Literature DB >> 9623663

Comparison of EGS4 and MCNP Monte Carlo codes when calculating radiotherapy depth doses.

P A Love1, D G Lewis, I A Al-Affan, C W Smith.   

Abstract

The Monte Carlo codes EGS4 and MCNP have been compared when calculating radiotherapy depth doses in water. The aims of the work were to study (i) the differences between calculated depth doses in water for a range of monoenergetic photon energies and (ii) the relative efficiency of the two codes for different electron transport energy cut-offs. The depth doses from the two codes agree with each other within the statistical uncertainties of the calculations (1-2%). The relative depth doses also agree with data tabulated in the British Journal of Radiology Supplement 25. A discrepancy in the dose build-up region may by attributed to the different electron transport algorithims used by EGS4 and MCNP. This discrepancy is considerably reduced when the improved electron transport routines are used in the latest (4B) version of MCNP. Timing calculations show that EGS4 is at least 50% faster than MCNP for the geometries used in the simulations.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9623663     DOI: 10.1088/0031-9155/43/5/024

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  1 in total

1.  A Comparison between Electron Gamma Shower, National Research Council/Easy Particle Propagation (EGSnrc/Epp) and Monte Carlo N-Particle Transport Code (MCNP) in Simulation of the INTRABEAM ® System with Spherical Applicators.

Authors:  E M Tegaw; Gh Geraily; S M Etesami; S Gholami; H Ghanbari; M Farzin; G F Tadesse; M Shojaei
Journal:  J Biomed Phys Eng       Date:  2021-02-01
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.