Evan B Asher1, Nikhil Panda2, Cao Thach Tran2, Michael Wu2.
Abstract
Entities:
Keywords: Fitness tracker; Implanted cardioverter-defibrillator; Magnet; Magnetic interference; Reed switch; Smart watch
Year: 2020 PMID: 33786313 PMCID: PMC7987898 DOI: 10.1016/j.hrcr.2020.12.002
Source DB: PubMed Journal: HeartRhythm Case Rep ISSN: 2214-0271
Figure 1Fitbit wristband magnet deactivating implanted cardioverter-defibrillator at depth of 2.4 cm.
Maximum distances, in centimeters, at which various magnets could deactivate the implanted cardioverter-defibrillator
| Magnet type | Distance at which ICD deactivates (cm) |
|---|---|
| Fitbit wristband | 2.4 |
| Apple Watch wristband | 2.0 |
| Donut magnet | 8.0 |
| Programming head | 7.0 |
ICD = implanted cardioverter-defibrillator.