Daniel F Craft1,2, Peter Balter1,2, Wendy Woodward2,3, Stephen F Kry1,2, Mohammad Salehpour1,2, Rachel Ger1,2, Mary Peters1,2, Garrett Baltz1,2, Erik Traneus4, Rebecca M Howell1,2.
Abstract
BACKGROUND ANDEntities:
Keywords: 3D printing; Compensator; Electrons; PMRT
Year: 2018 PMID: 33458415 PMCID: PMC7807570 DOI: 10.1016/j.phro.2018.11.005
Source DB: PubMed Journal: Phys Imaging Radiat Oncol ISSN: 2405-6316
Fig. 1Algorithm and compensator planning flow diagram. The four steps of the algorithm are shown, as well as the progression from step one to step two, the iteration between step two and step three until satisfactory plan conditions are met, and the final export and printing of the compensator as shown in step four.
. Comparison of dose metrics between conventional 3D plans and compensator based plans.
| Dose Metric | Average 3D Plan | Average Comp Plan | Average Difference (Comp-3D) | p-value |
|---|---|---|---|---|
| CTV D98 | 39.0 Gy | 44. 6 Gy | 5.6 Gy | <0.001 |
| CTV Mean Dose | 52.3 Gy | 53.1 Gy | 0.7 Gy | >0.1 |
| CTV D2 | 67.2 Gy | 62.5 Gy | −4.7 Gy | 0.06 |
| CTV V45 Gy | 89.1% | 97.1% | 8.0% | <0.001 |
| Heart Mean Dose | 2.8 Gy | 3.0 Gy | 1.6 Gy | >0.1 |
| Heart V30 Gy | 0.01% | 0.3% | 0.3% | 0.06 |
| Ips. Lung D2 | 52.1 Gy | 39.0 Gy | −13.1 Gy | <0.001 |
| Ips. Lung V20 Gy | 30.6% | 34.3% | 3.7% | >0.1 |
| Ips. Lung V40 Gy | 13.3% | 1.7% | −11.6% | <0.001 |
| Total Lung Mean Dose | 7.5 Gy | 7.6 Gy | 0.1 Gy | >0.1 |
| Skin Mean Dose | 40.8 Gy | 52.2 Gy | 11.5 Gy | <0.001 |
| Skin D2 | 57.7 Gy | 62.1 Gy | 4.4 Gy | 0.02 |
| HI | 56.4 | 35.9 | −20.5 | <0.001 |
| 110% Hotspot Volume | 257 cm3 | 170 cm3 | −87 cm3 | 0.08 |
| 130% Hotspot Volume | 32 cm3 | 5 cm3 | −27 cm3 | 0.09 |
| 150% Hotspot Volume | 8 cm3 | 0 cm3 | −8 cm3 | >0.1 |
Fig. 2Cumulative DVH comparison. The average of all compensator plan DVHs are shown with dotted lines, while the average of all conventional plans are shown as solid lines.
Average differences between calculated and measured dose profiles in film.
| Energy (MeV) | Average Error in 80% Isodose Width (mm) | Average Error in 80% Isodose Depth (mm) | Overall Average Error (mm) |
|---|---|---|---|
| 12 | 2.2 | 0.8 | 1.1 |
| 16 | 1.9 | 2.0 | 2.0 |
| 20 | 1.5 | 2.1 | 2.0 |
| Average | 1.9 | 1.7 | 1.7 |
Fig. 3Compensator plan delivery to a phantom. A) Photograph of the compensator in the cerrobend tray. B) Photograph of the compensator and phantom in treatment position, and C) Slice of the TPS view of the plan. In (C) the compensator is contoured in orange, the cerrobend in dark blue, the CTV in light blue, and the heart in pink. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.)
Summary of TLD dose calculation and measurement error.
| Phantom | Average In-field TLD Error | In-field TLD Error SD | Average Out-of-field TLD Error | Out-of-field Error SD | Average Skin TLD Error | Skin Error SD |
|---|---|---|---|---|---|---|
| 1 | 1% | 1% | 16% | 9% | NA | NA |
| 2 | 3% | 1% | 10% | 7% | 2% | 1% |
| 3 | 2% | 1% | 25% | 25% | 5% | 2% |
| 4 | 2% | 1% | 20% | 12% | 3% | 1% |
| 5 | 4% | 2% | 24% | 17% | 4% | 1% |
| 6 | 2% | 2% | 7% | 9% | 2% | 2% |
| Average | 2% | 2% | 17% | 13% | 3% | 1% |