Literature DB >> 33656261

Intra-fractional patient motion when using the Qfix Encompass immobilization system during HyperArc treatment of patients with brain metastases.

Riho Komiyama1, Shingo Ohira1,2, Hikari Ueda1, Naoyuki Kanayama1, Akira Masaoka1, Masaru Isono1, Yoshihiro Ueda1, Masayoshi Miyazaki1, Teruki Teshima1.   

Abstract

PURPOSE: This study investigated the intra-fractional motion (IM) of patients immobilized using the QFix Encompass Immobilization System during HyperArc (HA) treatment.
METHOD: HA treatment was performed on 89 patients immobilized using the Encompass. The IM during treatment (including megavoltage (MV) registration) was analyzed for six degrees of freedom including three axes of translation (anterior-posterior, superior-inferior (SI) and left-right (LR)) and three axes of rotation (pitch, roll, and yaw). Then, the no corrected IM (IMNC ) was retrospectively simulated (excluding MV registration) in three directions (SI, LR, and yaw). Finally, the correlation between the treatment time and the IM of the 3D vector was assessed.
RESULTS: The average IM in terms of the absolute displacement were 0.3 mm (SI), 0.3 mm (LR) and 0.2° (yaw) for Stereotactic radiosurgery (SRS), and 0.3 mm (SI), 0.2 mm (LR), and 0.2° (yaw) for stereotactic radiotherapy (SRT). The absolute maximum values of IM were <1 mm along the SI and LR axes and <1° along the yaw axis. The absolute maximum displacements for IMNC were >1 mm along the SI and LR axes and >1° along the yaw axis. In the correlation between the treatment time and the IM, the r-values were -0.025 and 0.027 for SRS and SRT respectively, along the axes of translation. For the axes of rotation, the r-values were 0.012 and 0.206 for SRS and SRT, respectively.
CONCLUSION: Encompass provided patient immobilization with adequate accuracy during HA treatment. The absolute maximum displacement IM was less than IMNC along the translational/rotational axes, and no statistically significant relationship between the treatment time and the IM was observed.
© 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  HyperArc; QFix EncompassTM immobilization system; intra-fractional patient motion; stereotactic radiosurgery; stereotactic radiotherapy

Mesh:

Year:  2021        PMID: 33656261      PMCID: PMC7984469          DOI: 10.1002/acm2.13143

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


  18 in total

1.  Accuracy evaluation of the optical surface monitoring system on EDGE linear accelerator in a phantom study.

Authors:  Pietro Mancosu; Antonella Fogliata; Antonella Stravato; Stefano Tomatis; Luca Cozzi; Marta Scorsetti
Journal:  Med Dosim       Date:  2016-03-16       Impact factor: 1.482

2.  A Systematic Analysis of Errors in Target Localization and Treatment Delivery for Stereotactic Radiosurgery Using 2D/3D Image Registration.

Authors:  Hao Xu; Stephen Brown; Indrin J Chetty; Ning Wen
Journal:  Technol Cancer Res Treat       Date:  2016-08-31

3.  Single-Isocenter Multiple-Target Stereotactic Radiosurgery: Risk of Compromised Coverage.

Authors:  Justin Roper; Vorakarn Chanyavanich; Gregory Betzel; Jeffrey Switchenko; Anees Dhabaan
Journal:  Int J Radiat Oncol Biol Phys       Date:  2015-07-20       Impact factor: 7.038

4.  A clinical comparison of patient setup and intra-fraction motion using frame-based radiosurgery versus a frameless image-guided radiosurgery system for intracranial lesions.

Authors:  Naren Ramakrishna; Florin Rosca; Scott Friesen; Evrim Tezcanli; Piotr Zygmanszki; Fred Hacker
Journal:  Radiother Oncol       Date:  2010-01-28       Impact factor: 6.280

Review 5.  Stereotactic radiosurgery for treatment of brain metastases. A report of the DEGRO Working Group on Stereotactic Radiotherapy.

Authors:  Martin Kocher; Andrea Wittig; Marc Dieter Piroth; Harald Treuer; Heinrich Seegenschmiedt; Maximilian Ruge; Anca-Ligia Grosu; Matthias Guckenberger
Journal:  Strahlenther Onkol       Date:  2014-04-09       Impact factor: 3.621

6.  Defining the optimal planning target volume in image-guided stereotactic radiosurgery of brain metastases: results of a randomized trial.

Authors:  John P Kirkpatrick; Zhiheng Wang; John H Sampson; Frances McSherry; James E Herndon; Karen J Allen; Eileen Duffy; Jenny K Hoang; Zheng Chang; David S Yoo; Chris R Kelsey; Fang-Fang Yin
Journal:  Int J Radiat Oncol Biol Phys       Date:  2014-10-21       Impact factor: 7.038

Review 7.  Treatment of Brain Metastases.

Authors:  Xuling Lin; Lisa M DeAngelis
Journal:  J Clin Oncol       Date:  2015-08-17       Impact factor: 44.544

8.  Evaluation of residual patient position variation for spinal radiosurgery using the Novalis image guided system.

Authors:  Jian-Yue Jin; Samuel Ryu; Jack Rock; Kathleen Faber; Qing Chen; Munther Ajlouni; Benjamin Movsas
Journal:  Med Phys       Date:  2008-03       Impact factor: 4.071

9.  Frameless linac-based stereotactic radiosurgery (SRS) for brain metastases: analysis of patient repositioning using a mask fixation system and clinical outcomes.

Authors:  Giuseppe Minniti; Claudia Scaringi; Enrico Clarke; Maurizio Valeriani; Mattia Osti; Riccardo Maurizi Enrici
Journal:  Radiat Oncol       Date:  2011-11-16       Impact factor: 3.481

10.  Evaluation and verification of the QFix EncompassTM couch insert for intracranial stereotactic radiosurgery.

Authors:  Karen Chin Snyder; Ilma Xhaferllari; Yimei Huang; M Salim Siddiqui; Indrin J Chetty; Ning Wen
Journal:  J Appl Clin Med Phys       Date:  2018-06-15       Impact factor: 2.102

View more
  3 in total

1.  Comparison of intrafractional motion with two frameless immobilization systems in surface-guided intracranial stereotactic radiosurgery.

Authors:  Chunhui Han; Arya Amini; Jeffrey Y C Wong; Jieming Liang; Kun Qing; W Tyler Watkins; Sean Zhang; Terence M Williams; An Liu
Journal:  J Appl Clin Med Phys       Date:  2022-04-19       Impact factor: 2.243

2.  Machine learning-based treatment couch parameter prediction in support of surface guided radiation therapy.

Authors:  Geert De Kerf; Michaël Claessens; Isabelle Mollaert; Wim Vingerhoed; Dirk Verellen
Journal:  Tech Innov Patient Support Radiat Oncol       Date:  2022-08-12

3.  Intra-fractional motion error during HyperArc stereotactic radiosurgery on patients with brain metastases: Comparison of open and full-face clamshell-style immobilization devices.

Authors:  Shingo Ohira; Riho Komiyama; Naoyuki Kanayama; Yoshihiro Ueda; Shoki Inui; Masayoshi Miyazaki; Masahiko Koizumi; Koji Konishi
Journal:  J Appl Clin Med Phys       Date:  2022-01-20       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.