Tracking of the beat-to-beat blood pressure changes by the Caretaker physiological monitor against invasive central aortic measurement.

Younghoon Kwon1, Patrick L Stafford2, Martin C Baruch3, Sung-Hoon Kim4, Yeilim Cho1, Sula Mazimba2, Lawrence W Gimple2.   

Abstract

OBJECTIVE: There is an unmet need for noninvasive continuous blood pressure (BP) monitoring technologies in various clinical settings. Continuous and noninvasive central aortic BP monitoring is technically not feasible currently, but if realized, would provide more accurate and real-time global hemodynamic information than any form of peripheral arterial BP monitoring in an acute care setting. As part of our efforts to develop such, herein we examined the tracking correlation between noninvasively-derived peripheral arterial BP by Caretaker device against invasively measured central aortic BP.
METHODS: Beat-to-beat BP by Caretaker was recorded simultaneously with central aortic BP measured in patients undergoing cardiac catheterization. Pearson's correlation was also derived for SBP and DBP. A trend comparison analysis of the beat-to-beat BP change was performed using a four-quadrant plot analysis with the exclusion zones of 0.5 mmHg/s to determine concordance, (i.e. the direction of beat-to-beat changes in SBP and DBP).
RESULTS: A total of 47 patients were included in the study. A total of 31 369 beats representing an average of 17.3 min of recording were used for analysis. The trend analysis yielded concordances of 84.4 and 83.5% for SBP and DBP, respectively. Respective correlations (Pearson's r) for SBP and DBP trends were 0.87 and 0.86 (P < 0.01). Tracking of beat-to-beat BP by Caretaker showed excellent concordance and correlation in the direction and the degree of BP change with central aortic BP, respectively.
CONCLUSION: This study supports the satisfactory performance of the Caretaker device in continuous tracking of central aortic BP beat-to-beat BP and provides a basis to develop an algorithm for absolute central aortic BP estimation in the future.
Copyright © 2021 Wolters Kluwer Health, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2022        PMID: 34569988      PMCID: PMC8741634          DOI: 10.1097/MBP.0000000000000568

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  26 in total

1.  Comparison of simultaneously recorded central and peripheral arterial pressure pulses during rest, exercise and tilted position in man.

Authors:  E J KROEKER; E H WOOD
Journal:  Circ Res       Date:  1955-11       Impact factor: 17.367

2.  The Accuracy of the CNAP® Device Compared with Invasive Radial Artery Measurements for Providing Continuous Noninvasive Arterial Blood Pressure Readings at a Medical Intensive Care Unit: A Method-Comparison Study.

Authors:  Karl-Heinz Smolle; Martin Schmid; Helga Prettenthaler; Christian Weger
Journal:  Anesth Analg       Date:  2015-12       Impact factor: 5.108

3.  Difference between central and peripheral blood pressure in healthy and hypertension-complicated pregnancy.

Authors:  Ludwina Szczepaniak-Chicheł; Wiesław Markwitz; Andrzej Tykarski
Journal:  Blood Press Monit       Date:  2016-04       Impact factor: 1.444

4.  Differential impact of blood pressure-lowering drugs on central aortic pressure and clinical outcomes: principal results of the Conduit Artery Function Evaluation (CAFE) study.

Authors:  Bryan Williams; Peter S Lacy; Simon M Thom; Kennedy Cruickshank; Alice Stanton; David Collier; Alun D Hughes; H Thurston; Michael O'Rourke
Journal:  Circulation       Date:  2006-02-13       Impact factor: 29.690

Review 5.  Role of pulse pressure amplification in arterial hypertension: experts' opinion and review of the data.

Authors:  Alberto P Avolio; Luc M Van Bortel; Pierre Boutouyrie; John R Cockcroft; Carmel M McEniery; Athanase D Protogerou; Mary J Roman; Michel E Safar; Patrick Segers; Harold Smulyan
Journal:  Hypertension       Date:  2009-06-29       Impact factor: 10.190

6.  The accuracy of central SBP determined from the second systolic peak of the peripheral pressure waveform.

Authors:  Stacey S Hickson; Mark Butlin; Fraz A Mir; Johann Graggaber; Joseph Cheriyan; Fakhar Khan; Andrew A Grace; John R Cockcroft; Ian B Wilkinson; Carmel M McEniery
Journal:  J Hypertens       Date:  2009-09       Impact factor: 4.844

Review 7.  Continuous noninvasive pulse wave analysis using finger cuff technologies for arterial blood pressure and cardiac output monitoring in perioperative and intensive care medicine: a systematic review and meta-analysis.

Authors:  Bernd Saugel; Phillip Hoppe; Julia Y Nicklas; Karim Kouz; Annmarie Körner; Julia C Hempel; Jaap J Vos; Gerhard Schön; Thomas W L Scheeren
Journal:  Br J Anaesth       Date:  2020-05-29       Impact factor: 9.166

8.  Central versus ambulatory blood pressure in the prediction of all-cause and cardiovascular mortalities.

Authors:  Chi-Ming Huang; Kang-Ling Wang; Hao-Min Cheng; Shao-Yuan Chuang; Shih-Hsien Sung; Wen-Chung Yu; Chih-Tai Ting; Edward G Lakatta; Frank Cp Yin; Pesus Chou; Chen-Huan Chen
Journal:  J Hypertens       Date:  2011-03       Impact factor: 4.844

9.  Central blood pressure for the management of hypertension: Is it a practical clinical tool in current practice?

Authors:  Hao-Min Cheng; Shao-Yuan Chuang; Tzung-Dau Wang; Kazuomi Kario; Peera Buranakitjaroen; Yook-Chin Chia; Romeo Divinagracia; Satoshi Hoshide; Huynh Van Minh; Jennifer Nailes; Sungha Park; Jinho Shin; Saulat Siddique; Jorge Sison; Arieska Ann Soenarta; Guru Prasad Sogunuru; Apichard Sukonthasarn; Jam Chin Tay; Boon Wee Teo; Yuda Turana; Narsingh Verma; Yuqing Zhang; Ji-Guang Wang; Chen-Huan Chen
Journal:  J Clin Hypertens (Greenwich)       Date:  2019-12-16       Impact factor: 3.738

10.  Pulse Decomposition Analysis of the digital arterial pulse during hemorrhage simulation.

Authors:  Martin C Baruch; Darren Er Warburton; Shannon Sd Bredin; Anita Cote; David W Gerdt; Charles M Adkins
Journal:  Nonlinear Biomed Phys       Date:  2011-01-12
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.