BACKGROUND: The purpose of this study was to document complications, outcomes, and 10-year survivorship of primary total hip arthroplasty (THA) using a direct anterior approach with an uncemented, straight, hydroxyapatite-coated stem and an uncemented cup. METHODS: A retrospective, consecutive series of 275 primary THAs through a direct anterior approach with traction table using Medacta Versafit cup and Quadra-H stem with a minimum of 10-year follow-up was identified. The cumulative 10-year survival of the implants was estimated using Kaplan-Meier estimator. All complications, reoperations, and failures were analyzed. Subjective and clinical outcomes (Subjective Hip Value, Western Ontario and McMaster Universities Osteoarthritis Index, and Harris Hip Score) were measured. RESULTS: Of 256 patients (275 hips, 143 men and 113 women) with a mean age of 63 (range, 24-85) years, 48 (19%) patients (52 hips) deceased not related to the surgery after a mean time 49 months (range, 3-118) postoperatively. At >10-year follow-up, 9 THAs were revised. The overall implant survival rate was 96.8% (95% confidence interval, 94.4-98.7) at 10 years. One cup and 1 stem were revised because of aseptic loosening. At the last follow-up, the median Subjective Hip Value was 90% (range, 20-100), the Western Ontario and McMaster Universities Osteoarthritis Index score reached a median of 0.2 points (range, 0-6.3), and the median Harris Hip Score points was 99 (range, 29-100). CONCLUSION: Primary THA through an anterior minimal invasive approach with the mentioned implants showed low revision rates and good to excellent clinical outcome after at least 10 years.
BACKGROUND: The purpose of this study was to document complications, outcomes, and 10-year survivorship of primary total hip arthroplasty (THA) using a direct anterior approach with an uncemented, straight, hydroxyapatite-coated stem and an uncemented cup. METHODS: A retrospective, consecutive series of 275 primary THAs through a direct anterior approach with traction table using Medacta Versafit cup and Quadra-H stem with a minimum of 10-year follow-up was identified. The cumulative 10-year survival of the implants was estimated using Kaplan-Meier estimator. All complications, reoperations, and failures were analyzed. Subjective and clinical outcomes (Subjective Hip Value, Western Ontario and McMaster Universities Osteoarthritis Index, and Harris Hip Score) were measured. RESULTS: Of 256 patients (275 hips, 143 men and 113 women) with a mean age of 63 (range, 24-85) years, 48 (19%) patients (52 hips) deceased not related to the surgery after a mean time 49 months (range, 3-118) postoperatively. At >10-year follow-up, 9 THAs were revised. The overall implant survival rate was 96.8% (95% confidence interval, 94.4-98.7) at 10 years. One cup and 1 stem were revised because of aseptic loosening. At the last follow-up, the median Subjective Hip Value was 90% (range, 20-100), the Western Ontario and McMaster Universities Osteoarthritis Index score reached a median of 0.2 points (range, 0-6.3), and the median Harris Hip Score points was 99 (range, 29-100). CONCLUSION: Primary THA through an anterior minimal invasive approach with the mentioned implants showed low revision rates and good to excellent clinical outcome after at least 10 years.
Authors: A Zimmerer; M Hauschild; R Nietschke; M M Schneider; G Wassilew; C Sobau; W Miehlke Journal: Arch Orthop Trauma Surg Date: 2020-10-12 Impact factor: 3.067