Complications and Effectiveness of Intramedullary Limb Lengthening: A Matched Pair Analysis of Two Different Lengthening Nails.

Peter H Thaller1, Felix Frankenberg2, Nikolaus Degen1, Chris Soo3, Florian Wolf1, Ekkehard Euler1, Julian Fürmetz1.   

Abstract

BACKGROUND: Intramedullary limb lengthening has become an accepted concept in reconstructive surgery, but as yet comparative clinical studies are missing. We compared the complications and effectiveness of two types of intramedullary limb lengthening devices (ISKD®; Fitbone®).
MATERIALS AND METHODS: In a retrospective series of 278 consecutive patients with internal limb lengthening, we found 17 matching pairs in terms of predefined matching parameters (group I with ISKD® and group II with Fitbone®). The surgeries were all performed with the same technique and managed with equivalent pre- and postoperative treatment protocols. The performance of the implants was evaluated using the distraction index and the weight-bearing index. Complications were rated according to Paley's classification for external lengthening.
RESULTS: The distraction index in group I (ISKD®) was 0.99 mm/day (range 0.55-1.67) and in the group II (Fitbone®) 0.55 mm/day (range 0.14-0.92) (p value = 0.001). The mean weight-bearing index differed between group I and group II from 32.0 day/cm (range 16.4-64.0) to 51.6 day/cm (25.8-95.0) (p value = 0.001). There were 17 recorded incidents in group I and 19 in group II during lengthening.
CONCLUSION: Specific technical handicaps of the two systems, such as the so-called runaway of the ISKD® and backtracking of the Fitbone® nails seem to result in different distraction index and weight-bearing index. Further comparative studies might induce technical progress in intramedullary limb lengthening. HOW TO CITE THIS ARTICLE: Thaller PH, Frankenberg F, Degen N, et al. Complications and Effectiveness of Intramedullary Limb Lengthening: A Matched Pair Analysis of Two Different Lengthening Nails. Strategies Trauma Limb Reconstr 2020;15(1):7-12.
Copyright © 2020; Jaypee Brothers Medical Publishers (P) Ltd.

Entities:  

Keywords:  Distraction osteogenesis; Fitbone®; ISKD®; Intramedullary lengthening nails; Limb lengthening

Year:  2020        PMID: 33363635      PMCID: PMC7744673          DOI: 10.5005/jp-journals-10080-1448

Source DB:  PubMed          Journal:  Strategies Trauma Limb Reconstr        ISSN: 1828-8928


Introduction

Intramedullary lengthening nails have several advantages in the management of leg length discrepancies such as lower infection rates, less soft tissue damage and pain, better joint movement, better alignment control, and more patient comfort compared to external devices.[1-6] With the rising number of intramedullary lengthening procedures over the past years, six different fully implantable intramedullary nails have been investigated in the literature, including the Bliskunov-Nail, the Albizzia-Nail, and the Phenix® nail, which are no longer available; and the Intramedullary Skeletal Kinetic Distractor (ISKD®, Orthofix, McKinney), which is now available again in Europe.[3,5,7,8] Currently, the Fitbone® nail (Wittenstein Intens1, Igersheim, Germany) and the Precice® nail (Ellipse Technologies Inc., CA, USA) are the most frequently implanted systems worldwide.[6,9-12] These nails are characterised by different actuation mechanisms, and each have their specific limitations. The Phenix® nail and the Precice® nails are magneto-mechanically driven systems; the Bliskunov-nail, the Albizzia-nail, and the ISKD® are mechanically driven; and the Fitbone® nail is a motorised (electromotive) system.[2,3,6-8,11,13] In the literature, the main problems of intramedullary lengthening devices mentioned are control of distraction rate, maintenance of distraction and implant failure.[8,9,14-16] The Fitbone® nail was developed at our clinic in 1990.[1] In 2006, we reported on 150 lengthening procedures with the Fitbone® nail.[17] A total of 144 (96%) patients reached the distraction goal, there was no infection, implant failure occurred in 3 cases, and 9 other instances of technical difficulty were reported. The unintended back tracking of the implant as described in this early paper was detected in nine (6%) cases. In 2009, we gathered first-hand experience with the mechanically driven ISKD® nail that was first described by Dean Cole in 2001.[2] At the time it was the most implanted system and by far the most published one. In our initial comparison of our early results to the Fitbone® nail, we also observed a few handicaps of ISKD® nail.[18] It was withdrawn from the US market, temporarily in 2009 and finally in 2012. The main reasons were jammed or non-distracting nails, leading to early consolidation,[14] and unintended high distraction rates (“runaway” or “accelerated” nail).[15] A scientific evaluation between intramedullary lengthening nails is difficult, as the number of implantations is low worldwide and techniques vary significantly between surgeons. By now, we were able to gather experience with five different intramedullary lengthening devices.[13] On those and other intramedullary lengthening nails, there are multiple publications of specific case series.[1-3,8,11,14] There are also studies comparing intramedullary lengthening with lengthening over nail (LON) or with LRS external fixator monorail system.[19,20] We found two matched-pair studies on LON vs external lengthening (Ilizarov) and one on intramedullary lengthening (Fitbone®) vs external lengthening (Taylor Spatial Frame).[21-23] But there are no matched pairs studies that compare specific intramedullary lengthening devices. This is the first matched pair analysis comparing two different intramedullary lengthening devices (ISKD® and Fitbone®).

Materials and Methods

In a retrospective review of a consecutive series, we found 278 patients who had lengthening of the lower limb between March 1999 and November 2011. The study was approved by our Institutional Review Board. From March 1999 to June 2009, we performed 241 limb lengthening procedures with a fully implantable electromechanical-actuated intramedullary nail (Fitbone®). From July 2009 to November 2011, 37 consecutive limb lengthening procedures were performed using a fully implantable mechanical-actuated intramedullary nail (ISKD®). All procedures were conducted by the same surgeon with an equivalent technique (see below).

Matching Criteria

The case matching was done as per agreement in five criteria: Equivalent surgical technique and pre- and postoperative treatment protocol Site of osteotomy (proximal femur, distal femur or proximal tibia) Simultaneous realignment of the mechanical axis (yes or no) Amount of lengthening (maximum variation 10%) Age (maximum variation 20%). All five criteria had to comply with both matching partners. Subsequent lengthening of another bone, implant failure due to malpractice by the patient or accidental trauma, or nicotine abuse during lengthening led to exclusion. Mismatch in sex and aetiology of leg length discrepancy were not included in the matching criteria.

Outcome Measures

Primary outcome measures included completed lengthening, distraction index (DI) in mm/day, and weight bearing index (WBI) in day/cm, as described in former publications.[4,5,8,9,12,15,24] We classified problems, obstacles, and complications as they are described for external lengthening by Paley in 1990.[25] Implant-related complications were specified as recommended by Lee et al.[26]

Surgical Technique and Treatment Protocol

Preoperative analysis, end-point-first planning, and operative technique with protective sleeves, rigid reaming, X-ray-grid verification were the same in all procedures. We have previously published detailed descriptions of our operative technique for the tibia and femur.[27-29] Postoperatively, patients received calcium and vitamin D, low-molecular-weight heparin and analgesics according to the WHO scheme (with the exception of NSAIDs). Prophylactic antibiotic treatment was maintained for as long as drains were in situ, in most cases for 1 day. All patients had regular physical therapy and were mobilised with partial weight bearing of 20 kg from the first day on. All patients were instructed to conduct lengthening procedures autonomously. The distraction process started on the fifth postoperative day and was controlled radiologically at least twice prior the patient's discharge to confirm that the distraction gap is opening. After discharge, clinical examination and radiological control were performed every week. In the majority of cases, the intended DI was 1 mm/day in all patients, with 0.33 mm lengthening applied three times daily. In very few cases the distraction speed was adjusted according to radiological signs of consolidation of the bony regenerate, conditions of joints and neurovascular structures. The range of motion of adjacent joints was controlled at each patient visit. After reaching the distraction goal, consolidation was monitored initially every 2 weeks and after partial consolidation every 6 weeks. As soon as one cortical side of the regenerate was deemed radiologically stable, weight bearing was increased gradually until full weight bearing was achieved within 6 weeks at the latest. The WI was calculated by the start of full weight bearing. Full consolidation and removal of the implants were achieved after 12–18 months. Six months after removal, a final reassessment of clinical and radiological parameters, including the leg length and alignment, was performed (Fig. 1).
Figs 1A to G

(A) Preoperative long-standing radiograph with post-traumatic shortening with varus deformity of the right femur; (B) Preoperative analysis; (C) End-point-first (EPF) planning for simultaneous lengthening and axis correction with retrograde lengthening nail (ISKD); (D) Ongoing lengthening, the blocking screw in the metaphyseal fragment stabilises the nail for the necessary valgisation, in the diaphyseal part, the medial and dorsal blocking screw enhances the friction at the rotating ISKD shaft for a better control of lengthening; (E) Complete lengthening achieved; (F) Long-standing radiograph 6 months after implant removal; (G) Final clinical result, follow-up 4.5 years after lengthening

Statistics

All matched variables were evaluated for statistical significance between groups using a Wilcoxon test for paired variables (SPSS 23). p values of less than 0.05 were regarded as statistically significant.

IRB approval

The study was carried out according to the Declaration of Helsinki. The Ethic Committee of the University of Munich approved this study with the ID number 8–16.

Results

From 37 ISKD® patients, we excluded two patients due to a simultaneous lengthening of the tibia or femur and two patients due to nicotine abuse during lengthening. From 33 remaining ISKD® patients, 17 patients (group I) could be matched to 17 Fitbone® patients (group II) respecting all five of the aforementioned criteria. The pairs of lengthening procedures were 4 proximal femoral, 7 distal femoral, and 6 proximal tibial (Table 1).
Table 1

Matching partner with patient characteristics and results of DI and WI

ISKD®Fitbone®
Study numberSexAge at surgery (year)Location of osteotomy (proximal/distal femur, tibia)Stroke (mm)DI (mm/day)BI (day/cm)Study numberSexAge at surgery (y)Location of osteotomy (proximal/distal femur, tibia)Stroke (mm)DI (mm/day)BI (day/cm)
I-1w14Proximal501.2525.20F-1w14Proximal550.7625.82
I-2m48Proximal500.6726.20F-2m41Proximal550.2942.55
I-3w25Proximal330.9226.67F-3m21Proximal350.9236.86
I-4m20Proximal271.0427.41F-4w18Proximal300.1472.67
I-5w39Distal520.7621.15F-5m43Distal480.2655.83
I-6w16Distal451.1316.44F-6w15Distal440.4143.41
I-7w63Distal451.6764.00F-7w49Distal490.6060.20
I-8w25Distal401.2932.75F-8w28Distal420.6538.10
I-9m29Distal281.1239.29F-9m24Distal280.2650.71
I-10m43Distal260.5531.54F-10w47Distal250.8160.00
I-11w47Distal210.6452.86F-11w36Distal220.5955.00
I-12w15Tibia501.0419.40F-12w15Tibia500.4346.80
I-13w15Tibia481.0923.13F-13w15Tibia500.6854.60
I-14w16Tibia440.8140.91F-14w15Tibia430.4336.51
I-15w38Tibia400.8732.75F-15m37Tibia440.8540.45
I-16m32Tibia391.0521.03F-16m25Tibia420.2695.00
I-17w17Tibia300.9743.67F-17w17Tibia320.7863.13
Mean follow-up of all selected patients was 2.5 years (range 1.3–7.0 years). All 34 patients completed the lengthening process. The average DI for ISKD® patients was 0.99 mm/day (range 0.55–1.67 mm/day) and for Fitbone® patients 0.55 mm/day (range 0.14–0.92 mm/day). The difference in DI between group I and II was significant (p = 0.001). Mean WBI was 32.0 day/cm (range 16.4–64.0 day/cm) in the ISKD group and 51.6 day/cm (range 25.8–95.0 day/cm) in the Fitbone® group. The WBI between group I and II was significantly different (p = 0.001). Differences between the three osteotomy sites independent from group I or II were not significant (p = 0.875). Detailed patient data and results are shown in Table 1. (A) Preoperative long-standing radiograph with post-traumatic shortening with varus deformity of the right femur; (B) Preoperative analysis; (C) End-point-first (EPF) planning for simultaneous lengthening and axis correction with retrograde lengthening nail (ISKD); (D) Ongoing lengthening, the blocking screw in the metaphyseal fragment stabilises the nail for the necessary valgisation, in the diaphyseal part, the medial and dorsal blocking screw enhances the friction at the rotating ISKD shaft for a better control of lengthening; (E) Complete lengthening achieved; (F) Long-standing radiograph 6 months after implant removal; (G) Final clinical result, follow-up 4.5 years after lengthening Patient I-7 had had a long history of previous surgeries with a wider cavity. She additionally suffered from severe agitation and excessive postoperative movement due to drug addiction and unexpected withdrawal symptoms after the surgery. The acute motor agitation triggered accidental distractions in the wider cavity. In an attempt to slow down lengthening, three blocking screws were implanted in an attempt to increase friction between the nail and cortical bone. Additionally, an epidural catheter permitted intermediate motor paresis. In this elderly patient, insufficient callus formation was managed by an exchange to a straightened conventional intramedullary nail and bone grafting after 6 months, with no further problems and full consolidation.

Complications

Regarding Paley's classification for external lengthening, we noted no complication in either of the groups. In the ISKD® group (I), we observed 11 problems in 7 patients and 6 obstacles in 4 patients. Problems in group I: 3× pain treated in an inpatient setting with peridural catheter or other analgesics, 1× accelerated nail treated with peridural catheter to reduce patient movements, 2× equinus treated with physiotherapy successfully, 1× broken interlocking bolt, 2× temporary peroneal nerve irritation, 1× superficial wound defect, and 1× thrombosis of the popliteal vein. Obstacles in group I: There were two cases of early consolidation treated by re-osteotomy, which did not affect the final outcome. Aforementioned accelerated nail treated with peridural catheter received additionally three blocking screws to increase the friction. In the Fitbone® group (II) there were 13 problems in 11 patients and 6 obstacles in 5 patients. Problems in group II: 1× defect in external controller, 7× back tracking of the nail (DI below 0.5 mm/day), 3× temporary peroneal nerve irritation, 1× periostitis, 1× equinus foot. Obstacles in group II: 3× loosening of interlocking bolts (retightening or exchange), 2× insufficient regenerate by following cancellous bone grafting, 1× osteophyte requiring removal. There were neither breakage nor deep infection in either of the groups. But using Lee's classification for complications with internal lengthening devices we noted 15 implant related complications.[26] There were four implant related complications with ISKD®: 3× type I distraction control related (1× accelerated nail, 2× early consolidations), 1× type II stability related (broken interlocking bolt). Additionally, there were 11 implant-related complications with Fitbone®: 8× type I distraction control related (1× defect controller, 7× back tracking), 3× type II stability related (loosening of interlocking bolts). No implant related complication affected the final outcome.

Discussion

Until now, there are only case series on individual implants to evaluate treatment efficacy and complications of intramedullary limb lengthening devices. This is the first comparison of two different implants (ISKD® and Fitbone®) in a matched pair analysis. Our results show a similar rate of problems, obstacles, and complications but indicate significant differences in the lengthening and consolidation phase (DI and WBI) between the two devices. A problem frequently reported with the ISKD® is the insufficient control of distraction speed. An unintentionally distracting nail with a DI of more than 1.5 mm/day was named as an “accelerated nail” or “runaway nail”. From 2001 to 2012, the ISKD® nail was the most implanted intramedullary lengthening device. The earliest ISKD® complication, a 10% implant failure rate, was reported by Cole et al., with further studies demonstrating a significant risk of early consolidation resulting from non-distracting ISKD®s.[2] More recently, excessive distraction by runaway nails has been reported, with resultant non-union and insufficient bone regeneration.[2,5,30] Runaway and non-distracting ISKD® incidence have been reported to be up to 50%.[30] In 2011, Burghardt et al. published the largest retrospective study about 242 ISKD® lengthening procedures.[14] 15 ISKD®s (6.2%) failed mechanically, with 10 cases of device fracture. After a further 4 devices reportedly failed to lengthen, thought to be due to an error in the manufacturing process, a worldwide recall of the implant was enacted. However, no cases of “runaway nail” were reported. Matching partner with patient characteristics and results of DI and WI In 2006, we reported on 297 Fitbone® applications worldwide. 181 implantations were performed at our university hospital and 150 cases were retrospectively evaluated. A fracture of the nail occurred in 2% of the cases and unintended back tracking in 5%.[17] Fitbone® lengthening devices are not widely available; they are restricted to few clinical centres worldwide, and with the formation of a new group we introduced the ISKD in 2009. Krieg et al. observed three (9.4 %) cases of backtracking in the consolidation phase of 32 Fitbone® applications.[12] Backtracking seems to be caused by thousands of repetitive micro-movements at the telescoping part of the nail. These micro-movements are transmitted through the gearbox to the electric motor and cause shortening until the bone is consolidated. In a recently published study of 23 Fitbone® implants, no backtracking was reported and thought to be attributed to a minor technical improvement made by the manufacturer intended to prevent backtracking of the system.[31] Future independent studies with a larger sample size might further determine if this problem has been solved. In a number of Fitbone® implantations, we observed signs of nail corrosion at the site, where the telescoping part leaves the main body of the nail. In some cases, we observed osteolysis at this site, and in very few cases, a painful tumorigenic reaction which resolved immediately after removal.[18] But this was not observed in the patients included in this matched pairs study. For the ISKD®, the DI is reported to be from 1.1 to 1.5 mm/day for the femur and 0.5 to 0.9 mm/day for the tibia.[5,15,24,32] Case series with the Fitbone® demonstrate a DI from 1.0 to 1.1 mm/day.[10,33] In literature, the WBI ranges from 23.3 to 47.8 days/cm for ISKD® and 30.5 to 35.0 days/cm for the Fitbone®.[10,30,32,33] The WBI is not reported in every case series with intramedullary limb lengthening and is often described as the healing index. Historically, the terms healing index, consolidation index and WBI are related to the type of distraction in regard to external vs internal implants. Some authors mention a consolidation index rather than the WBI for fully implantable nails or lengthening over nail techniques. When reviewing the literature on Fitbone or ISKD® devices, our findings were somewhat unexpected. In the ISKD® group, we could find no incidences of distraction failure, and there were seven cases of backtracking in the Fitbone® group, representing 41% of cases. In our results, only one ISKD® patient had a DI of more than 1.5 mm/day (patient I-7). Our hypothesis is that none of our cases had been over reamed more than 0.5 mm for implantation of “straight” lengthening nails. Since the medullary cavities of the lower extremity are anatomically curved (femur to anterior and tibia to posterior), flexible reaming results in excessive over-reaming in an attempt to implant the straight (non-curved) device. There were no cases of accelerated lengthening in the Fitbone® group. Unintended lengthening rarely seems possible with the Fitbone® technique. Frequent backtracking of the Fitbone® nail resulted in a range of DI from 0.14 to 0.9 mm/day. In the case of the patient with a DI of 0.14 mm/day, backtracking had already occurred twice during the distraction phase. In order to prevent loss of lengthening during the consolidation of the regenerate, a continuous distraction was necessary even after reaching the targeted length. In the other cases of backtracking, only one had a DI of more than 0.5 mm/day. We hypothesize that perhaps the design of the Fitbone®, with no intrinsic mechanical mechanism for preventing backtracking could be the reason for this.

Limitations

The main limitation of the study is a low number of matched pairs, due to challenging matching criteria and the lower caseload in the ISKD® group. We chose the matching criteria like this: (1) Because there are significantly diverging techniques, protocols, and a certain learning curve, for consistency, all patients had to be operated by the same surgeon using the same protocol. (2) Bone formation cannot easily be compared between different osteotomy sites, as it depends on the available blood supply at each location.[34] (3) Simultaneous realignment was matched because of this risk of impaired bone formation by more offset and/or wider open wedges between fragments. (4) The amount of lengthening has a huge influence on the entire treatment, and so we accepted a variance not more than 10%.[22] (5) The osteogenic potential decreases over the span of a lifetime,[35] therefore we matched patients with a maximum variance of 20% in age. But aetiology was not included in the matching criteria, which can be an important factor for example in cases of CFD (congenital femoral deficiency) with poorer callus formation or Ollier's disease with quicker formation. There are several other factors which might influence the results of limb lengthening procedures like the conditions of soft tissues, aetiology of the deformity, bone mineral density, and others. Still, we choose only the above-mentioned major matching criteria to prevent an even smaller number of matching partners. We estimate that with more matching criteria or tighter ranges for the specific criteria the number needed to treat would grow higher than any institution can provide. Another important limitation is that one of the implants has been withdrawn from the market, and the other is not freely available. Nevertheless, we consider the data to be relevant with regard to other internal lengthening systems and believe that further studies may benefit from the methodology used. One of the issues raised during the study was the lack of an appropriate classification system for internal lengthening systems to facilitate the evaluation of the systems. Although Lee's classification implements specific complications of internal lengthening devices, not all complications such as regenerative insufficiency have been addressed.

Conclusion

Due to a low incidence of limb lengthening procedures, there is a lack of valid studies on intramedullary lengthening devices. Direct comparison of specific devices with equivalent treatment protocols and a higher caseload might provide further important information on the efficacy and complications of these devices. Regarding the lack of control of ISKD® nails in the literature, our results indicate a better control of distraction and patients might have benefited by rigid reaming. Our rate of backtracking of Fitbone® nails was higher than previously reported. We recommend close follow-up in all procedures in order to facilitate the early detection of specific problems such as backtracking, jamming, implant instability, late or early consolidation, contractures, or neurovascular lesions. Ethical committee approval was obtained prior to the study. Informed consent was obtained from all individual participants included in the study.
  30 in total

1.  Mechanical failure of the Intramedullary Skeletal Kinetic Distractor in limb lengthening.

Authors:  R D Burghardt; J E Herzenberg; S C Specht; D Paley
Journal:  J Bone Joint Surg Br       Date:  2011-05

2.  Precision of the PRECICE internal bone lengthening nail.

Authors:  Yatin M Kirane; Austin T Fragomen; S Robert Rozbruch
Journal:  Clin Orthop Relat Res       Date:  2014-12       Impact factor: 4.176

3.  Limb lengthening with fully implantable magnetically actuated mechanical nails (PHENIX(®))-preliminary results.

Authors:  Peter Helmut Thaller; Julian Fürmetz; Florian Wolf; Thorsten Eilers; Wolf Mutschler
Journal:  Injury       Date:  2013-10-28       Impact factor: 2.586

4.  Impacted Steel Sleeves for a minimally invasive approach in intramedullary nailing.

Authors:  F Wolf; S Bösl; N Degen; J Fürmetz; P H Thaller
Journal:  Injury       Date:  2019-08-01       Impact factor: 2.586

5.  Intramedullary skeletal kinetic distractor in the treatment of leg length discrepancy--a review of 16 cases and analysis of complications.

Authors:  Kemble Wang; Elton Edwards
Journal:  J Orthop Trauma       Date:  2012-09       Impact factor: 2.512

6.  [Intramedullary distraction of the femur (preliminary report)].

Authors:  A I Bliskunov
Journal:  Ortop Travmatol Protez       Date:  1983-10

7.  The intramedullary skeletal kinetic distractor (ISKD): first clinical results of a new intramedullary nail for lengthening of the femur and tibia.

Authors:  J D Cole; D Justin; T Kasparis; D DeVlught; C Knobloch
Journal:  Injury       Date:  2001-12       Impact factor: 2.586

8.  Outcomes following femoral lengthening: An initial comparison of the Precice intramedullary lengthening nail and the LRS external fixator monorail system.

Authors:  M Laubscher; C Mitchell; A Timms; D Goodier; P Calder
Journal:  Bone Joint J       Date:  2016-10       Impact factor: 5.082

9.  Improved comfort in lower limb lengthening with the intramedullary skeletal kinetic distractor. Principles and preliminary clinical experiences.

Authors:  Stefan Hankemeier; Hans-Christoph Pape; Thomas Gosling; Tobias Hufner; Martinus Richter; Christian Krettek
Journal:  Arch Orthop Trauma Surg       Date:  2004-01-27       Impact factor: 3.067

10.  Intramedullary leg lengthening with a motorized nail.

Authors:  Andreas H Krieg; Ulrich Lenze; Bernhard M Speth; Carol C Hasler
Journal:  Acta Orthop       Date:  2011-05-11       Impact factor: 3.717

View more
  1 in total

1.  Limb reconstruction in a resource-limited environment.

Authors:  Nando Ferreira; Sanjeev Sabharwal; Gamal Ahmed Hosny; Hemant Sharma; Ashok Johari; Vasudevan P Nandalan; Mauro Vivas; Mangal Parihar; Selvadurai Nayagam; David Ferguson; Jan Duedal Rölfing
Journal:  SICOT J       Date:  2021-12-31
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.