Jinsoo Uh1, Mo Kadbi2, Chia-Ho Hua1.
Abstract
BACKGROUND ANDEntities:
Keywords: Breathing irregularity; Four-dimensional magnetic resonance imaging (4D MRI); Pediatric patients; Prospective gating; Respiration characteristics; Retrospective sorting
Year: 2019 PMID: 33458284 PMCID: PMC7807601 DOI: 10.1016/j.phro.2019.09.004
Source DB: PubMed Journal: Phys Imaging Radiat Oncol ISSN: 2405-6316
Information on the respiratory signals used for the phantom experiments and the performance results.
| Patient | Sex | Age | Source of respiratory waveform | General anesthesia | Resp. rate (breaths/min) | Diaph. motion (mm) | Breathing irregularity | Scan time (min) | Deviation (mm) | |||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Mean | 95th percentile | Max | ||||||||||||
| Pro. | Retro. | Pro. | Retro. | Pro. | Retro. | |||||||||
| 1 | F | 5 | 4D CT | Yes | 28.7 | 5.2 | 7.4 | 4.9 | 0.18 | 0.33 | 0.50 | 1.02 | 1.21 | 1.94 |
| 2 | M | 5 | 4D MRI | Yes | 16.8 | 12.5 | 12.6 | 6.1 | 0.44 | 1.00 | 1.21 | 3.43 | 2.12 | 6.97 |
| 3 | M | 5 | 4D CT | Yes | 22.1 | 4.4 | 8.9 | 4.6 | 0.15 | 0.29 | 0.40 | 0.91 | 0.98 | 1.67 |
| 4 | F | 6 | 4D CT | Yes | 15.6 | 6.3 | 13.2 | 6.5 | 0.20 | 0.30 | 0.49 | 0.93 | 1.08 | 1.86 |
| 5 | M | 6 | 4D CT | Yes | 16.6 | 6.5 | 7.3 | 6.1 | 0.18 | 0.36 | 0.46 | 1.13 | 0.91 | 2.19 |
| 6 | M | 12 | 4D CT | No | 20.6 | 14.8 | 9.7 | 5.3 | 0.41 | 0.80 | 1.07 | 2.48 | 2.27 | 4.86 |
| 7 | F | 12 | 4D MRI | No | 16.9 | 7.5 | 35.8 | 6.6 | 0.34 | 0.68 | 0.89 | 2.40 | 3.10 | 4.61 |
| 8 | F | 13 | 4D CT | No | 18.4 | 10.5 | 5.5 | 5.9 | 0.34 | 0.92 | 0.85 | 3.13 | 2.46 | 6.81 |
| 9 | F | 13 | 4D MRI | No | 25.9 | 6.7 | 29.4 | 4.3 | 0.34 | 0.57 | 0.95 | 1.79 | 2.32 | 4.65 |
| 10 | F | 15 | 4D CT | No | 10.1 | 17.6 | 22.6 | 7.7 | 0.54 | 0.86 | 1.35 | 2.66 | 3.77 | 5.33 |
| 11 | M | 15 | 4D CT | No | 20.5 | 12.2 | 13.9 | 4.7 | 0.41 | 0.83 | 1.07 | 2.44 | 2.08 | 5.27 |
| 12 | M | 15 | 4D CT | No | 14.7 | 9.3 | 28.7 | 6.9 | 0.44 | 0.52 | 1.19 | 1.49 | 3.57 | 3.64 |
| 13 | F | 17 | 4D CT | No | 8.5 | 16.3 | 31.6 | 15.1 | 0.63 | 0.55 | 1.47 | 1.79 | 3.44 | 3.89 |
| 14 | F | 17 | 4D MRI | No | 17.7 | 6.6 | 20.5 | 5.8 | 0.27 | 0.57 | 0.73 | 2.03 | 1.32 | 3.48 |
| 15 | F | 17 | 4D CT | No | 10.9 | 8.5 | 22.8 | 5.9 | 0.33 | 0.51 | 0.89 | 1.46 | 1.63 | 2.41 |
| 16 | M | 17 | 4D MRI | No | 6.1 | 30.2 | 26.3 | 7.7 | 1.62 | 2.22 | 4.56 | 7.13 | 8.43 | 12.93 |
| 17 | F | 17 | 4D CT | No | 21.0 | 4.3 | 21.2 | 5.9 | 0.18 | 0.34 | 0.48 | 1.00 | 0.81 | 2.40 |
| 18 | M | 18 | 4D CT | No | 12.2 | 8.0 | 7.2 | 5.5 | 0.25 | 0.55 | 0.66 | 1.94 | 3.05 | 3.62 |
| 19 | M | 19 | 4D CT | No | 13.1 | 17.1 | 16.8 | 5.9 | 0.53 | 0.97 | 1.22 | 2.60 | 2.81 | 7.28 |
| 20 | M | 20 | 4D CT | No | 12.6 | 13.8 | 11.4 | 6.4 | 0.33 | 0.76 | 0.92 | 2.65 | 2.12 | 6.01 |
| 21 | F | 20 | 4D CT | No | 16.9 | 10.0 | 27.2 | 7.4 | 0.34 | 0.61 | 0.90 | 1.90 | 1.56 | 3.92 |
| 22 | F | 20 | 4D CT | No | 10.6 | 11.6 | 13.5 | 5.2 | 0.35 | 0.82 | 0.82 | 2.40 | 1.33 | 5.96 |
| 23 | M | 24 | 4D CT | No | 6.7 | 18.0 | 52.4 | 12.7 | 1.27 | 1.12 | 3.35 | 3.86 | 14.31 | 8.85 |
| Min. | 5 | 6.1 | 4.3 | 5.5 | 4.3 | 0.15 | 0.29 | 0.40 | 0.91 | 0.81 | 1.67 | |||
| Max. | 24 | 28.7 | 30.2 | 52.4 | 15.1 | 1.62 | 2.22 | 4.56 | 7.13 | 14.31 | 12.93 | |||
| Median | 15 | 16.6 | 10.0 | 16.8 | 5.9 | 0.34 | 0.61 | 0.90 | 2.03 | 2.12 | 4.61 | |||
Abbreviations: Diaph. motion: peak-to-peak diaphragm motion; F: female; M: male; Max.: maximum; Min.: minimum, Pro.: prospective gating; Resp.: respiration, Retro.: retrospective sorting.
Patients younger than 7 years underwent general anesthesia with intravenously administered propofol (150–270 μg/kg/min).
Breathing irregularity was quantified by the coefficient of variation of the peak-to-peak amplitude in the respiratory signal.
Patient-specific clinical parameters and comparison of 4D MRI with 4D CT.
| Patient | Diagnosis | Treatment site | 4D MRI parameters | |||
|---|---|---|---|---|---|---|
| Orientation | FOV (mm2) | No. of slices | Resolution (mm3) | |||
| 2 | Neuroblastoma | Chest apex | Coronal | 300 × 300 | 44 | 1.5 × 1.5 × 4.0 |
| 4 | Neuroblastoma | Adrenal | Transverse | 300 × 300 | 42 | 1.8 × 1.8 × 4.0 |
| 7 | Rhabdomyosarcoma | Porta hepatis | Coronal | 335 × 307 | 40 | 1.5 × 1.5 × 5.0 |
| 9 | Rhabdomyosarcoma | Chest wall | Sagittal | 300 × 220 | 30 | 2.0 × 2.0 × 6.0 |
| 14 | Rhabdomyosarcoma | Mediastinum | Coronal | 291 × 361 | 46 | 1.5 × 1.5 × 5.0 |
| 16 | Hodgkin lymphoma | Mediastinum | Coronal | 350 × 344 | 52 | 1.5 × 1.5 × 4.0 |
Abbreviations: 4D CT: four-dimensional computed tomography; 4D MRI: four-dimensional magnetic resonance imaging; CTV: clinical target volume; FOV: field of view; ITV: internal target volume.
Breathing irregularity was quantified by the coefficient of variation of the peak-to-peak amplitude in the respiratory signal.
Fig. 1Correlations between respiratory characteristics (rate, amplitude, and irregularity) and the performance of prospectively gated 4D MRI (scan time and mean deviation). Breathing irregularity indicates the coefficient of variation of the peak-to-peak amplitude in the respiratory waveforms.
Fig. 2Examples of 4D MRI scans prospectively gated from coronal (A and B) or transverse (C) image slices. Images of a patient with relatively small organ motion (Patient 7; peak-to-peak diaphragm motion: 7.5 mm) (A) showed no apparent sorting artifacts, whereas images of another patient with greater motion (Patient 16; peak-to-peak diaphragm motion: 30.2 mm) (B) did show artifacts. The signal from liver was dark for a patient (Patient 4) who had received multiple blood transfusions, possibly because of iron overload (C).
Fig. 3Comparison of sorting accuracy with prospective gating and retrospective sorting for a representative case (Patient 11). The solid line indicates the mean displacement at each phase, the deviation from which was calculated to evaluate the sorting error.