| Literature DB >> 33437308 |
Paweł Czajkowski1, Grzegorz Zwierzchowski2,3, Tomasz Piotrowski2,3.
Abstract
PURPOSE: The aim of the study was to assess selected methods of image registration available in the RaySearch software and their impact on the accuracy of mapping of doses deposited in the bladder during brachytherapy (BRT) of cervical cancer in images used during external beam radiotherapy (EBRT).Entities:
Keywords: brachytherapy; cervix; deformable image registration; dose mapping
Year: 2020 PMID: 33437308 PMCID: PMC7787205 DOI: 10.5114/jcb.2020.101693
Source DB: PubMed Journal: J Contemp Brachytherapy ISSN: 2081-2841
Volumes of the bladder based on the contours made during treatment planning in the BrachyVision and Eclipse systems, and reconstructed after data migration to the RaySearch system
| Bladder (cm3) | ||||||
|---|---|---|---|---|---|---|
| Patient | CBCT images | CT images | ||||
| BV | RS | Δ(RS-BV) | E | RS | Δ(RS-E) | |
| 1 | 56.5 | 57.9 | 1.3 | 91.6 | 92.6 | 1.0 |
| 2 | 87.9 | 88.3 | 0.5 | 209.8 | 209.4 | –0.4 |
| 3 | 127.5 | 128.4 | 1.0 | 34.8 | 35.4 | 0.6 |
| 4 | 111.3 | 112.0 | 0.7 | 124.1 | 124.3 | 0.2 |
| 5 | 46.4 | 46.6 | 0.2 | 354.5 | 355.0 | 0.5 |
| 6 | 37.9 | 38.3 | 0.4 | 56.3 | 56.6 | 0.3 |
| 7 | 61.2 | 62.4 | 1.2 | 51.3 | 52.7 | 1.4 |
| 8 | 74.5 | 75.2 | 0.7 | 215.9 | 218.1 | 2.2 |
| 9 | 48.1 | 48.5 | 0.4 | 70.2 | 71.6 | 1.3 |
| 10 | 203.6 | 204.2 | 0.6 | 236.4 | 237.7 | 1.3 |
BV – BrachyVision, RS – RaySearch, E – Eclipse
Fig. 1Bladder volumes during BRT (CBCT images) and during EBRT (CT images)
Fig. 2DSC for bladder obtained for the four methods of registering CT images from CBCT
Number of observations according to the extent of DSC values for the four registration methods of CT images from CBCT
| Number of observations | ||||
|---|---|---|---|---|
| DSC (range) | Rigid | ANACONDA | MORFEUS | |
| Reg_1 | Reg_2a | Reg_2b | Reg_3 | |
| [0.00, 0.70] | 10 | 1 | 1 | 0 |
| [0.70, 0.90] | 0 | 2 | 2 | 0 |
| [0.90, 0.95] | 0 | 3 | 2 | 1 |
| [0.95, 1.00] | 0 | 4 | 5 | 9 |
Fig. 3Differences between corresponding points (from reference and deformed meshes) for three mesh-based registrations
Doses measured for bladder on CBCT images (initial BRT plan) and corresponding doses mapped by three registration methods, and displayed for bladder on CT images
| Maximum dose (Gy) measured in 2 cm3 of the bladder | ||||
|---|---|---|---|---|
| Patient | CBCT images# | CT images@ | ||
| Reg_2a | Reg_2b | Reg_3 | ||
| 1 | 5.20 | 5.26 | 5.36 | 5.32 |
| 2 | 9.43 | 10.60 | 11.05 | 11.84 |
| 3 | 6.54 | 6.78 | 7.21 | 7.46 |
| 4 | 3.62 | 3.98 | 3.75 | 3.41 |
| 5 | 9.25 | 19.75 | 12.09 | 12.05 |
| 6 | 4.99 | 4.33 | 4.53 | 4.83 |
| 7 | 2.94 | 2.04 | 2.28 | 2.96 |
| 8 | 5.97 | 7.38 | 7.32 | 6.83 |
| 9 | 3.32 | 3.76 | 3.71 | 3.44 |
| 10 | 7.50 | 7.67 | 7.83 | 7.64 |
BRT doses measured in bladder contoured on CBCT; @BRT doses, mapped on CT and measured in bladder contoured on CT
Fig. 4Differences between planned (CBCT) and mapped (CT) doses measured in 2 cm3 of the bladder
Fig. 5CBCT image of region treated during brachytherapy
Fig. 6Alignment of bladder contour from CBCT image (dashed line) with the corresponding reference contour delineated in the CT image (continuous line) for A) rigid registration, Reg_1, B) hybrid registration, Reg_2b, and C) biomechanical registration, Reg_3
Fig. 7Example of errors in dose mapping procedure