Potential interference of small neodymium magnets with cardiac pacemakers and implantable cardioverter-defibrillators.

Thomas Wolber1, Salome Ryf, Christian Binggeli, Johannes Holzmeister, Corinna Brunckhorst, Roger Luechinger, Firat Duru.   

Abstract

BACKGROUND: Magnetic fields may interfere with the function of cardiac pacemakers and implantable cardioverter-defibrillators (ICDs). Neodymium-iron-boron (NdFeB) magnets, which are small in size but produce strong magnetic fields, have become widely available in recent years. Therefore, NdFeB magnets may be associated with an emerging risk of device interference.
OBJECTIVE: We conducted a clinical study to evaluate the potential of small NdFeB magnets to interfere with cardiac pacemakers and ICDs.
METHODS: The effect of four NdFeB magnets (two spherical magnets 8 and 10 mm in diameter, a necklace made of 45 spherical magnets, and a magnetic name tag) was tested in forty-one ambulatory patients with a pacemaker and 29 patients with an ICD. The maximum distance at which the magnetic switch of a device was influenced was observed.
RESULTS: Magnetic interference was observed in all patients. The maximum distance resulting in device interference was 3 cm. No significant differences were found with respect to device manufacturer and device types.
CONCLUSION: Small NdFeB magnets may cause interference with cardiac pacemakers and ICDs. Patients should be cautioned about the interference risk associated with NdFeB magnets during daily life.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17198980     DOI: 10.1016/j.hrthm.2006.09.012

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  11 in total

1.  [Patients suffering from atrial fibrillation in Germany. Characteristics, resource consumption and costs].

Authors:  T Reinhold; S Rosenfeld; F Müller-Riemenschneider; S N Willich; T Meinertz; P Kirchhof; B Brüggenjürgen
Journal:  Herz       Date:  2012-08       Impact factor: 1.443

2.  Incidence of magnet mode in patients with implantable cardioverter defibrillators.

Authors:  Gesa von Olshausen; Johanna Schorr; Christian Grebmer; Severin Weigand; Patrick Blažek; Amir Brkic; Lena Friedrich; Verena Semmler; Christof Kolb; Carsten Lennerz
Journal:  J Interv Card Electrophysiol       Date:  2019-10-16       Impact factor: 1.900

3.  Intermittent, erratic behaviour of an implantable cardioverter defibrillator secondary to a hidden magnetic source of interference.

Authors:  Roy Beinart; Mary L Guy; Patrick T Ellinor
Journal:  Europace       Date:  2011-05-15       Impact factor: 5.214

4.  Accuracy of fit of 3-to-3 retainers after adhesive fixation using a neodymium-iron-boron magnet chain.

Authors:  Wolfram Hahn; Wiebke Wasser-Merkel; Katharina Lange; Rudolf M Gruber; Dietmar Kubein-Meesenburg; Dankmar Ihlow
Journal:  J Orofac Orthop       Date:  2011-10       Impact factor: 1.938

Review 5.  [Electromagnetic interference : Pacemakers, cardiac resynchronization therapy devices, implantable cardioverter-defibrillator].

Authors:  Bernd Nowak; Oliver Przibille; Andreas Napp
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2022-07-04

6.  Effects of external electrical and magnetic fields on pacemakers and defibrillators: from engineering principles to clinical practice.

Authors:  Roy Beinart; Saman Nazarian
Journal:  Circulation       Date:  2013-12-24       Impact factor: 29.690

7.  Interference between surgical magnetic drapes and pacemakers: an observational study comparing commercially available devices and a new magnetically isolated drape.

Authors:  Valerie Zaphiratos; Hubert Chiasson; Pierre Drolet; Bruno Benzaquen; Jacques Lapointe; Louis-Philippe Fortier
Journal:  Biomed Eng Online       Date:  2016-07-19       Impact factor: 2.819

Review 8.  Magnetic anchor guidance for endoscopic submucosal dissection and other endoscopic procedures.

Authors:  Mohamed Mortagy; Neal Mehta; Mansour A Parsi; Seiichiro Abe; Tyler Stevens; John J Vargo; Yutaka Saito; Amit Bhatt
Journal:  World J Gastroenterol       Date:  2017-04-28       Impact factor: 5.742

9.  Smart wearable device accessories may interfere with implantable cardiac devices.

Authors:  Evan B Asher; Nikhil Panda; Cao Thach Tran; Michael Wu
Journal:  HeartRhythm Case Rep       Date:  2020-12-13

Review 10.  Interference by Modern Smartphones and Accessories with Cardiac Pacemakers and Defibrillators.

Authors:  Fahd Nadeem; Cao Thach Tran; Estelle Torbey; Daniel Philbin; Carlos Morales; Michael Wu
Journal:  Curr Cardiol Rep       Date:  2022-01-27       Impact factor: 2.931

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.