| Literature DB >> 33713171 |
S Sawall1,2, L Klein3,4, E Wehrse5,6, L T Rotkopf5,6, C Amato3,5, J Maier3, H-P Schlemmer5,6, C H Ziener5,6, S Heinze7, M Kachelrieß3,5.
Abstract
OBJECTIVE: To evaluate the dual-energy (DE) performance and spectral separation with respect to iodine imaging in a photon-counting CT (PCCT) and compare it to dual-source CT (DSCT) DE imaging.Entities:
Keywords: Angiography; Contrast media; Iodine; Phantoms, imaging; Tomography, X-ray computed
Mesh:
Substances:
Year: 2021 PMID: 33713171 PMCID: PMC8379121 DOI: 10.1007/s00330-021-07786-0
Source DB: PubMed Journal: Eur Radiol ISSN: 0938-7994 Impact factor: 5.315
Fig. 1Left: Experimental photon-counting CT and the used semi-anthropomorphic phantom. Right: Semi-anthropomorphic liver phantom used in the experiments. The numbers indicate the iodine concentration in the vials measured in mg/mL. Zero denotes pure water
Acquisition and reconstruction parameters for all experiments conducted with phantoms and the post-mortem angiography study (FOM, field of measurement; FOV, field of view; AEC, automatic exposure control)
| DSCT | PCCT | |
|---|---|---|
| Tube voltagephantoms | 80 kV/140 kV + Sn, 100 kV/140 kV + Sn | 80 kV, 100 kV, 120 kV, 140 kV |
| Tube voltagepost-mortem | - | 140 kV |
| CTDI32 cm | 11 mGy | 11 mGy |
| AEC | off | off |
| FOM | 500 mm/330 mm | 275 mm |
| FOV | 275 mm | 275 mm |
| Image matrix | 512 × 512 px | 512 × 512 px |
| Pixel size | 0.54 mm | 0.54 mm |
| Slice thickness | 1.50 mm | 1.50 mm |
| Slice increment | 0.75 mm | 0.75 mm |
| Kernel | D40f | D40f |
Fig. 2R as function of threshold for all available tube voltages. Colors encode phantom sizes, i.e., the S-phantom (green), the M-phantom (blue), and the L-phantom (red). Black dashed-dotted lines show R measured in a reference dual-source dual-energy CT system and represent the average overall phantom sizes
Fig. 3Dose-normalized contrast-to-noise ratio (CNRD) of iodine maps (colored solid lines) and noise quotient (RelNoise = σLow/σHigh, dashed lines) for all phantoms sizes, all tube voltages, and all thresholds. Colors encode phantom sizes, i.e., the S-phantom (green), the M-phantom (blue), and the L-phantom (red). Black dashed-dotted lines were acquired using the M-phantom in a reference dual-source dual-energy CT system
Fig. 4Iodine maps for all available tube voltages of the S-phantom acquired at the same dose. Images were acquired at the optimal threshold according to Fig. 3. Magnifications of the yellow dashed region of interest are shown in the top right of each figure (C = 7 mg/mL, W = 30 mg/mL)
Fig. 5Post-mortem CT angiography using a tube voltage of 140 kV for thresholds of 50 keV (left column), 70 keV (middle column), and 90 keV (right column). Low energy bins (top row), high energy bins (middle row), and iodine maps (bottom row) (bin images C = 300 HU, W = 1400 HU; material maps C = 15 mg/mL, W = 30 mg/mL)