Continuous blood pressure measurement using the pulse transit time: Comparison to intra-arterial measurement.

Andreas Patzak1, Yuri Mendoza, Heiko Gesche, Martin Konermann.   

Abstract

Continuous blood pressure (BP) measurement allows the investigation of transient changes in BP and thus may give insights into mechanisms of BP control. We validated a continuous, non-invasive BP measurement based on the pulse transit time (PTT), i.e., BP(PTT), by comparing it with the intra-arterial BP (BP(i.a.)) measurement. Twelve subjects (five females and seven males) were included. BP(i.a.) was obtained from the radial artery using a system from ReCor Medical. Systolic and diastolic BP were calculated using the PTT (BP(PTT), SOMNOscreen). (PTT) was determined from the electrocardiogram and the peripheral pulse wave. The BP was modulated by application of increasing doses of dobutamine (5, 10, 20 μg/kg body mass). Systolic BP(PTT) and systolic BP(i.a.) correlated significantly (R = 0.94). The limits of agreement in the Bland-Altman plot were ± 19 mmHg; the mean values differed by 1 mmHg. The correlation coefficient for the diastolic BP measurements was R = 0.42. The limits of agreement in the Bland-Altman plot were ± 18 mmHg, with a mean difference of 5 mmHg in favour of the BP(PTT). The study demonstrates a significant correlation between the measurement methods for systolic BP. The results encourage the application of PTT-based BP measurement for the evaluation of BP dynamics and pathological BP changes.

Entities:  

Keywords:  Blood pressure; pulse transit time; validation

Mesh:

Year:  2015        PMID: 25857601     DOI: 10.3109/08037051.2015.1030901

Source DB:  PubMed          Journal:  Blood Press        ISSN: 0803-7051            Impact factor:   2.835


  18 in total

1.  Interrelationships between pulse arrival time and arterial blood pressure during postural transitions before and after spaceflight.

Authors:  Katelyn N Wood; Danielle K Greaves; Richard L Hughson
Journal:  J Appl Physiol (1985)       Date:  2019-08-15

2.  Nocturnal blood pressure fluctuations measured by using pulse transit time in patients with severe obstructive sleep apnea syndrome.

Authors:  Jennifer Gehring; Heiko Gesche; Gesine Drewniok; Gert Küchler; Andreas Patzak
Journal:  Sleep Breath       Date:  2017-08-22       Impact factor: 2.816

3.  Pulse transit time in pregnancy: a new way to diagnose and classify sleep disordered breathing?

Authors:  Brittany N Link; Celine Eid; Maggie H Bublitz; Martino F Pengo; Myriam Salameh; Karin S Ludwig; Richard P Millman; Lance Dworkin; Ghada Bourjeily
Journal:  Sleep       Date:  2019-05-01       Impact factor: 5.849

4.  The B-Score is a novel metric for measuring the true performance of blood pressure estimation models.

Authors:  Tomas L Bothe; Andreas Patzak; Niklas Pilz
Journal:  Sci Rep       Date:  2022-07-16       Impact factor: 4.996

5.  Introducing Contactless Blood Pressure Assessment Using a High Speed Video Camera.

Authors:  In Cheol Jeong; Joseph Finkelstein
Journal:  J Med Syst       Date:  2016-01-20       Impact factor: 4.460

Review 6.  Blood pressure monitoring in sleep: time to wake up.

Authors:  Younghoon Kwon; Patrick L Stafford; Diane C Lim; Sungha Park; Sung-Hoon Kim; Richard B Berry; David A Calhoun
Journal:  Blood Press Monit       Date:  2020-04       Impact factor: 1.430

7.  Pulse arrival time, a novel sleep cardiovascular marker: the multi-ethnic study of atherosclerosis.

Authors:  Younghoon Kwon; Christopher Wiles; B Eugene Parker; Brian R Clark; Min-Woong Sohn; Sara Mariani; Jin-Oh Hahn; David R Jacobs; James H Stein; Joao Lima; Vishesh Kapur; Andrew Wellman; Susan Redline; Ali Azarbarzin
Journal:  Thorax       Date:  2021-04-16       Impact factor: 9.102

8.  Quantitative Assessment of the Impact of Blood Pulsation on Intraocular Pressure Measurement Results in Healthy Subjects.

Authors:  Robert Koprowski; Lei Tian
Journal:  J Ophthalmol       Date:  2017-01-30       Impact factor: 1.909

9.  Nocturnal blood pressure and nocturnal blood pressure fluctuations: the effect of short-term CPAP therapy and their association with the severity of obstructive sleep apnea.

Authors:  Frauke Picard; Petroula Panagiotidou; Anne-Beke Tammen; Anamaria Wolf-Pütz; Maximilian Steffen; Hanno Julian Gerhardy; Sebastian Waßenberg; Rolf Michael Klein
Journal:  J Clin Sleep Med       Date:  2022-02-01       Impact factor: 4.062

10.  Cuffless blood pressure estimation using only a smartphone.

Authors:  Kenta Matsumura; Peter Rolfe; Sogo Toda; Takehiro Yamakoshi
Journal:  Sci Rep       Date:  2018-05-08       Impact factor: 4.379

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.