Beat-to-beat blood pressure measurement using a cuffless device does not accurately reflect invasive blood pressure.

Mohammed A Moharram1, Luke C Wilson1, Michael Ja Williams1, Sean Coffey1.   

Abstract

BACKGROUND: The availability of an accurate continuous cuffless blood pressure (BP) monitor would provide an alternative to both invasive continuous BP and 24-h intermittent cuff-based BP monitors. We investigated the accuracy of a cuffless beat to beat (BtB) device compared to both invasive BP (iBP) and brachial cuff BP (cBP) measurements.
METHODS: Patients undergoing clinically indicated coronary angiography (CA) and/or percutaneous coronary intervention (PCI) were recruited. After calibration to an initial cBP reading, BP was measured simultaneously using a BtB device (SOMNOtouch NIBP), brachial artery iBP, and cBP at two time points.
RESULTS: The study was terminated early due to a significant bias. Recordings from 14 participants (11 males, mean age 68.4 years) were analysed. Readings from BtB BP were higher than iBP. The bias between BtB BP and iBP was 34.3 mmHg (95%CI: 27.0, 41.5) and 23.6 mmHg (95%CI: 16.8, 30.4) for SBP and DBP respectively. A similar bias was seen between BtB BP and cBP, but cBP and iBP were largely in agreement.
CONCLUSIONS: In patients undergoing CA/PCI, significant differences were detected between BtB BP and both invasively measured and cuff BP. The non-invasive BtB BP measurement device tested is not suitable for clinical or research use.
© 2020 The Authors.

Entities:  

Keywords:  BP, blood pressure; BtB, beat to beat; Continuous blood pressure monitor; Cuffless blood pressure; DBP, diastolic blood pressure; PTT, pulse transit time; Pulse arrival time; SBP, systolic blood pressure; SOMNOtouch NIBP; cBP, brachial cuff blood pressure; iBP, invasive blood pressure

Year:  2020        PMID: 33447759      PMCID: PMC7803068          DOI: 10.1016/j.ijchy.2020.100030

Source DB:  PubMed          Journal:  Int J Cardiol Hypertens        ISSN: 2590-0862


  21 in total

1.  Continuous blood pressure measurement using the pulse transit time: Comparison to intra-arterial measurement.

Authors:  Andreas Patzak; Yuri Mendoza; Heiko Gesche; Martin Konermann
Journal:  Blood Press       Date:  2015-04-10       Impact factor: 2.835

2.  An examination of calibration intervals required for accurately tracking blood pressure using pulse transit time algorithms.

Authors:  B M McCarthy; C J Vaughan; B O'Flynn; A Mathewson; C Ó Mathúna
Journal:  J Hum Hypertens       Date:  2013-05-23       Impact factor: 3.012

Review 3.  A Literature Review on Current and Proposed Technologies of Noninvasive Blood Pressure Measurement.

Authors:  Ramtanu Mukherjee; Sanchita Ghosh; Bharat Gupta; Tapas Chakravarty
Journal:  Telemed J E Health       Date:  2017-08-07       Impact factor: 3.536

4.  Accuracy of 24-hour ambulatory blood pressure monitoring by a novel cuffless device in clinical practice.

Authors:  Philipp Krisai; Annina Salome Vischer; Leo Kilian; Andrea Meienberg; Michael Mayr; Thilo Burkard
Journal:  Heart       Date:  2018-09-18       Impact factor: 5.994

Review 5.  Assessment and management of blood-pressure variability.

Authors:  Gianfranco Parati; Juan E Ochoa; Carolina Lombardi; Grzegorz Bilo
Journal:  Nat Rev Cardiol       Date:  2013-02-12       Impact factor: 32.419

6.  [Validation of continuous blood pressure measurements by pulse transit time: a comparison with invasive measurements in a cardiac intensive care unit].

Authors:  S Bartsch; D Ostojic; H Schmalgemeier; T Bitter; N Westerheide; S Eckert; D Horstkotte; O Oldenburg
Journal:  Dtsch Med Wochenschr       Date:  2010-11-24       Impact factor: 0.628

7.  Brachial and Radial Systolic Blood Pressure Are Not the Same.

Authors:  Matthew K Armstrong; Martin G Schultz; Dean S Picone; J Andrew Black; Nathan Dwyer; Philip Roberts-Thomson; James E Sharman
Journal:  Hypertension       Date:  2019-05       Impact factor: 10.190

8.  Toward Ubiquitous Blood Pressure Monitoring via Pulse Transit Time: Theory and Practice.

Authors:  Ramakrishna Mukkamala; Jin-Oh Hahn; Omer T Inan; Lalit K Mestha; Chang-Sei Kim; Hakan Töreyin; Survi Kyal
Journal:  IEEE Trans Biomed Eng       Date:  2015-06-05       Impact factor: 4.538

9.  Weighing Scale-Based Pulse Transit Time is a Superior Marker of Blood Pressure than Conventional Pulse Arrival Time.

Authors:  Stephanie L-O Martin; Andrew M Carek; Chang-Sei Kim; Hazar Ashouri; Omer T Inan; Jin-Oh Hahn; Ramakrishna Mukkamala
Journal:  Sci Rep       Date:  2016-12-15       Impact factor: 4.379

Review 10.  The use of photoplethysmography for assessing hypertension.

Authors:  Mohamed Elgendi; Richard Fletcher; Yongbo Liang; Newton Howard; Nigel H Lovell; Derek Abbott; Kenneth Lim; Rabab Ward
Journal:  NPJ Digit Med       Date:  2019-06-26
View more
  2 in total

1.  Elucidation of obstructive sleep apnoea related blood pressure surge using a novel continuous beat-to-beat blood pressure monitoring system.

Authors:  Younghoon Kwon; Martin Baruch; Patrick L Stafford; Heather Bonner; Yeilim Cho; Sula Mazimba; Jeongok G Logan; Daichi Shimbo; Sung-Ha Park; Gen-Min Lin; Ali Azarbarzin; David A Calhoun; Richard Berry; Robert M Carey
Journal:  J Hypertens       Date:  2022-03-01       Impact factor: 4.844

2.  Tracking of the beat-to-beat blood pressure changes by the Caretaker physiological monitor against invasive central aortic measurement.

Authors:  Younghoon Kwon; Patrick L Stafford; Martin C Baruch; Sung-Hoon Kim; Yeilim Cho; Sula Mazimba; Lawrence W Gimple
Journal:  Blood Press Monit       Date:  2022-02-01       Impact factor: 1.444

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.