Literature DB >> 16798417

Reporting and analyzing statistical uncertainties in Monte Carlo-based treatment planning.

Indrin J Chetty1, Mihaela Rosu, Marc L Kessler, Benedick A Fraass, Randall K Ten Haken, Feng-Ming Spring Kong, Daniel L McShan.   

Abstract

PURPOSE: To investigate methods of reporting and analyzing statistical uncertainties in doses to targets and normal tissues in Monte Carlo (MC)-based treatment planning. METHODS AND MATERIALS: Methods for quantifying statistical uncertainties in dose, such as uncertainty specification to specific dose points, or to volume-based regions, were analyzed in MC-based treatment planning for 5 lung cancer patients. The effect of statistical uncertainties on target and normal tissue dose indices was evaluated. The concept of uncertainty volume histograms for targets and organs at risk was examined, along with its utility, in conjunction with dose volume histograms, in assessing the acceptability of the statistical precision in dose distributions. The uncertainty evaluation tools were extended to four-dimensional planning for application on multiple instances of the patient geometry. All calculations were performed using the Dose Planning Method MC code.
RESULTS: For targets, generalized equivalent uniform doses and mean target doses converged at 150 million simulated histories, corresponding to relative uncertainties of less than 2% in the mean target doses. For the normal lung tissue (a volume-effect organ), mean lung dose and normal tissue complication probability converged at 150 million histories despite the large range in the relative organ uncertainty volume histograms. For "serial" normal tissues such as the spinal cord, large fluctuations exist in point dose relative uncertainties.
CONCLUSIONS: The tools presented here provide useful means for evaluating statistical precision in MC-based dose distributions. Tradeoffs between uncertainties in doses to targets, volume-effect organs, and "serial" normal tissues must be considered carefully in determining acceptable levels of statistical precision in MC-computed dose distributions.

Entities:  

Mesh:

Year:  2006        PMID: 16798417     DOI: 10.1016/j.ijrobp.2006.03.039

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  15 in total

1.  Dosimetric characteristics of the INTRABEAM ® system with spherical applicators in the presence of air gaps and tissue heterogeneities.

Authors:  Eyachew Misganew Tegaw; Somayeh Gholami; Gilnaz Omyan; Ghazale Geraily
Journal:  Radiat Environ Biophys       Date:  2020-03-31       Impact factor: 1.925

2.  Dosimetric impact of statistical uncertainty on Monte Carlo dose calculation algorithm in volumetric modulated arc therapy using Monaco TPS for three different clinical cases.

Authors:  Mohandass Palanisamy; Khanna David; Manigandan Durai; Narendra Bhalla; Abhishek Puri
Journal:  Rep Pract Oncol Radiother       Date:  2019-02-18

3.  Pitfalls in the beam modelling process of Monte Carlo calculations for proton pencil beam scanning.

Authors:  Carla Winterhalter; Adam Aitkenhead; David Oxley; Jenny Richardson; Damien C Weber; Ranald I MacKay; Antony J Lomax; Sairos Safai
Journal:  Br J Radiol       Date:  2020-02-06       Impact factor: 3.039

4.  Automated Monte-Carlo re-calculation of proton therapy plans using Geant4/Gate: implementation and comparison to plan-specific quality assurance measurements.

Authors:  Adam H Aitkenhead; Peter Sitch; Jenny C Richardson; Carla Winterhalter; Imran Patel; Ranald I Mackay
Journal:  Br J Radiol       Date:  2020-07-29       Impact factor: 3.039

5.  Evaluation of dose prediction error and optimization convergence error in four-dimensional inverse planning of robotic stereotactic lung radiotherapy.

Authors:  Mark K H Chan; Dora L W Kwong; Anthony Tong; Eric Tam; Sherry C Y Ng
Journal:  J Appl Clin Med Phys       Date:  2013-07-08       Impact factor: 2.102

6.  Quantification of internal dosimetry in PET patients: individualized Monte Carlo vs generic phantom-based calculations.

Authors:  Sara Neira; Jacobo Guiu-Souto; Pablo Díaz-Botana; Paulino Pais; Carlos Fernández; Virginia Pubul; Álvaro Ruibal; Cristian Candela-Juan; Araceli Gago-Arias; Miguel Pombar; Juan Pardo-Montero
Journal:  Med Phys       Date:  2020-07-14       Impact factor: 4.071

7.  Verification of a Monte Carlo dose calculation engine in proton minibeam radiotherapy in a passive scattering beamline for preclinical trials.

Authors:  Consuelo Guardiola; Ludovic De Marzi; Yolanda Prezado
Journal:  Br J Radiol       Date:  2020-01-06       Impact factor: 3.039

8.  Quantification of internal dosimetry in PET patients II: Individualized Monte Carlo-based dosimetry for [18F]fluorocholine PET.

Authors:  Sara Neira; Jacobo Guiu-Souto; Paulino Pais; Sofía Rodríguez Martínez de Llano; Carlos Fernández; Virginia Pubul; Álvaro Ruibal; Miguel Pombar; Araceli Gago-Arias; Juan Pardo-Montero
Journal:  Med Phys       Date:  2021-07-29       Impact factor: 4.506

9.  Validation of a Software Upgrade in a Monte Carlo Treatment Planning System by Comparison of Plans in Different Versions.

Authors:  P Mohandass; D Khanna; D Manigandan; Narendra Kumar Bhalla; Abhishek Puri
Journal:  J Med Phys       Date:  2018 Apr-Jun

10.  Feasibility of a GATE Monte Carlo platform in a clinical pretreatment QA system for VMAT treatment plans using TrueBeam with an HD120 multileaf collimator.

Authors:  Boram Lee; Seonghoon Jeong; Kwangzoo Chung; Myonggeun Yoon; Hee Chul Park; Youngyih Han; Sang Hoon Jung
Journal:  J Appl Clin Med Phys       Date:  2019-09-23       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.