The importance of spectral separation: an assessment of dual-energy spectral separation for quantitative ability and dose efficiency.

Bernhard Krauss1, Katharine L Grant, Bernhard T Schmidt, Thomas G Flohr.   

Abstract

INTRODUCTION: One method to acquire dual-energy (DE) computed tomography (CT) data is to perform CT scans at 2 different x-ray tube voltages, typically 80 and 140 kV, either as 2 separate scans, by means of rapid kV switching, or with the use of 2 x-ray sources as in dual-source CT (DSCT) systems. In DSCT, it is possible to improve spectral separation with tin prefiltration (Sn) of the high-kV beam. Recently, x-ray tube voltages beyond the established range of 80 to 140 kV were commercially introduced, which enable additional voltage combinations for DE acquisitions, such as 80/150 Sn or 90/150 Sn kV. Here, we investigate the DE performance of several x-ray tube voltages and prefilter combinations on 2 DSCT scanners and the impact of the spectra on quantitative analysis and dose efficiency.
MATERIALS AND METHODS: Circular phantoms of different sizes (10-40 cm in diameter) equipped with cylindrical inserts containing water and diluted iodine contrast agent (14.5 mg/cm) were scanned using 2 different DSCT systems (SOMATOM Definition Flash and SOMATOM Force; Siemens AG, Forchheim, Germany). Five x-ray tube voltage combinations (80/140 Sn, 100/140 Sn, 80/150 Sn, 90/150 Sn, and 100/150 Sn kV) were investigated, and the results were compared with the previous standard acquisition technique (80/140 kV). As an example, 80/140 Sn kV means that 1 x-ray tube of the DSCT system was operated at 80 kV, whereas the other was operated at 140 kV with additional tin prefiltration (Sn). Dose values in terms of computed tomography dose index (CTDIvol) were kept constant between the different voltage combinations but adjusted with regard to object size according to automatic exposure control recommendations. Reconstructed images were processed using linear blending of the low- and high-kV CT images to combined images, as well as 3-material decomposition techniques to generate virtual noncontrast (VNC) images and iodine images. Contrast and pixel noise were evaluated, as well as DE ratios, which are defined as the CT value at low kV divided by the CT value at high kV.
RESULTS: For the 10-, 20-, 30-, and 40-cm phantom, dose values in terms of CTDIvol were 1.2, 2.6, 7.3, and 21.6 mGy, respectively. In the combined images, those obtained with tin filtration showed lower noise values at similar iodine enhancement levels than did images obtained without tin filtration. The largest differences in noise were observed for the larger phantoms, in particular the 40-cm phantom. Dual-energy ratios for iodine increased with decreasing voltages of the low-kV beam and with increasing voltages of the high-kV beam, and they increased when tin prefiltration was added. In case of the 20-cm phantom, DE ratios ranged from 2.0 at 80/140 kV to 3.4 at 80/150 Sn kV. The noise level of the VNC images was strongly correlated with the inverse of the DE ratio. Irrespective of the phantom size, the lowest noise values were measured for 80/150 Sn kV. DISCUSSION: Dual-source CT systems enable DE data to be acquired using a variety of voltage combinations. Combined (or mixed) DE images provide an image impression similar to standard 120 kV images, yet the noise level depends on the DE voltage combination that is selected. Noise in decomposed VNC images is strongly influenced by the DE ratio, and it improves substantially with tin filtration of the high-voltage beam.

Entities:  

Mesh:

Year:  2015        PMID: 25373305     DOI: 10.1097/RLI.0000000000000109

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  32 in total

1.  Single- and dual-energy CT of the abdomen: comparison of radiation dose and image quality of 2nd and 3rd generation dual-source CT.

Authors:  Julian L Wichmann; Andrew D Hardie; U Joseph Schoepf; Lloyd M Felmly; Jonathan D Perry; Akos Varga-Szemes; Stefanie Mangold; Damiano Caruso; Christian Canstein; Thomas J Vogl; Carlo N De Cecco
Journal:  Eur Radiol       Date:  2016-05-10       Impact factor: 5.315

2.  Interventional dual-energy imaging-Feasibility of rapid kV-switching on a C-arm CT system.

Authors:  K Müller; S Datta; M Ahmad; J-H Choi; T Moore; L Pung; C Niebler; G E Gold; A Maier; R Fahrig
Journal:  Med Phys       Date:  2016-10       Impact factor: 4.071

3.  A noise-optimized virtual monochromatic reconstruction algorithm improves stent visualization and diagnostic accuracy for detection of in-stent re-stenosis in lower extremity run-off CT angiography.

Authors:  Stefanie Mangold; Carlo N De Cecco; U Joseph Schoepf; Ricardo T Yamada; Akos Varga-Szemes; Andrew C Stubenrauch; Damiano Caruso; Stephen R Fuller; Thomas J Vogl; Konstantin Nikolaou; Thomas M Todoran; Julian L Wichmann
Journal:  Eur Radiol       Date:  2016-03-09       Impact factor: 5.315

4.  Coronary CT Angiography: Variability of CT Scanners and Readers in Measurement of Plaque Volume.

Authors:  Rolf Symons; Justin Z Morris; Colin O Wu; Amir Pourmorteza; Mark A Ahlman; João A C Lima; Marcus Y Chen; Marissa Mallek; Veit Sandfort; David A Bluemke
Journal:  Radiology       Date:  2016-09-16       Impact factor: 11.105

5.  The Effect of Patient Diameter on the Dual-Energy Ratio of Selected Contrast-Producing Elements.

Authors:  Jack W Lambert; Paul F FitzGerald; Peter M Edic; Yuxin Sun; Peter J Bonitatibus; Robert E Colborn; Benjamin M Yeh
Journal:  J Comput Assist Tomogr       Date:  2017 May/Jun       Impact factor: 1.826

6.  Longitudinal reproducibility of attenuation measurements on virtual unenhanced images: multivendor dual-energy CT evaluation.

Authors:  Simon Lennartz; Anushri Parakh; Jinjin Cao; Avinash Kambadakone
Journal:  Eur Radiol       Date:  2021-06-10       Impact factor: 5.315

7.  An Image-Domain Contrast Material Extraction Method for Dual-Energy Computed Tomography.

Authors:  Jack W Lambert; Yuxin Sun; Robert G Gould; Michael A Ohliger; Zhixi Li; Benjamin M Yeh
Journal:  Invest Radiol       Date:  2017-04       Impact factor: 6.016

8.  A noise-optimized virtual monoenergetic reconstruction algorithm improves the diagnostic accuracy of late hepatic arterial phase dual-energy CT for the detection of hypervascular liver lesions.

Authors:  Carlo N De Cecco; Damiano Caruso; U Joseph Schoepf; Domenico De Santis; Giuseppe Muscogiuri; Moritz H Albrecht; Felix G Meinel; Julian L Wichmann; Philip F Burchett; Akos Varga-Szemes; Douglas H Sheafor; Andrew D Hardie
Journal:  Eur Radiol       Date:  2018-02-19       Impact factor: 5.315

9.  Dual Energy Differential Phase Contrast CT (DE-DPC-CT) Imaging.

Authors:  Xu Ji; Ran Zhang; Ke Li; Guang-Hong Chen
Journal:  IEEE Trans Med Imaging       Date:  2020-10-28       Impact factor: 10.048

Review 10.  Dual-energy CT: theoretical principles and clinical applications.

Authors:  Andrea Agostini; Alessandra Borgheresi; Alberto Mari; Chiara Floridi; Federico Bruno; Marina Carotti; Nicolò Schicchi; Antonio Barile; Stefania Maggi; Andrea Giovagnoni
Journal:  Radiol Med       Date:  2019-12-02       Impact factor: 3.469