatives have been
extensively investigated for the development of a variety of sensors.[33,34] However, the development of amide DA film dosimeters for blood irradiation
has not been explored yet. Accordingly, the rationale of our work
was to design a novel amide-modified DA-based film dosimeter viable
for routine blood dosimetry. Particularly, amide containing headgroups
were chosen because of the presence of an extra lone pair which is
responsible for the enhanced hydrogen bonding, which leads to improved
radiation response of the DA monomers. The results of the role of
amide head groups on the radiation-induced color transitions helped
in the choice of a suitable amide-substituted DA monomer for the development
of a potential film dosimeter for blood irradiator. The quantitative
study of color transition was performed by using a high-resolution
scanner. A comparative analysis of the radiation sensitivity of various
amide terminated DA derivatives was done by studying their optical
density (OD) and intensity profiles obtained by image digitization
of the scanned images by developing a color quantification algorithm
in MATLAB software. Measurement of the OD of all the exposed films
in a single scan enabled efficient dosimetry applications. The objective
of this work was to synthesize amide-substituted DA derivatives and
then to employ them as a coating for the preparation of films which
exhibit an unprecedented high colorimetric response to blood irradiation
doses from 5 to 50 Gy. The effect of pre- and post-irradiation storage
conditions on the developed film dosimeters was also investigated.
Results
and Discussion
Synthesis of DA Monomers
The commercially
available
DA monomer 10,12-pentacosadiynoic acid (PCDA) was chosen as the base
molecule. The high reactivity of the carboxylic acid head group present
in PCDA allows the substitution by various side groups. The DA monomers 1–4 (Figure A) used in this study were synthesized by a single-step conversion
of DA carboxylic acids into amides (Figure B).
Figure 1
(A) Structures of the synthesized DA monomers
(1–4) investigated for colorimetric film dosimetry.
(B) Reaction scheme
for the synthesis of amine-substituted DA monomers.
(A) Structures of the synthesized DA monomers
(1–4) investigated for colorimetric film dosimetry.
(B) Reaction scheme
for the synthesis of amine-substituted DA monomers.The amides were prepared by treating the carboxylic acid
of DA
with ethyldimethylaminopropylcarbodiimide hydrochloride (EDC·HCl)
to activate it, followed by the addition of amide.[35] The overall reaction led to the formation of the amide
linkage. The influence of bulky amide headgroups on the radiation
sensitivity was studied by substituting PCDA with aromatic and aliphatic
amide groups. PCDA functionalized with aniline (PCDA-AN 1), tetrahydronaphthylamine (PCDA-TNAP 2), and naphthylamine
(PCDA-NAP 3) groups were synthesized to compare their
colorimetric response toward radiation doses. Aminofluorene (PCDA-AFL 4)-functionalized DA compound was synthesized to probe how
extended aromatic interactions and hydrogen bonding influence the
radiation sensitivity. Spectroscopic data of synthesized monomers
are as follows:
Figure A,B shows the morphology of PCDA-AFL 4 acquired by field electron scanning electron microscopy
(FESEM) before and after radiation-induced polymerization, respectively. Figure A shows embedded
structures in a multilayered morphology, which can be ascribed to
the strong headgroup interactions caused by the hydrogen bonding with
π–π aromatic interactions. After radiation-induced
polymerization (Figure B), uniformly spread intercalated flake-like structures can be noticed.
Although there is no significant change in the morphology of DA monomers
post irradiation, an increase in the population of particles is noticed
which may be attributed to the advancement of monomers toward each
other for the formation of the long PDA chain. From the FESEM images,
it can be inferred that the size of the PCDA-AFL polymers lies in
the range of several micrometers.
Figure 2
(A) Morphology of PCDA-AFL 4 before radiation exposure.
(B) Morphology of PCDA-AFL 4 after radiation exposure.
(A) Morphology of PCDA-AFL 4 before radiation exposure.
(B) Morphology of PCDA-AFL 4 after radiation exposure.
Fabrication of the DA-Based Film Dosimeter
The incorporation
of synthesized DA monomers in the form of films allows immediate radiation
dosimetric applications. The synthesized DA monomers were introduced
in polyvinyl alcohol (PVA) solution to prepare an emulsion, which
was later coated to obtain thick and sturdy film dosimeters. The inert
nature of the PVA matrix does not interfere with the radiation-induced
conformational changes in the guest DA monomers, and hence provides
a good binder solution for the film preparation. A Tinuvin P-based
UV absorber was added to protect the emulsion from pre-irradiation
polymerization because of the environmental light.[36] Because the variation in the thickness of the films may
result in the change in the absorbance and hence radiation response,
thus, it is essential to obtain a highly uniform thickness film for
precise radiation dose measurement.[37,38] Therefore,
instead of conventional film preparation techniques like spin-coating,
Langmuir–Blodgett film deposition, and solvent casting method,
an automatic film applicator with variable thickness were used for
coating DA-based films.[33] Using an automatic
film applicator, the DA-embedded PVA emulsion was uniformly coated
on a transparent polyethylene terephthalate (PET) sheet. The semi-transparent
nature of the obtained films is most suitable to distinctly identify
the appearance of light and dark shades of blue coloration upon exposure
to low and high radiation doses. A 200 μm thick, flexible yet
self-standing film with a 100 μm top coating of DA-based emulsion
on a 100 μm PET sheet was prepared which enabled easy handling
during dosimetry. With the perspective of blood irradiator dosimetry
applications, several film dosimeters of different sizes can be pasted
on the blood bags in various axes and analyzed simultaneously. Figure represents the experimental
procedure and properties like flexibility and transparency of the
prepared film dosimeters.
Figure 3
Pictorial representation of the experimental
conditions for the
preparation of DA–PVA emulsion followed by coating it using
an automatic film applicator unit for uniform thickness (i) flexible
nature of the developed film dosimeters which can take the irregular
shape of the blood bags, (ii) developed film dosimeters can be easily
cut in to any shape and pasted on the blood bags with cello tape,
and (iii) transparent nature of the films is shown by writing the
doses on a paper placed beneath the films.
Pictorial representation of the experimental
conditions for the
preparation of DA–PVA emulsion followed by coating it using
an automatic film applicator unit for uniform thickness (i) flexible
nature of the developed film dosimeters which can take the irregular
shape of the blood bags, (ii) developed film dosimeters can be easily
cut in to any shape and pasted on the blood bags with cello tape,
and (iii) transparent nature of the films is shown by writing the
doses on a paper placed beneath the films.
Colorimetric Response to Radiation Doses
The radiation-induced
polymerization of the DA-based film dosimeter resulted in the formation
of blue colored PDAs which displayed an increase in the intensity
of blue color with the radiation doses. The UV–visible (UV–vis)
absorption spectra were acquired to study the presence of various
absorbance peaks and their evolution with the increase in radiation
doses. Figure A shows
the UV–vis absorption spectra of PCDA and synthesized DA monomers 1–4 after exposure to 5 Gy γ radiation dose.
It is evident from the spectra that PCDA-AFL 4 is most
sensitive to radiation and displays maximum absorbance at a low dose
of 5 Gy. The visible absorption spectra represent the sensitivity
order of PCDA-AFL 4 > PCDA-NAP 3 >
PCDA-TNAP 2 > PCDA-AN 1 > PCDA. It
can be noticed that
after exposure γ radiation PCDA-AFL 4 and PCDA-NAP 3 show significant absorbance in the blue region unlike PCDA
and PCDA-AN 1 in which color transition becomes observable
only at higher doses. It can be elucidated that the DAs having different
side groups display variation in the extent of polymerization and
hence colorimetric sensitivity to radiation dose. It is presumed that
the shift in the main absorbance peak in spectra is contributed by
the conformational change of the DA backbone from planar to nonplanar.
This shift can also be attributed to the change in the effective conjugation
length of the PDAs.[39] However, the maximum
absorption wavelength (λmax) in all of them remains
similar (650–660 nm), which suggests that the basic mechanism
of radiation-induced color transition with respect to the DA main
chain is irrespective of the side groups and remains the same. Figure B shows the absorption
spectra of the film dosimeters prepared with PCDA-AFL 4 at various radiation doses ranging from 0 to 50 Gy which are used
in blood irradiation. A maximum absorption wavelength (λmax) at 660 nm can be seen with a shoulder peak at 600 nm.
The absorbance peak at 660 nm is associated with the blue color of
the amide dosimeter. The appearance of intense blue color upon exposure
to γ-radiation is because of the electron delocalization within
the conjugated backbone. Previous studies suggest that the colorless
to blue color transition is attributed to the changes in orientation,
conformation, and packing caused by the propagation of PDA chains.[44] However, the exact detailed mechanism of color
transition has not been fully determined yet.[40]
Figure 4
(A)
Visible spectra of DA-based film dosimeters. (B) Visible spectroscopic
monitoring of PCDA-AFL 4 film dosimeter upon exposure
to different γ radiation doses along with the photographs showing
the progression of blue coloration with increasing radiation dose.
(A)
Visible spectra of DA-based film dosimeters. (B) Visible spectroscopic
monitoring of PCDA-AFL 4 film dosimeter upon exposure
to different γ radiation doses along with the photographs showing
the progression of blue coloration with increasing radiation dose.A minor peak at ∼550 nm in the 50 Gy absorbance
spectrum
is also observed, which is attributed to the red phase of the PDA.
The absorbance peak observed at 600 nm in the spectra is because of
an intermediate form between the red and blue phases of the PDA. The
absorbance spectra observations confirm the presence of two main chromatic
phases of PDA, that is, the blue phase, which is observed immediately
after the radiation-induced topochemical polymerization of the amide-substituted
DA monomers and a red phase which results after irradiation to high
radiation doses. It can be noticed from the spectra that as the radiation
doses increase, there is a concomitant increase in the absorbance
peak of blue color at 660 nm. These spectral changes corroborate with
the visual observation as shown in the inset photographic images.
Influence of Amide Headgroups on Radiation Sensitivity
In
order to allow DA units to polymerize, a specific molecular arrangement
is required to meet the criteria for topochemical reactions.[40] It is found that the incorporation of substituent
pendant groups to the DA unit can promote polymerization.[41] The ordering, orientation, and packing of the
side chains induce stress to the DA backbone which markedly influences
the optical, electronic, and various physical properties. In our study,
various colorimetric compounds were synthesized by choosing PCDA as
the basic moiety and altering its carboxylic acid group by various
amide-based head groups.As compared to pristine PCDA, all the
developed amide dosimeters exhibited better visual blue coloration
upon exposure to radiation. This is supposedly because of the intramolecular
hydrogen bonding of the amidehydrogen and carbonyl groups of the
carboxylic moiety. The hydrogen bonding tends to form an expanded
π-surface which further extends the π–π stacking
in amide dosimeters.[42] Moreover, the existence
of strong intermolecular hydrogen bonding in the amide-substituted
PCDA is useful for the ordered self-assembly of DA, which is essential
for topochemical polymerization and leads to enhanced colorimetric
radiation response. The improvement in the radiation response from
PCDA-AN 1 and PCDA-TNAP 2 can be seen because
of an increase in the number of rings. Further, in PCDA-NAP 3 two aromatic rings were introduced, the enhanced aromaticity
and conjugation increase the π–π stacking of amide
dosimeters, which results in significant blue coloration upon exposure
to radiation. The substitution of a bulkier, three-ring system as
a pendant group in PCDA-AFL 4 exhibits the most significant
coloration among all the developed film dosimeters at 5 Gy radiation
dose exposure. Thus, it can be proposed that the degree of aromatic
interactions, conjugation and hydrogen bonding of the head groups
can have a pronounceable effect on the radiation-induced polymerization
of the DA chains which leads to chromatic transitions.[34]Raman spectra of the PCDA-AFL 4 film dosimeter was
analyzed to identify the presence of various bonds and monitor their
progression with the increase in radiation doses. The presence of
peaks at 2095 and 1450 cm–1 are attributed to the
characteristic conjugated alkyne-alkene bands of C≡C and C=C,
respectively[43] (Figure A). It is evident that with the increase
in the radiation doses no major shift in the peak position has been
observed but a significant escalation in the intensity of C=C
and C≡C bond intensity peaks can be noticed. Figure B shows the change in the peak
intensity as a function of radiation doses, it can be seen that after
radiation exposure the C=C bond intensity is notably more than
the C≡C bond intensity, which implies that with increasing
doses, one of the C≡C bond in the monomer cleavages and cross-links
with another monomer via C=C bonds. It can be proposed that
because of the radiation exposure the amide-substituted DA molecule
undergoes 1,4 addition polymerization of the conjugated triple bonds
which gives rise to PDA units cumulated with double bonds. Depending
upon the radiation dose, these PDA units undergo cross-linking through
C=C bonds by forming repeating units.
Figure 5
(A) Raman spectra of
the PCDA-AFL 4 film dosimeter
upon exposure to γ radiation doses. (B) Progression of Raman
signal corresponding to C=C (1450 cm–1) and
C≡C bonds (2095 cm–1) in the PCDA-AFL film
dosimeter.
(A) Raman spectra of
the PCDA-AFL 4 film dosimeter
upon exposure to γ radiation doses. (B) Progression of Raman
signal corresponding to C=C (1450 cm–1) and
C≡C bonds (2095 cm–1) in the PCDA-AFL film
dosimeter.
Digital Colorimetric Analysis
for Film Dosimetry
In
digital colorimetric analysis (DCA), quantification of the colorimetric
dose response of the developed dosimeter was done by two-dimensional
OD measurement using a high-resolution Epson 10000XL flatbed scanner.
With the increase in the radiation dose, the blue color of the film
intensifies. Hence, OD is the most preferred measurement quantity.[44] DCA was carried out by scanning the indigenously
developed film dosimeters separately in each channel, that is, red
(R), green (G), and blue (B) thereby extracting the RGB channel values
for each pixel within the dosimeters in the scanned images. The effect
of background light was subtracted by taking a white background, also
the background noises from the scanner bed and the uncoated film were
subtracted. Figure shows the post irradiation intensity profile of the developed film
dosimeter in terms of pixel values of red, blue, and green channels.
It confirms the homogenous dose distribution across the surface of
the film dosimeter. A Wiener filter, which is a MATLAB tool, was applied
to remove the additive noise and blurring.[45] Because the absorption spectrum of the radiochromic film exhibits
maximum peak in the red region of the visible spectrum, thus, the
red channel of the scanner was chosen to monitor the changes in OD.
Figure 6
Intensity
profile of the PCDA-AFL 4 film dosimeter
post irradiation in red, blue, and green channels.
Intensity
profile of the PCDA-AFL 4 film dosimeter
post irradiation in red, blue, and green channels.A MATLAB code was written to evaluate the changes in the
amount
of coloration because of radiation exposure by calculating the net
OD for all the samples from the pixel values. Figure represents the variation in OD of the developed
film dosimeters with radiation doses and the data were fitted by using
second degree polynomial. The uncertainty in OD measurement defined
as the standard deviation of the mean value of OD was evaluated by
the statistical analysis of a series of measurements.[46] The mean and standard deviation were computed over a set
of averaged measured intensities over the entire surface of 10 unexposed
and exposed films read 5 times each. The statistical uncertainty in
OD was obtained to be 1.4%.
Figure 7
DCA using a high-resolution Epson scanner.
DCA using a high-resolution Epson scanner.
Reproducibility and Uniformity of the Amide
Film Dosimeter
The PCDA-AFL 4 amide film dosimeters
were scanned
on different days to investigate the film reproducibility. Five sets
of films comprising film pieces were irradiated with selected known
doses and scanned over a period of 10 days. The mean OD for the film
dosimeters was obtained and normalized with respect to the OD of the
first day. Figure A shows the results for the film reproducibility. It can be seen
that the film response changes very slightly with time with an average
darkening of 0.2 ± 0.059% (1 standard deviation) after 10 days.
The study shows that the amide film dosimeters exhibit stable and
reproducible response over a period of 10 days.
Figure 8
(A) Reproducibility of
PCDA-AFL film dosimeters. (B) Dose response
curve of the PCDA-AFL film dosimeter obtained on different days. (C)
Dose response curve obtained at three different ROIs on the PCDA-AFL
films.
(A) Reproducibility of
PCDA-AFL film dosimeters. (B) Dose response
curve of the PCDA-AFL film dosimeter obtained on different days. (C)
Dose response curve obtained at three different ROIs on the PCDA-AFL
films.The quantification of the doses
from OD acquired by the high-resolution
flatbed scanner is fast, cost effective, and practically more feasible
in a clinical setup.[47] However, it is important
to study the reproducibility of this method in the developed amide
dosimeters. The reproducibility of the DCA was investigated by repeatedly
scanning film pieces irradiated at 0, 5, 15, 20, 30, and 50 Gy on
different days: after 1, 3, 5, and 7 days. All the films were positioned
at the center of the scanner in a reproducible manner by using a cut-out
template. The template was useful in minimizing the nonuniform response
of the measurement because of the inconsistent positioning of the
films. The settings of the scanner were maintained the same on all
days. Figure B shows
the dose response curve obtained on different days. The average variation
for all the measurements performed on different days was 1.18%.Uniformity of the films plays an important role in obtaining a
consistent dose response over the complete surface of the films. Therefore,
the uniformity of the films was investigated by comparing the dose
response measured at 3 different regions of interest (ROIs) on different
locations on a single piece of the film. The ROIs were taken in the
central area of the film to avoid the artifacts from the edges of
the films. The physical deformities in the film were also avoided.
The OD was measured using the GretagMacbeth densitometer. An average
variation of 1.53% was obtained at 3 different ROIs (Figure C), which indicates that there
is limited variation in scatter and dose sensitivity.
Storage Conditions
for the Amide Dosimeters
The effect
of pre-irradiation and post-irradiation storage conditions on the
developed films was monitored by measuring L, a, and b values of the films stored under
different conditions.[48] To establish the
effect of temperature and room light, three sets of films were taken
in a Petri dish, one set of films was stored at room temperature (RT)
in normal room light, the second set was kept in an oven at 50 °C
in the dark and the third set was kept in a refrigerator at 4 °C. Figure shows the variation
in L (degree of lightness or darkness) values in
all three sets measured at different time intervals pre- and post-irradiation.
It can be seen that the L values of the pre- and
post-exposed films stored at RT and at 4 °C remained essentially
unchanged during the whole period of observation. The L values of the film dosimeters stored at 50 °C changed about
2%. The thermal reactivity of the amide derivatives of the DA monomer
can be responsible for the slight decrease in L values
before irradiation. It can be concluded that the developed film dosimeters
have a long shelf life both pre- and post-exposure in normal environmental
conditions and hence can be easily stored without the requirement
of any special storage conditions.
Figure 9
Pre- and post-irradiation L values of the PCDA-AFL 4 film dosimeter at different
storage conditions.
Pre- and post-irradiation L values of the PCDA-AFL 4 film dosimeter at different
storage conditions.
Conclusions
The
DA-based colorimetric film dosimeter developed in this work
enables visual detection and measurement of low to high range of radiation
doses used in blood irradiation. By leveraging the ability of DA to
undergo radiation-induced polymerization which leads to chromatic
transitions, a series of amide-functionalized DA compounds were synthesized
and later used to prepare films. The effect of variation of the bulky
amide head group on the radiation sensitivity was studied. It was
found that the aminofluorene-substituted DA monomer exhibited significant
color transition upon exposure to 5 Gy γ radiation and the color
intensity increased till 50 Gy. Using the developed film dosimeter,
the measurement of radiation of doses can be simply made by quantifying
the color transition from white to different intensities of blue coloration
in terms of OD. The amount of coloration was quantified by digital
image processing using MATLAB code. In a routine dosimetry situation,
this could be an easy tool to investigate if the blood bags are adequately
irradiated to the recommended doses. The developed film dosimeters
exhibit a long shelf life under normal storage conditions. They can
be cut into any size and pasted on the blood bags to record the absolute
doses and relative dose distribution at the different planes, which
can be useful for the time-to-time quality assurance of the blood
irradiator.
Experimental Section
Materials and Instruments
PCDA was
purchased from Sigma-Aldrich
and 1-ethyl-3-(3-dimethylaminopropyl)carbodiimide hydrochloride (EDC·HCl)
was purchased from TCI Chemicals. 1H NMR (400 MHz) and 13C NMR (100 MHz) spectra were recorded in CDCl3. Chemical shifts for protons and carbons are reported in ppm from
tetramethylsilane and referenced to the carbon resonance of the solvent.
The differential UV–vis absorption spectra of the film dosimeters
were recorded using a PerkinElmer Lambda 1050 spectrophotometer equipped
with a 150 mm integrating sphere with 8° reflectance. The spot
size of a spectrophotometer was 0.4 cm wide and 1.2 cm high for a
3 cm × 3 cm film sample. FESEM images of DA monomers studied
here obtained on Carl Zeiss (GeminiSEM 500). Freshly prepared compounds
were deposited on silicon wafers. Samples were sputter-coated on Quorum
pure Au target, sputter time was 60 s and sputter current was 60 mA.
A Horiba LabRam HR revolution Raman spectrometer with 785 nm laser
50 mW power was used to record the Raman spectra of the developed
film dosimeters. The acquisition time was 2 s. High-resolution mass
spectra were recorded on a Applied Biosystems QSTAR Elite Hybrid (QqTOF)
LC/MS/MS system with an electrospray mass spectrometer using positive
electrospray ionization mode (ESI+). The L, a, and b values of the film dosimeters
were measured using NCS Colourpin 11 color reader. It was connected
to the mobile phone through Bluetooth and an online app was downloaded
to read the color values of the films on a mobile phone. The Fourier
transform infrared spectra of the film dosimeters were obtained from
a Thermo Scientific spectrometer in attenuated total reflection mode.
Preparation of DA Monomers (1–4)
In
an oven-dried round bottom flask, PCDA 1.0 equiv in dichloromethane
(DCM) was added dropwise in a solution of EDC·HCl 1.5 equiv in
DCM at 0 °C. After 30 min, the substituted amine 1.0 equiv was
added to the reaction mixture. The resulting reaction mixture was
run at RT overnight. The progression of the reaction was monitored
by thin-layer chromatography analysis; after the complete consumption
of the starting material. The reaction mixture was diluted with DCM
(10 mL) and water (15 mL). The layers were separated, and the organic
layer was washed with brine solution and dried over sodium sulfate
(Na2SO4). The organic layer was concentrated
under reduced pressure. The crude material so obtained was purified
by column chromatography on silica gel (100–200) (hexane/ethyl
acetate; 70/30).
Fabrication of the DA Monomer Film Dosimeter
A 5% PVA
solution in distilled water was prepared. The synthesized DA monomer
dissolved in ethyl acetate was added to the PVA solution in a 1:1
ratio. The solution was stirred and heated at 60 °C for 3 h to
form a turbid emulsion. A Tinuvin P-based UV absorber (0.1%) was added.
The emulsion was coated on a transparent sheet using an automatic
film applicator and left for drying overnight at RT. The obtained
film dosimeters were cut to pieces as per the requirement. The thickness
of the dried layer of the obtained films was 100 μm.
Irradiation
with Blood Irradiator
The irradiation was
performed by Gammacell 3000 Elan, blood irradiator (Best Theratronics
Ltd., Ontario, Canada). It contains one sealed 137Cs radiation
source with a nominal activity of 53.7 TBq (1450 Ci) inside a steel-encased
lead shield with a dose rate of 5 Gy/min at the canister midplane.
Digital Colorimetric Analysis
All the samples were
scanned by using an Epson 10000XL flatbed scanner in reflective mode.
The images were digitalized in 48 bits RGB format at a resolution
of 72 dpi and saved in uncompressed TIFF format. The image processing
was performed using an open-source program written in MATLAB 7.4.
The analysis was performed in the red channel component of the image.
Authors: S Vandana; V S Shaiju; S D Sharma; S Mhatre; S Shinde; G Chourasiya; Y S Mayya Journal: Appl Radiat Isot Date: 2010-09-17 Impact factor: 1.513
Authors: A Kessinger; J O Armitage; L W Klassen; J D Landmark; J M Hayes; A E Larsen; D T Purtilo Journal: J Surg Oncol Date: 1987-11 Impact factor: 3.454