Literature DB >> 28535919

Precision IORT - Image guided intraoperative radiation therapy (igIORT) using online treatment planning including tissue heterogeneity correction.

Frank Schneider1, Frederic Bludau2, Sven Clausen3, Jens Fleckenstein3, Udo Obertacke2, Frederik Wenz3.   

Abstract

BACKGROUND: To the present date, IORT has been eye and hand guided without treatment planning and tissue heterogeneity correction. This limits the precision of the application and the precise documentation of the location and the deposited dose in the tissue. Here we present a set-up where we use image guidance by intraoperative cone beam computed tomography (CBCT) for precise online Monte Carlo treatment planning including tissue heterogeneity correction.
MATERIALS AND METHODS: An IORT was performed during balloon kyphoplasty using a dedicated Needle Applicator. An intraoperative CBCT was registered with a pre-op CT. Treatment planning was performed in Radiance using a hybrid Monte Carlo algorithm simulating dose in homogeneous (MCwater) and heterogeneous medium (MChet). Dose distributions on CBCT and pre-op CT were compared with each other. Spinal cord and the metastasis doses were evaluated.
RESULTS: The MCwater calculations showed a spherical dose distribution as expected. The minimum target dose for the MChet simulations on pre-op CT was increased by 40% while the maximum spinal cord dose was decreased by 35%. Due to the artefacts on the CBCT the comparison between MChet simulations on CBCT and pre-op CT showed differences up to 50% in dose.
CONCLUSIONS: igIORT and online treatment planning improves the accuracy of IORT. However, the current set-up is limited by CT artefacts. Fusing an intraoperative CBCT with a pre-op CT allows the combination of an accurate dose calculation with the knowledge of the correct source/applicator position. This method can be also used for pre-operative treatment planning followed by image guided surgery.
Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Intrabeam; Intraoperative radiotherapy; Kypho IORT; Radiance; Treatment planning

Mesh:

Year:  2017        PMID: 28535919     DOI: 10.1016/j.ejmp.2017.04.017

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  3 in total

1.  Dosimetric characteristics of the INTRABEAM ® system with spherical applicators in the presence of air gaps and tissue heterogeneities.

Authors:  Eyachew Misganew Tegaw; Somayeh Gholami; Gilnaz Omyan; Ghazale Geraily
Journal:  Radiat Environ Biophys       Date:  2020-03-31       Impact factor: 1.925

2.  A Comparison between Electron Gamma Shower, National Research Council/Easy Particle Propagation (EGSnrc/Epp) and Monte Carlo N-Particle Transport Code (MCNP) in Simulation of the INTRABEAM ® System with Spherical Applicators.

Authors:  E M Tegaw; Gh Geraily; S M Etesami; S Gholami; H Ghanbari; M Farzin; G F Tadesse; M Shojaei
Journal:  J Biomed Phys Eng       Date:  2021-02-01

3.  Treatment Planning in Intraoperative Radiation Therapy (IORT): Where Should We Go?

Authors:  Carlo Cavedon; Renzo Mazzarotto
Journal:  Cancers (Basel)       Date:  2022-07-20       Impact factor: 6.575

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.