Literature DB >> 33413463

Mid-term subsidence and periprosthetic radiolucency of the AMIStem: a 5-year EBRA-FCA analysis.

Julian Hasler1, Andreas Flury2, Dimitris Dimitriou2, Iris Holweg2, Naeder Helmy2, Michael Finsterwald2.   

Abstract

BACKGROUND: There has been an evolution in cementless total hip arthroplasty (THA) with newer short stem designs aimed to preserve metaphyseal bone stock and facilitate implantation through minimally invasive approaches. While early subsidence has been correlated to aseptic loosening in conventional stems, there is a paucity of data regarding short stems. The current study aims to report on stem subsidence and mid-term clinical outcomes of a cementless, metaphyseal-anchored short femoral stem, specifically designed for the direct anterior approach (DAA).
METHODS: Ninety-four consecutive patients (100 hips) with a minimum follow-up of 5 years following cementless THA were included in this single-center retrospective study. Subsidence was evaluated using the "Ein-Bild-Roentgen-Analyse" (EBRA). Periprosthetic radiolucency allocated to the zones of Charnley and Gruen was assessed. Additionally, demographic and implant-related factors potentially associated with increased subsidence and clinical outcomes were evaluated.
RESULTS: At the last follow-up, the average stem subsidence was 1.98 ± 1.20 mm, with 48% of the implants demonstrating subsidence of > 2 mm. Periprosthetic radiolucency of > 2 mm was found in 26% of the implants in zone 1 and in 9% in zone 7, respectively. Neither the amount of subsidence nor proximal periprosthetic radiolucency was associated with aseptic loosening or worse clinical outcomes.
CONCLUSIONS: Comparable to other proximally fixed short stem designs, the highest subsidence was observed within the first 3 months following implantation. No demographic or implant-related factors were found to have a statistically significant influence on stem subsidence. Periprosthetic radiolucency and subsidence of the AMISstem is not correlated with worse clinical outcomes at 5-year follow-up.

Entities:  

Keywords:  Cementless; EBRA; Metaphyseal-anchored femoral stem; Subsidence; Total hip arthroplasty

Mesh:

Substances:

Year:  2021        PMID: 33413463      PMCID: PMC7792117          DOI: 10.1186/s13018-020-02104-8

Source DB:  PubMed          Journal:  J Orthop Surg Res        ISSN: 1749-799X            Impact factor:   2.359


  42 in total

1.  Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030.

Authors:  Steven Kurtz; Kevin Ong; Edmund Lau; Fionna Mowat; Michael Halpern
Journal:  J Bone Joint Surg Am       Date:  2007-04       Impact factor: 5.284

2.  Does early micromotion of femoral stem prostheses matter? 4-7-year stereoradiographic follow-up of 84 cemented prostheses.

Authors:  J Kärrholm; B Borssén; G Löwenhielm; F Snorrason
Journal:  J Bone Joint Surg Br       Date:  1994-11

3.  A randomised trial comparing a short and a standard-length metaphyseal engaging cementless femoral stem using radiostereometric analysis.

Authors:  R W McCalden; A Korczak; L Somerville; X Yuan; D D Naudie
Journal:  Bone Joint J       Date:  2015-05       Impact factor: 5.082

Review 4.  Survival of short hip stems with a "modern", trochanter-sparing design - a systematic literature review.

Authors:  Elke Rometsch; Pieter K Bos; Bart W Koes
Journal:  Hip Int       Date:  2012 Jul-Aug       Impact factor: 2.135

5.  Early migration and late aseptic failure of proximal femoral prostheses.

Authors:  M A Freeman; P Plante-Bordeneuve
Journal:  J Bone Joint Surg Br       Date:  1994-05

6.  Uncemented Total Hip Arthroplasty Through the Direct Anterior Approach: Analysis of a Consecutive Series of 275 Hips With a Minimum Follow-Up of 10 Years.

Authors:  Stefan Rahm; Timo Tondelli; Sylvain Steinmetz; Pascal Schenk; Claudio Dora; Patrick O Zingg
Journal:  J Arthroplasty       Date:  2019-01-31       Impact factor: 4.757

7.  No increased stem subsidence after arthroplasty in young patients with femoral head osteonecrosis: 41 patients followed for 1-9 years.

Authors:  Akif Ince; Johannes Lermann; Sascha Göbel; Nicole Wollmerstedt; Christian Hendrich
Journal:  Acta Orthop       Date:  2006-12       Impact factor: 3.717

8.  Future young patient demand for primary and revision joint replacement: national projections from 2010 to 2030.

Authors:  Steven M Kurtz; Edmund Lau; Kevin Ong; Ke Zhao; Michael Kelly; Kevin J Bozic
Journal:  Clin Orthop Relat Res       Date:  2009-04-10       Impact factor: 4.176

9.  Prospective five-year subsidence analysis of a cementless fully hydroxyapatite-coated femoral hip arthroplasty component.

Authors:  Martin Clauss; Catherine Van Der Straeten; Marc Goossens
Journal:  Hip Int       Date:  2014-01-24       Impact factor: 2.135

10.  Migration analysis of a metaphyseal-anchored short femoral stem in cementless THA and factors affecting the stem subsidence.

Authors:  Michael O Schaer; Michael Finsterwald; Iris Holweg; Dimitris Dimitriou; Alexander Antoniadis; Naeder Helmy
Journal:  BMC Musculoskelet Disord       Date:  2019-12-12       Impact factor: 2.362

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.