E M Tegaw1,2, Gh Geraily1,3, S M Etesami4, S Gholami1,3, H Ghanbari5, M Farzin3,6, G F Tadesse1,7, M Shojaei1.
Abstract
BACKGROUND: Online Monte Carlo (MC) treatment planning is very crucial to increase the precision of intraoperative radiotherapy (IORT). However, the performance of MC methods depends on the geometries and energies used for the problem under study.Entities:
Keywords: Computer Simulation; EGSnrc/Epp, Radiotherapy; INTRABEAM® System; MCNP4c; Monte Carlo Method; Monte Carlo N-Particle Transport; Simulation; Spherical Applicators; Statistical Uncertainty
Year: 2021 PMID: 33564639 PMCID: PMC7859382 DOI: 10.31661/jbpe.v0i0.2008-1171
Source DB: PubMed Journal: J Biomed Phys Eng ISSN: 2251-7200
Figure 1A model of simulation set-up similar to the measurement of depth-doses for the purpose of simulation validation. These sketches are not drawn to show the corresponding scales.
Figure 2A model of simulation set-up in the presence of a 2 mm air gap between the spherical applicators and tumor bed. These sketches are not drawn to show the corresponding scales.
Figure 3Monte Carlo N-Particle Transport Code version 4c (MCNP4c) and Electron Gamma Shower, National Research Council/Easy particle propagation (EGSnrc/Epp) was validated based on the measured depth-dose profile data using both 1.5- and 5-cm-diameter applicators. Depth is measured from the applicator surface (depth=0 corresponds to applicator surfaces)
Simulation parameters of Monte Carlo N-Particle Transport Code version 4c (MCNP4c) and Electron Gamma Shower, National Research Council/Easy particle propagation (EGSnrc/Epp)
| Parameter | EGSnrc/Epp | MCNP4c |
|---|---|---|
| 5e10 | 1e9 | |
| 90.26 | 117.89 | |
| Less than 0.5 | Less than 2 |
EGSnrc/Epp: Electron Gamma Shower, National Research Council/Easy particle propagation, MCNP4c: Monte Carlo N-Particle Transport Code version 4c
Figure 4Depth-dose profiles without and with a 2-mm air gap, for 1.5- and 5-cm-diameter applicators for both Monte Carlo N-Particle Transport Code version 4c (MCNP4c) and Electron Gamma Shower, National Research Council/Easy particle propagation (EGSnrc/Epp) codes. Depth is measured from the applicator surface (depth = 0, corresponds to the applicator surfaces)