Literature DB >> 33353399

Comparison of the image quality and radiation dose of different scanning modes in head-neck CT angiography.

Tianle Zhang1, Xue Geng2, Dongxue Li1, Yize Xu1, Yongxia Zhao1.   

Abstract

OBJECTIVES: To analyze and compare the radiation dose and image quality of different CT scanning modes on head-neck CT angiography.
METHODS: A total of 180 patients were divided into Group A and Group B. The groups were further subdivided according to different scanning modes: subgroups A1, A2, A3, B1, B2, and B3. Subgroups A1 and B1 used conventional CT protocol, subgroups A2 and B2 used the kV-Assist scan mode, and subgroups A3 and B3 used the dual-energy gemstone spectral imaging protocol. The CT dose index and dose-length product were recorded. The objective image quality and subjective image evaluation was conducted by two independent radiologists.
RESULTS: The signal-to-noise ratios, contrast-to-noise ratios, and subjective scores of subgroups A3 and B3 were higher than the other subgroups. In subgroups B1 and B2, the subjective scores of 9 patients and 12 patients were lower than 3, respectively. The subjective scores of subgroups B1 and B2 were lower than the other subgroups. There was no statistically significant difference in signal-to-noise ratios, contrast-to-noise ratios, and subjective scores between subgroups A1 and A2. The effective dose of subgroup A2 was 41.7 and 36.4% lower than that in subgroups A1 and A3, respectively (p < 0.05). In Group B, there were no statistically significant differences in CT dose indexvol, dose-length product, and ED among the subgroups (p > 0.05).
CONCLUSION: In the head-neck CT angiography, the kV-Assist scan mode is recommended for patients with body mass index between 18.5 and 34.9 kg m-2; gemstone spectral imaging scanning mode is recommended for patients with body mass index ≥34.9 kg m-2.

Entities:  

Keywords:  Head-neck CTA; Image quality; Radiation dose; Scan mode

Mesh:

Year:  2021        PMID: 33353399      PMCID: PMC8231676          DOI: 10.1259/dmfr.20200428

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   3.525


  29 in total

1.  Iterative reconstruction in head CT: image quality of routine and low-dose protocols in comparison with standard filtered back-projection.

Authors:  A Korn; M Fenchel; B Bender; S Danz; T K Hauser; D Ketelsen; T Flohr; C D Claussen; M Heuschmid; U Ernemann; H Brodoefel
Journal:  AJNR Am J Neuroradiol       Date:  2011-10-27       Impact factor: 3.825

2.  Feasibility of low-concentration iodinated contrast medium with lower-tube-voltage dual-source CT aortography using iterative reconstruction: comparison with automatic exposure control CT aortography.

Authors:  Hee Jeong Shin; Song Soo Kim; Jae-Hwan Lee; Jae-Hyeong Park; Jin-Ok Jeong; Seon Ah Jin; Byung Seok Shin; Kyung-Sook Shin; Moonsang Ahn
Journal:  Int J Cardiovasc Imaging       Date:  2015-11-30       Impact factor: 2.357

3.  Aorto-iliac multidetector-row CT angiography with low kV settings: improved vessel enhancement and simultaneous reduction of radiation dose.

Authors:  B Wintersperger; T Jakobs; P Herzog; S Schaller; K Nikolaou; C Suess; C Weber; M Reiser; C Becker
Journal:  Eur Radiol       Date:  2004-12-21       Impact factor: 5.315

4.  The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103.

Authors: 
Journal:  Ann ICRP       Date:  2007

5.  Initial phantom study comparing image quality in computed tomography using adaptive statistical iterative reconstruction and new adaptive statistical iterative reconstruction v.

Authors:  Kyungjae Lim; Heejin Kwon; Jinhan Cho; Jongyoung Oh; Seongkuk Yoon; Myungjin Kang; Dongho Ha; Jinhwa Lee; Eunju Kang
Journal:  J Comput Assist Tomogr       Date:  2015 May-Jun       Impact factor: 1.826

6.  Coronary CT angiography radiation dose trends: A 10-year analysis to develop institutional diagnostic reference levels.

Authors:  Zhu Xiao Lin; Chang Sheng Zhou; U Joseph Schoepf; Marwen Eid; Taylor M Duguay; William T Greenberg; Song Luo; Wei Quan; Fan Zhou; Guang Ming Lu; Long Jiang Zhang
Journal:  Eur J Radiol       Date:  2019-02-13       Impact factor: 3.528

7.  Effect of an Arm Traction Device on Image Quality and Radiation Exposure during Neck CT: A Prospective Study.

Authors:  Y J Choi; J H Lee; D H Yoon; H J Kim; K J Seo; K-H Do; J H Baek
Journal:  AJNR Am J Neuroradiol       Date:  2017-11-09       Impact factor: 3.825

8.  Successful Dose Reduction Using Reduced Tube Voltage With Hybrid Iterative Reconstruction in Pediatric Abdominal CT.

Authors:  Sheila C Berlin; Dayna M Weinert; Pauravi S Vasavada; Claudia Martinez-Rios; Roshni A Parikh; Michael A Wien; David W Jordan; Ronald D Novak
Journal:  AJR Am J Roentgenol       Date:  2015-08       Impact factor: 3.959

9.  Exposure to low dose computed tomography for lung cancer screening and risk of cancer: secondary analysis of trial data and risk-benefit analysis.

Authors:  Cristiano Rampinelli; Paolo De Marco; Daniela Origgi; Patrick Maisonneuve; Monica Casiraghi; Giulia Veronesi; Lorenzo Spaggiari; Massimo Bellomi
Journal:  BMJ       Date:  2017-02-08

10.  Radiation Dose Optimization in Pediatric Chest CT: Major Indicators of Dose Exposure in 1695 CT Scans over Seven Years.

Authors:  Michael Esser; Sabine Hess; Matthias Teufel; Mareen Sarah Kraus; Sven Schneeweiß; Sergios Gatidis; Juergen F Schaefer; Ilias Tsiflikas
Journal:  Rofo       Date:  2018-10-11
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.