| Literature DB >> 33458632 |
Jason Davis1, Bernard Thibault2, Iqwal Mangat3, Benoit Coutu4, Matthew Bennett5, Francois Philippon6, Roopinder Sandhu7, Laurence Sterns8, Vidal Essebag9, Pablo Nery10, George Wells11, Raymond Yee12, Derek Exner13, Andrew Krahn5, Ratika Parkash1.
Abstract
BACKGROUND: Premature or rapid battery depletion may compromise the performance and reliability of an implantable cardioverter defibrillator (ICD), potentially resulting in harm or death to patients. We sought to describe the outcomes and clinical management of devices included in the Abbott ICD Premature Battery Depletion Advisory, using data from a Canadian registry.Entities:
Year: 2020 PMID: 33458632 PMCID: PMC7801196 DOI: 10.1016/j.cjco.2020.09.008
Source DB: PubMed Journal: CJC Open ISSN: 2589-790X
Patient demographics
| Characteristic | n (%) or mean ± SD |
|---|---|
| Patient age (y) | 65.6 ± 11.9 |
| Male | 2160 (80.7) |
| ICD indication | |
| Primary prevention | 1665 (62) |
| Secondary prevention | 879 (33) |
| Unknown | 134 (5) |
| LVEF (%,) | 32.2 ± 13.2 |
| Previous therapy | |
| ATP only | 276 (10.3) |
| Shocks only | 118 (4.4) |
| ATP with shocks | 553 (58.3) |
| Pacing-dependent | 335 (12.5) |
| Remote monitoring | 1716 (64.1) |
| Device models | |
| Fortify | 670 (25.0) |
| Fortify Assura | 1032 (38.5) |
| Quadra Assura | 444 (16.6) |
| Quadra Assura MP | 11 (0.4) |
| Unify | 239 (8.9) |
| Unify Assura | 167 (6.2) |
| Unify Quadra | 115 (4.3) |
ATP, anti-tachycardia pacing; ICD, implantable cardioverter defibrillator; LVEF, left ventricular ejection fraction; SD, standard deviation.
Figure 1Outcomes for patients with Abbott Premature Battery Depletion Advisory device. PBD, premature battery depletion; ERI, elective replacement indicator.
Figure 2Cardiac device survival according to Abbott Battery Performance Alert (BPA) vs device end of life (EOL). P = 0.053.
Abbott advisory premature battery depletion rate by device model and device class
| Model | Failure (n, %) |
|---|---|
| ICD (n = 1701) | 34 (2.0) |
| Fortify (n = 669) | 13 (1.9 |
| Fortify Assura (n = 1032) | 21 (2.0) |
| CRT device (n = 974) | 9 (0.9) |
| Quadra Assura (n = 442) | 7 (1.6) |
| Quadra Assura MP (n = 11) | 0 (0) |
| Unify (n = 239) | 1 (0.4) |
| Unify Assura (n = 167) | 0 (0) |
| Unify Quadra (n = 115) | 1 (0.9) |
CRT, cardiac resynchronization therapy; ICD, implantable cardioverter defibrillator.