Ørjan G Martinsen1,2, Håvard Kalvøy2, Dindar S Bari3, Christian Tronstad2.
Abstract
A circuit is presented that enables measurement of skin electrical conductance, susceptance, and potential simultaneously beneath the same monopolar electrode. Example measurements are shown to confirm the function of the circuit. The measurements are also in accordance with earlier findings that changes in skin conductance and potential do not always correspond and hence contain unique information.Entities:
Keywords: Electrodermal activity; bioimpedance; electrodes
Year: 2019 PMID: 33584891 PMCID: PMC7851977 DOI: 10.2478/joeb-2019-0016
Source DB: PubMed Journal: J Electr Bioimpedance ISSN: 1891-5469
Figure 1Circuit for simultaneous measurement of skin conductance, susceptance and potential.
Fig. 2Measurements on a 20-year-old female healthy volunteer. Measured conductance in green, susceptance in blue and potential in red (right hand Y-axis).
Fig. 3Measurements on a 19-year-old male healthy volunteer. Measured conductance in green, susceptance in blue and potential in red (right hand Y-axis).