Zhengzheng Xu1, Gregory Warrell1, Soyoung Lee1, Valdir Colussi1,2, Yiran Zheng1, Rodney Ellis1,2, Mitchell Machtay1,2, Tarun Podder1,2.
Abstract
PURPOSE: Dosimetric accuracy is critical when switching a patient treated with stereotactic body radiation therapy (SBRT) or stereotactic fractionated radiotherapy (SRT) among beam-matched linacs. In this study, the dose delivery accuracy of volumetric modulated arc therapy (VMAT) plans for SBRT/SRT patients were evaluated on three beam-matched linacs.Entities:
Keywords: zzm321990SBRTzzm321990; beam-matching; small field dosimetry
Mesh:
Year: 2018 PMID: 30402983 PMCID: PMC6333115 DOI: 10.1002/acm2.12492
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Illustration of ArcCHECK with (a) film holder and (b) A1SL micro ion chamber inserted for planar and point dose measurements.
Output factors for 6 MV photon beam of different field sizes for three beam‐matched linacs
| Field size (cm × cm) | Linac 1 | Linac 2 | Linac 3 | Maximal difference (%) |
|---|---|---|---|---|
| 1 × 1 | 0.699 | 0.705 | 0.708 | 1.3 |
| 2 × 2 | 0.808 | 0.806 | 0.810 | 0.5 |
| 3 × 3 | 0.847 | 0.845 | 0.844 | 0.4 |
| 4 × 4 | 0.876 | 0.876 | 0.875 | 0.1 |
| 5 × 5 | 0.905 | 0.907 | 0.906 | 0.2 |
| 10 × 10 | 1.000 | 1.000 | 1.000 | 0.0 |
| 20 × 20 | 1.093 | 1.095 | 1.094 | 0.1 |
| 30 × 30 | 1.136 | 1.137 | 1.137 | 0.1 |
Difference in PDD10 for beams and MLC leaf transmission factors among beam‐matched linacs
| 6 MV | Linac 1 | Linac 2 | Linac 3 | Maximal difference |
|---|---|---|---|---|
| PDD10 (2 × 2 cm2) | 60.5% | 60.6% | 60.1% | 0.5% |
| PDD10 (5 × 5 cm2) | 63.9% | 63.3% | 63.9% | 0.6% |
| PDD10 (10 × 10 cm2) | 67.3% | 67.3% | 67.4% | 0.2% |
| PDD10 (30 × 30 cm2) | 71.2% | 71.5% | 71.6% | 0.4% |
| MLC leaf transmission | 0.7% | 0.4% | 0.6% | 0.3% |
Beam flatness and symmetry of different field sizes and depths for beam‐matched linacs
| Linac 1 | Linac 2 | Linac 3 | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Crossplane | Inplane | Crossplane | Inplane | Crossplane | Inplane | |||||||
| d: 1.5 cm | Flatness (%) | Symmetry (%) | Flatness (%) | Symmetry (%) | Flatness (%) | Symmetry (%) | Flatness (%) | Symmetry (%) | Flatness (%) | Symmetry (%) | Flatness (%) | Symmetry (%) |
| 2 × 2 | 6.8 | 0.6 | 9.5 | 0.7 | 6.2 | 0.6 | 11.0 | 0.3 | 5.2 | 0.5 | 11.3 | 0.4 |
| 5 × 5 | 2 | 0.2 | 2 | 0.2 | 2.5 | 0.0 | 1.8 | 0.5 | 1.2 | 0.6 | 2.1 | 0.1 |
| 10 × 10 | 0.4 | 0.3 | 1.4 | 0.1 | 1.8 | 0.4 | 1.8 | 0.6 | 0.7 | 0.3 | 0.8 | 0.1 |
| 30 × 30 | 3 | 0.2 | 3.9 | 0.5 | 2.0 | 0.1 | 3.0 | 0.1 | 2.5 | 0.1 | 2.9 | 0.1 |
| d: 10 cm | ||||||||||||
| 2 × 2 | 7.2 | 0.4 | 10.3 | 0.7 | 6.5 | 0.4 | 11.6 | 0.6 | 5.8 | 0.6 | 11.9 | 0.0 |
| 5 × 5 | 2.8 | 0.6 | 3.2 | 0 | 2.8 | 0.8 | 3.7 | 0.5 | 2.2 | 0.4 | 3.3 | 0.1 |
| 10 × 10 | 2.1 | 0.8 | 2.3 | 0.6 | 1.9 | 0.0 | 2.6 | 0.1 | 2.3 | 0.2 | 2.0 | 0.0 |
| 30 × 30 | 1.8 | 0.5 | 1.4 | 0.2 | 1.7 | 0.1 | 1.6 | 0.2 | 1.4 | 0.0 | 1.8 | 0.2 |
d, depth.
Figure 2Distributions of gamma passing rate variation of SBRT/SRT VMAT plans delivered on three beam‐matched linacs and measured with ArcCHECK on three different days.
Figure 3Distributions of absolute dose gamma passing rates of SBRT/SRT VMAT plans delivered on three beam‐matched linacs. (a) Passing rates of ArcCHECK measurements; (b) Passing rates of EBT3 film measurements. Solid red line: 95% passing rate threshold for 3%/2 mm gamma criteria. Dashed red line: 90% passing rate threshold for 2%/2 mm gamma criteria.
Difference in passing rates of ArcCHECK and film measurements and point doses among beam‐matched linacs
| L1 vs L2 | L1 vs L3 | L2 vs L3 | Mean ± SD | |
|---|---|---|---|---|
| 3%/2 mm (ArcCHECK) | 0.2 ± 0.5% | 0.3 ± 0.6% | 0.2 ± 0.3% | 0.3 ± 0.5% |
| r+ | 0.15 | 0.23 | 0.19 | – |
| 2%/2 mm (ArcCHECK) | 0.6 ± 1.8% | 0.9 ± 2.1% | 0.8 ± 1.5% | 0.8 ± 1.7% |
| r | 0.19 | 0.20 | 0.24 | – |
| 3%/2 mm (Film) | 1.2 ± 1.8% | 1.7 ± 1.5% | 1.5 ± 1.2% | 1.6 ± 1.3% |
| r | 0.23 | 0.14 | 0.26 | – |
| 2%/2 mm (Film) | 2.3 ± 1.6% | 1.9 ± 2.2% | 2.5 ± 1.7% | 2.2 ± 1.9% |
| r | 0.20 | 0.16 | 0.21 | – |
| Point dose difference 1 | −1.3 ± 0.8% | −2.1 ± 0.6% | −0.9 ± 1.1% | −1.5 ± 0.8% |
| r | 0.22 | 0.17 | 0.15 | – |
| Point dose difference 2 | −0.1 ± 0.5% | 0.2 ± 0.3% | 0.1 ± 0.3% | 0.1 ± 0.5% |
| r | 0.14 | 0.21 | 0.11 | – |
L1, linac 1; L2, linac 2; L3, linac 3; (+) r value, Pearson correlation coefficient.
Point dose difference 1 is the difference between measured dose with ion chamber and TPS calculation.
Point dose difference 2 is the difference in measured doses between two beam‐matched linacs.